[Page 1] No. 2D730-148E”O SERVICE MANUAL FOR DIAGNOSTIC ULTRASOUND SYSTEM MODELS SSA-700A/SSA-770A (2D730-148E”O) TOSHIBA MEDICAL SYSTEMS CORPORATION @I TOSHIBA MEDICAL SYSTEMS CORPORATION 2001-2008 ALL RIGHTS RESERVED
[Page 2] IMPORTANT! 1. No part of this manual may be copied or reprinted, in whole or in pat-t, without written permission. 2. The contents of this manual are subject to change without prior notice and without our legal obligation. C-l
[Page 3] REVISION RECORD REV. DATE REASON PAGE SER. DOC. (MM/W) No. PRODUCT. /AUTHOR CHANGED INI. 1 07/‘01 1 Mr. Kuramata - ------ *A 1 I/‘01 support of VI .5 Mr. Tamura *B 04/‘02 Support of V2.0 for SSA-770A (and additions) Mr. Ogasawarai 1 1 *C 06/‘02 Mr. Shibanuma *D 1 I/‘02 support of v3.0 Mr. Ogasawara *E 07/‘03 support of v4.0 Mr. Nakai *F 08/‘03 Change of the AC/DC power supply etc. Mr. Fujiwara *G 1 IO/‘03 ID ue to change of company name *H IO/‘03 Change of the AC/DC power supply Mr. Fujiwara *I 1 12/‘03 Support of V5.0 *J 02/‘04 Addition of section 17 Mr. Nakai “K 1 06/‘04 Changes to section 17 *L 02/‘05 Support of V6.0 Mr. Nakai KD-WPX *M 12/‘05 support of v7.0 Mr. Nakai *N 09’06 Addition of systems with the LCD monitor Mr. Nakai 02/‘08 Support of Doppler Measurement Preset TM-WP2 Mr. Nakai R-l
[Page 4] Safety Precautions 1. Meaning of Signal Words ln this manual, the signal words DANGER, WARNING, and CAUTION are used regarding safety and other important instructions. The signal words and their meanings are defined as follows. Please understand their meanings clearly before reading this manual. Signal word Meaning Indicates an imminently hazardous situation which, if not avoided, will ADANGER result in death or serious injury. Indicates a potentially hazardous situation which, if not avoided, could AWARNING result in death or serious injury. Indicates a potentially hazardous situation which, if not avoided, may ACAUTION result in minor or moderate injury. CAUTION Indicates a potentially hazardous situation which, if not avoided, may result in property damage. 2. Meaning of Safety Symbols Symbol 1 Description Type-B applied part ir I * Type B when Type-B applied part is connected. The PCG sensor and pulse sensor that can be connected to this system are Type-B applied parts. Type-BF applied part fi cl * Type BF when Type-BF applied part is connected. All ultrasound transducers, ECG cables, and respiration sensors that can be connected to this system are Type-BF applied parts. “Attention” (Refer to the operation manual.) A! S-l
[Page 5] 3. Safety Precautions Please observe the following precautions to ensure the safety of service engineers as well as operators when using this system. &ANGER: D o no t use flammable gases such as anesthetics, or flammable liquids such I as ethanol, near this product, because there is danger of explosion. I ,&ARNING: 1. Information concerning the terminal to which the potential equalization conductor is to be connected ( ): To use another medical device in combination with this system, an equipotential wire for connecting to an equipotential bus must be supplied. For more information, contact your TOSHIBA representative. * Be sure to connect the potential-equalization lead wire before inserting the equipment power plug into the receptacle. Also, be sure to remove the equipment power plug from the receptacle before disconnecting the wire to avoid electrical shock. 2. Information concerning the functional earth terminal (6): Use of the functional earth terminal as protective earth is not allowed in Europe. The system should be connected according to local requirements. 3. Do not connect this system to outlets with the same circuit breakers and fuses that control current to devices such as life-support systems. If this system malfunctions and generates an overcurrent, or when there is an instantaneous current at power ON, the circuit breakers and fuses of the building’s supply circuit may be tripped. 4. Do not connect to the system transducers other than those specified by TOSHIBA, to prevent accidents such as fire. 5. Do not subject the transducers to knocks. Use of defective transducers may cause an electric shock. ACAUTION: 1. Malfunctions due to radio waves (1) Use of radio-wave-emitting devices in the proximity of this kind of medical electronic system may interfere with its operation. Do not bring or use devices which generate radio waves, such as cellular telephones, transceivers, and radio controlled toys, in the room where the system is installed. (2) If a user brings a device which generates radio waves near the system, they must be instructed to immediately turn OFF the device. This is necessary to ensure the proper operation of the system. s-2
[Page 6] A - CAUTION: 2. Be sure to install the system on a level floor and lock the casters. If this is not done, the system may move, injuring the service personnel. 3. Move the system forward or backward only. If the system is moved to the left or right, it may fall, causing injury. In addition, be sure to fix the moving sections such as the operating panel before moving the system. If such sections move, they may catch the hands or fingers of the persons moving the system, possibly causing injury. 4. To prevent electric shock, do not connect the peripheral units (video printer, VCR, etc.) to an external outlet. Peripheral units should be connected to the service outlet of the system or the optional isolation transformer. For the connection procedures, contact your TOSHIBA representative. 5. The service outlet of the main unit is intended solely for recommended peripheral units. Do not connect devices other than recommended peripheral units to the service outlet. Connecting devices other than those recommended may cause the consumption current to exceed the power capacity of the main unit, possibly resulting in malfunction. 6. Do not place any objects on top of the monitor. They may fall, causing injury. 7. Do not allow fluids such as water to contact the system or peripheral devices. Electric shock may result. 8. Do not handle the system with wet or moist hands. Electric shock may result. 9. The monitor is a heavy unit. Two persons are required when mounting or removing the monitor. If a worker mounts or removes the monitor alone, he/she may be injured. 10. Be sure to turn OFF the power breaker and disconnect the mains plug from the outlet before removing the covers of the main unit and performing internal wiring. 11. When discarding any part of this system, follow all applicable local regulations. 12. Wear protective gloves to ensure safety and prevent infection when performing unit replacement, board replacement, internal wiring, or cleaning. 13. Wait at least 30 seconds after turning OFF the power breaker and disconnecting the mains plug from the outlet before replacing the battery. Otherwise, an electric shock may result. CAUTION: I. When replacing the AC/DC tray of the SSA-770A, check the revision number of the AC/DC tray supplied for replacement and follow the appropriate procedure according to the revision number. 1 .I When the new AC/DC tray is BSM31-2013E (no revision) The AC/DC tray is not compatible and must not be used to replace the existing tray. s-3
[Page 7] CAUTION: 1.2 If AC/DC tray BSM31-2013E-G or BSM31-2013E-I is supplied (1) Wind the cables around the core as described below. (a) Wind the FAN/RELAY signal cables (two lines) between the AC/DC tray and transformer tray ONE time around the core (TWO turns). * When the cable is passed through the core the first time it is counted as one turn. Therefore, winding of the cable one time corresponds to TWO turns. (b) Wind the LED signal cable (which is connected to the top of the AC/DC tray) between the AC/DC tray and the transformer tray ONE time around the core (TWO turns). * When the cable is passed through the core the first time it is counted as one turn. Therefore, winding of the cable one time corresponds to TWO turns. (2) The metal sheet for EMC compliance should not be installed. 1) (b) Core for the LED signal cable (1) (a) Core for the FAN/RELAY Wind the cable ONE time around s-4
[Page 8] CAUTION: 1.3 When the new AC/DC tray is BSM31-2013E-H or -J or later (revision or later) (1) Wind the cables around the core as described below. Note the number of times each cable is to be wound around the core. (a) Wind the FAN/RELAY signal cables (two lines) between the AC/DC tray and transformer tray FOUR times around the core (FIVE turns). * When the cable is passed through the core the first time it is counted as one turn. Therefore, winding of the cable four times corresponds to five turns. (b) Wind the LED signal cable (which is connected to the top of the AC/DC tray) between the AC/DC tray and the transformer tray ONE time around the core (TWO turns). * When the cable is passed through the core the first time it is counted as one turn. Therefore, winding of the cable one time corresponds to TWO turns. (2) The metal sheet for EMC compliance should be installed. Press the metal sheet firmly against the rack so that the upper tab is flat and then secure the sheet using screws1 and 2. (2) Press the metal sheet for EMC compliance firmly against the rack so that this tab is flat and then secure the sheet using screws 1 and 2. /;1) (b) Core for the LED signal cable Wind the cable ONE time around the core (TWO turns). \I----:.:, (1) (a) Core for the FAN/RELAY Screw (M3*8 mm) Wind the cable FOUR times around the core (FIVE turns). Screw (M3*8 mm) S-5
[Page 9] CAUTION: 2. Voltage setting for replacing the transformer tray and TRANS-unit The switches used to select the output voltage according to the input voltage are installed on the transformer tray and the TRANS-unit. When the switches of the replacement units (transformer tray and TRANS-unit) are not set correctly, the power supply may not start up correctly, or a malfunction of the connection equipment may result. Be sure to check the settings. For the transformer tray 100 to 120 VAC “115” must be displayed on Sl and S3. The direction of S2 must be the same as that of Sl 230 VAC “230” must be displayed on Sl and S3. The direction of S2 must be the same as that of S3 For the TRANS-unit 100 to 120 VAC “115” must be displayed. “115” must be displayed. The direction must be the same as that of Sl and S2 230 VAC “230” must be displayed. “230” must be displayed. The direction must be the same as that of Sl and S2 S-6
[Page 10] CONTENTS Safety precautions ________________________________________------------------------------------------------------------------- S-1 1. INTRODUCTION ________________________________________-------------------------------------------------------------- 1-I 1.1 Outline ________________________________________------------------------------------------------------------------- 1-I 1.2 PWB Configuration of Each Unit ________________________________________---------------------------------- I-2 2. BLOCK DIAGRAM ________________________________________------------------------------------------------------------ 2-I 3. DISASSEMBLY AND REMOVAL PROCEDURES ________________________________________--------------------- 3-I 3.1 Name of Each Section ------__--------------~~~--------------~~~--------------~~--------------~~~------------ 3-l 3.2 Removing the Covers ________________________________________------------------------------------------------ 3-5 3.2.1 For the SSA-77OA-------------------------------------------------------------------------------- 3-5 3.2.2 For the SSA-7OOA-------------------------------------------------------------------------------- 3-7 3.3 Removing the Handle ________________________________________---------------------------------------------- 3-10 3.3.1 For the SSA-77OA------------------------------------------------------------------------------ 3-l 0 3.3.2 For the SSf+7OOA ________________________________________-------------------------------------- 3-l 1 3.4 Removing the Monitor Unit ________________________________________--------------------------------------- 3-12 3.4.1 Removing the monitor unit (for systems with the CRT monitor)---------------------- 3-12 3.4.2 Removing the monitor unit (for systems with the LCD monitor)---------------------- 3-14 3.4.3 Removing the LCD monitor arm ________________________________________--------------------- 3-15 3.4.4 Adjusting the LCD monitor arm vertical movement resistance ----------------------- 3-l 8 3.5 Removing the Operating Panel ________________________________________---------------------------------- 3-19 3.5.1 For the SSf+77OA ________________________________________-------------------------------------- 3-l 9 3.5.2 For the SSf+7OOA ________________________________________-------------------------------------- 3-23 3.6 Disassembling the Operating Panel ________________________________________---------------------------- 3-24 3.6.1 Removing the lower cover ________________________________________---------------------------- 3-24 3.6.2 Removing the upper panel ________________________________________--------------------------- 3-27 3.6.3 Removing the handle -----------___--------------~~~--------------~~--------------~~~---------- 3-28 3.6.4 Removing the trackball section ________________________________________---------------------- 3-29 3.6.5 Removing the PWBs (B) and (G) -------------___--------------~~--------------~~~---------- 3-30 3.6.6 Removing the keyboard----------------------------------------------------------------------- 3-31 -a-
[Page 11] 3.6.7 Removing the keyboard ________________________________________------------------------------- 3-32 3.6.8 Removing the LCD /.l,SSY ________________________________________---------------------------- 3-33 3.6.9 Removing the PWBs (C), (D), and (E) ________________________________________------------- 3-34 3.6.10 Removing the near/away slide lever ________________________________________--------------- 3-35 3.6.11 Removing the up/down slide lever ________________________________________------------------ 3-36 3.6.12 Mounting the switch cover ________________________________________---------------------------- 3-37 3.7 Removing the PWB Rack--------------------------------------------------------------------------------- 3-38 3.8 Removing the PWBs ________________________________________----------------------------------------------- 3-40 3.9 Removing the TI--------------------------------------------------------------------------------------------- 3-41 3.10 Removing the 10 -------------__--------------~~~--------------~~~--------------~~--------------~~~---------- 3-42 3.11 Removing the HDD Unit ________________________________________------------------------------------------- 3-43 3.12 Removing the Battery-------------------------------------------------------------------------------------- 3-44 3.13 Removing the AC-DC Unit ________________________________________---------------------------------------- 3-45 3.14 Removing the PSA and PSD ________________________________________------------------------------------- 3-48 3.15 Removing the Transformer ________________________________________--------------------------------------- 3-49 3.16 Removing the MO ________________________________________-------------------------------------------------- 3-50 3.17 Removing the Front panel ________________________________________---------------------------------------- 3-51 3.17.1 For the SSf+77OA ________________________________________-------------------------------------- 3-51 3.17.2 For the SSf!,-7OOA ________________________________________-------------------------------------- 3-54 3.18 Removing the Rear panel ________________________________________----------------------------------------- 3-55 3.19 Removing the PHYSlO Unit ________________________________________-------------------------------------- 3-56 3.19.1 For the SSf+77()A ________________________________________-------------------------------------- 3-56 3.19.2 For the SSf+7OOA ________________________________________-------------------------------------- 3-57 3.20 Removing the Caster Units ________________________________________--------------------------------------- 3-58 3.20.1 For the SSA-77OA------------------------------------------------------------------------------ S-58 3.20.2 For the SSf!,-7()()p, ________________________________________-------------------------------------- 3-61 3.21 Cleaning the Air Filters -----__--------------~~~--------------~~~--------------~~--------------~~~---------- 3-62 3.22 Cleaning the panel ----------__--------------~~~--------------~~~--------------~~--------------~~~---------- 3-63 -b-
[Page 12] 3.22.1 Cleaning of key tops ________________________________________----------------------------------- 3-63 3.22.2 Cleaning procedures for parts other than key tops ------------------------------------- 3-63 3.23 Adjusting the Operating Panel Locking Wire ________________________________________----------------- 3-64 3.23.1 Adjusting the up/down slide wire ________________________________________-------------------- 3-64 3.23.2 Adjusting the near/away slide wire ________________________________________----------------- 3-66 3.24 Replacing the Operating Panel Locking Wire ________________________________________----------------- 3-67 3.24.1 Replacing the up/down slide wire----------------------------------------------------------- 3-67 3.24.2 Replacing the near/away slide wire ________________________________________----------------- 3-72 3.25 Removing the CD-R Unit---------------------------------------------------------------------------------- 3-80 3.26 Cleaning the MO Drive -----__--------------~~~--------------~~~--------------~~--------------~~~---------- 3-82 3.27 Cleaning the palm Switch ________________________________________----------------------------------------- 3-83 3.27.1 Cleaning procedures for the old-type palm switch -------------------------------------- 3-83 3.27.2 Cleaning procedures for the new-type palm switch------------------------------------- 3-85 4. OpER/T!,TlON OF EACH PWB ________________________________________---------------------------------------------- 4-l 4.1 Front-End Unit ________________________________________--------------------------------------------------------- 4-l 4.1.1 TI ________________________________________------------------------------------------------------------- 4-5 4.1.2 TR ________________________________________------------------------------------------------------------ 4-g 4.1.3 RCB ________________________________________-------------------------------------------------------- 4-13 4.1.4 CB ________________________________________---------------------------------------------------------- 4-17 4.2 BE Unit ________________________________________---------------------------------------------------------------- J-23 4.2.1 BE ________________________________________---------------------------------------------------------- J-24 4.2.2 VI (Video Interface) ________________________________________------------------------------------ 4-35 4.2.3 RM (Realtime Manager) ________________________________________------------------------------ 4-41 4.2.4 1QDA.S (Ip) ________________________________________----------------------------------------------- 4-44 4.3 System-R&t& Boar& ----__--------------____________________--------------~~--------------~~~---------- 4-47 4.3.1 I/O -------------------------------------------------------------------------------------------------- 4-47 4.3.2 Front panel --------__--------------~~~--------------~~~--------------~~--------------~~~---------- 4-50 4.3.3 Rear panel --------__--------------~~~--------------~~~--------------~~--------------~~~---------- 4-50 -c-
[Page 13] 5. OPERATING PANEL ________________________________________--------------------------------------------------------- 5-I 6. POWER SUPPLY UNIT ________________________________________------------------------------------------------------ 6-I 6.1 power Supply Unit Configuration ________________________________________---------------------------------- 6-I 6.2 Block Diagram of the power Supply Unit ________________________________________------------------------ 6-I 6.3 Unit Specifications ________________________________________---------------------------------------------------- 6-2 6.4 Transformer Tray ________________________________________----------------------------------------------------- 6-3 6.5 AC/DC Tray --___--------------_____________________------------~~~--------------~~--------------~~~------------ 6-6 6.6 pSA/pSD Unit ________________________________________--------------------------------------------------------- 6-7 6.7 LEDs for Error Display ------__--------------~~~--------------~~~--------------~~--------------~~~---------- 6-l 2 7. MONITOR -------------___--------------~~--------------~~~--------------~~~--------------~~--------------~~~------------ 7-l 7.1 CRT Monitor ________________________________________----------------------------------------------------------- 7-l 7.2 LCD Monitor---------------------------------------------------------------------------------------------------- 7-5 8. PHySlO MODULE ________________________________________------------------------------------------------------------ 8-I 9. CONNECTION DRAWING FOR UNITS ________________________________________---------------------------------- 9-l 10, SOFTWARE ORG/T!,NlZATlON ________________________________________------------------------------------------- 10-l IO.1 Overall Folder Configuration ________________________________________------------------------------------- 1 O-1 IO.2 Version Information ________________________________________------------------------------------------------- IO-2 IO.3 Error Handling ________________________________________------------------------------------------------------- 1 O-3 10.3.1 Error log operation ________________________________________------------------------------------- 1 O-3 10.3.2 Error messages ________________________________________----------------------------------------- 1 O-4 10.3.3 Error levels (severity) ________________________________________---------------------------------- 1 O-4 10.3.4 Error log operation (facility) ________________________________________--------------------------- 1 O-5 11, SETTING THE PRESET MENU ________________________________________----------------------------------------- 11-I 11 ,I Outline of Each preset ________________________________________--------------------------------------------- 11-I 1 I.2 Setting and Changing presets ------------___--------------~~~--------------~~--------------~~~---------- 1 I-5 II.3 preset Editors ________________________________________------------------------------------------------------- 1 I-10 1 I.4 Description of Preset Parameters ----------------------------------------------------------------------l l-l 4 12, BACK1 NG Up THE USERS DATA------------------------------------------------------------------------------- 12-I -d-
[Page 14] 12.1 Backing Up the preset Data ________________________________________-------------------------------------- 12-I 12.1.1 Backup ________________________________________---------------------------------------------------- 12-I 12.1 ,‘J R&oration ________________________________________----------------------------------------------- 12-I 12.2 Backing Up the Image Data ________________________________________-------------------------------------- 12-I 12.2.1 Backup ________________________________________---------------------------------------------------- 12-I 12.2.2 R&oration ________________________________________----------------------------------------------- 12-I 13, N EJWORK CONNECTION --------__--------------~~~--------------~~~--------------~~--------------~~~---------- 13-l 13.1 Changing the PC Name ________________________________________------------------------------------------- 13-I 14, U PGRADl NG THE SOFTWARE --__--------------______________________------------~~--------------~~~---------- 14-l 15, SELECTION OF THE PANEL SWITCH LAYOUT -------------___--------------~~--------------~~~---------- 15-I 15.1 Standard panel Switch Layout ________________________________________----------------------------------- 15-I 15.2 Changing the panel Configuration File----------------------------------------------------------------- 15-I 15.3 Panel Switch Layout and Switch Functions After Configuration Change----------------------- 15-2 15.4 Attaching the Label ________________________________________------------------------------------------------- 15-7 16, DATE/TIME SETTING CHANGE ________________________________________---------------------------------------- 16-I 16.1 To Change the System Time Only ________________________________________------------------------------ 16-I 16.2 To Change the System Date and Time ________________________________________------------------------ 16-I 17, 3D POSITION SENSOR ________________________________________--------------------------------------------------- 17-I 17.1 3D position Sensor Unit ________________________________________------------------------------------------- 17-I 17.2 Magnetic Field Dist&ion Check TooI ________________________________________-------------------------- 17-5 -e-
[Page 15] 1. INTRODUCTION 1.1 Outline The SSA-700A/SSA-770A (hereinafter referred to as Aplio) functions in the 2D, M, FFT Doppler, and color blood-flow modes. The system is classified into the following five main systems depending on the combined units. (1) Front-end unit (2) Back-end unit (3) System-related unit (4) Power supply unit (5) Monitor l-l
[Page 16] 1.2 PWB Configuration of Each Unit Table l-l PWB configuration of each unit of the Aplio Unit PWB name Drawing No. YW No. Remarks Front-end unit TI PM30-30382 YWM 1030 The TI-ASSY (including PM30-29950) can be replaced during field service. TRO PM30-32040 YWM 1069 PM30-32263 YWM1116 PM30-32469 YWMI 171 Compatible with PM30-32263 TRI PM30-32040 YWM 1069 PM30-32263 YWMIIIG PM30-32469 YWMI 171 Compatible with PM30-32263 TR2 PM30-32040 YWM 1069 PM30-32263 YWM1116 PM30-32469 YWMI 171 Compatible with PM30-32263 RCB PM30-32038 YWM 1098 PM30-32088 YWMI 109 Compatible with PM30-32038 CB PM30-30297 YWM0857 Option Back-end unit BE PM30-30385 YWM 1058 VI PM30-30386 YWM 1059 PM30-32039 YWMI 101 This PWB is compatible with PM30-30386. RM PM30-29855 BSM31-2996E IQDAS PM30-32120 YWMI 105 Option System related unit BP PM30-30381 YWM 1057 PM30-32036 YWMI 100 This PWB is compatible with PM30-30381. IO PM30-32037 YWM 1099 The IO-ASSY (including PM30-30120) can be replaced during field service. PM30-32210 YMM1112 The service unit is the same as above. Compatible with PM30-32037 FRONT PM30-30357 YWM1081 The F-U-PANEL-ASSY PANEL (including PM30-30010) can be replaced during field service. REAR PM30-30358 YWM0985 The REAR-PANEL-ASSY PANEL (including PM30-29990) can be replaced during field service. l-2
[Page 17] Block Diagram - - - Goods o n market TI(Transducer IF) BE (Back End) RXBF DRU EP kl SC 1 _ .. Physio BP (Backplane) DC48 V Ma,n TRS PSA 100,230 v (select) 5”. 3 3 ” orl3V DC , AC-OUT, Monitor . AC-OUTZ. Color Printer b (/GOUT3 “CR) . (AC-OUT4 B&W Printer) Skin cage rsoGiia, - - 1 (Option) ‘I:
[Page 18] 3. DISASSEMBLY AND REMOVAL PROCEDURES 3.1 Name of Each Section (1) SSA-770A (systems with the CRT monitor)
[Page 19] SSA-770A (systems with the LCD monitor) 3-2
[Page 20] (3) SSA-700A (systems with the CRT monitor)
[Page 21] (4) SSA-700A (systems with the LCD monitor) 3-4
[Page 22] 3.2 Removing the Covers 3.2.1 For the SSA-770A (b) Remove the two M4 screws <A> to remove the left side cover. * The cover can be removed by pulling the bottom of the cover toward you (2) Removing the right side cover Remove the right side cover in the same manner as the left side cover. (3) Removing the left side cover for peripheral units It is necessary to remove the left side cover in advance. Remove the two M4 screws <B> to remove the left side cover for the peripheral units. * The cover can be removed by lifting it up and then pulling it toward you. (4) Removing the right side cover for the peripheral units It is necessary to remove the right side cover in advance Remove the two M4 screws CC> to remove the right side cover for the peripheral units. * The cover can be removed by lifting it up and then pulling it toward you. (5) Removing the front cover When the transducers are connected, remove them. It is necessary to remove the right and left side covers in advance. (a) When the blank cover is installed, loosen the M3 screw CD> to remove the blank cover with the printer base installed. (b) Remove the four M4 screws <E> to remove the front cover. (6) Removing the rear cover for the power supply It is necessary to remove the right and left side covers in advance. Remove the four M4 screws cF> to remove the rear cover for the power supply. (7) Removing the base front cover It is necessary to remove the right and left side covers and the front cover in advance. (a) Remove the two M4 screws cG> (b) Loosen the two M4 screws <H> to remove the base front cover 3-5
[Page 23] Removing the base rear cover It is necessary to remove the right and left side covers and the rear cover for the power supply in advance. Loosen the two M4 screws cl> to remove the base rear cover. 3-6
[Page 24] 3.2.2 For the SSA-700A (b) Remove the two M4 screws <A> to remove the left side cover. * The cover can be removed by pulling the bottom of the cover toward you (2) Removing the right side cover Remove the right side cover in the same manner as the left side cover. (3) Removing the front cover When the transducers are connected. remove them. It is necessary to remove the right and left side covers in advance. (a) When the blank cover is installed, loosen the M3 screw cB> to remove the blank cover with the printer base installed. (b) Remove the four M4 screws CC> to remove the front cover (4) Removing the rear cover for the power supply It is necessary to remove the right and left side covers in advance. Remove the four M4 screws CD> to remove the rear cover for the power supply. (5) Removing the base front cover It is necessary to remove the right and left side covers and the front cover in advance. (a) Remove the two M4 screws cE>. (b) Loosen the two M4 screws cF> to remove the base front cover. (6) Removing the base rear cover It is necessary to remove the right and left side covers and the rear cover for the power supply in advance. Loosen the two M4 screws cG> to remove the base rear cover. (7) Removing the peripheral unit rack Remove the six M4 screws <H> to remove the rack. 3-7
[Page 25] (8) Removing the handle covers (a) Remove the four M4 screws <I> and the M4 screw <J> to remove the peripheral unit rack retainer. (b) Remove the four M4 screws <K> to remove the handle cover (A). (c) Remove the four M4 screws CL> to remove the handle cover (B). (9) Removing the handle Remove the two M6 screws CM> and loosen the two M4 screws cN> to remove the handle. (10) Removing the top cover Remove the two M4 screws CO=- to remove the top cover. (11) Removing the top-left side cover Remove the M4 screws cP> and cQ> (1 each) to remove the top-left side cover. (12) Removing the top-right side cover Remove the top-right side cover in the same manner as the top-left side cover. (13) Removing the cover for the peripheral units Remove the M4 screws CR> and cS> (2 each) to remove the cover for the peripheral units. (14) Removing the sliding cover Remove the two M4 screws CT> to remove the sliding cover. 3-8
[Page 26] 1~ Handle cover <B> unit rack Handle cover <A> : rack retainer Base rear cover Top-right side cover I I ‘.,I / Screw cQ> >??I Base front cover cover Rubber cap p< Screw -=P> -Screw <A> -Screw <t>
[Page 27] 3.3 Removing the Handle 3.3.1 For the SSA-770A (1) Removing the handle (a) Remove the caps <A> and <B> set in the handle and the rear cover. (b) Remove (or loosen) the two M6 bolts <A>. (c) Remove the two M6 bolts <B> to remove the handle. Cap <B> 7 Bolt <B> - (6B6-12SDM) + ’ Haidle Bolt <Ii> (6B6-12SDM) 3-10
[Page 28] 3.3.2 For the SSA-700A (1) Remove the six M4 screws <A> that secure the middle shelf in order to remove it. (2) Remove the two M4 screws <B> to remove the reinforcement rib. (3) Remove the four M4 screws CC> to remove the middle shelf retaining plate. (4) Remove the four M4 screws CD> to remove the handle cover (front). (5) Remove the four M4 screws cE> to remove the handle cover (rear). (6) Remove the two M6 screws cF> to remove the handle. Screw <A> Middle shelf
[Page 29] 3.4 Removing the Monitor Unit hAUTION: T wo persons are required when removing the monitor unit. If a worker removes the monitor alone, he/she may be injured. 3.4.1 Removing the monitor unit (for systems with the CRT monitor) (1) Removing the monitor rear cover Loosen the two screws CA> and four screws cB> that secure the monitor rear cover to remove the cover. (2) Removing the support rear cover The support rear cover engages with the support front cover. Disengage the upper part of the cover and remove the cover by pulling it backward. (3) Removing the cables (refer to the next page) (a) Loosen the two screws CC> to remove the pinch shield cover. (b) Remove the two screws CD> to remove the clamps. (c) Remove the signal, audio, and power cables that are connected to the monitor unit. (4) Removing the monitor unit from the system Remove the four M6 bolts <E> to remove the monitor unit with the boss installed. - Screw <A> (two) Support rear cover 3-12
[Page 30] No. 2D730-148E*0 Audio cable Screw CC> Pinch shield cover (for power cable) Signal cable Power cable Clamp (for signal and audio cables) (5) Removing the monitor unit from the boss Remove the two M6 flat head screws cE> to pull the monitor unit from the boss. - Screw cE> (PS6-1 OSCR) 3-13
[Page 31] No. 2D730-148E”0 3.4.2 Removing the monitor unit (for systems with the LCD monitor) (1) Remove the screw <A> and then remove the connector cover. (2) Disconnect the power cable, DVI cable, and audio cable. (3) Attach the connector cover and secure it with the screw <A>. * The screw <A> is a self-tapping screw. Special care is required when tightening it. (4) Remove the four caps and the four M4 screws cB> to remove the speakers (one each on the right and left of the monitor). (5) Remove the two M4 screws ~0. (6) Loosen the two M4 screws CD> and then remove the LCD monitor from the arm. Power cable -/5 Screw CD> (SBPPCI OSCR) -Connector cover LCD monitor Screw <A> opaa~er (one eacn on the right and left- of the monitor) (Velcro fastener is provided on the monitor.) (SBPPClPSCRM\ 0 \- -LCD monitor arm 3-14
[Page 32] 3.4.3 Removing the LCD monitor arm (1) Remove the two caps and the two M3 screws <A> to remove arm cover I. (2) Remove the two M3 screws <B> to remove arm cover 2. (3) Remove the two M3 screws CC> to remove arm base cover I. (4) Raise the stopper and remove the M3 screw CD> and the M3 screw cE>. Arm cover 1 Arm cover 2 IL Arm base cover 1 Stopper *Screw +,‘,,, -Screw cc;=- (PS3-8SCR) 3-15
[Page 33] (5) Rotate the arm biaxial rotation column in the direction indicated by the arrow in the figure below. Lift arm base cover 2 to separate it from the stopper. Arm base cover 2 I Biaxial rotation column (6) Remove the two M3 screws <F> and the clamps cG> and <Hz. (7) Remove the three M4 screws cl> and the clamps cJ>, cK>, and CL>. Screw cF> P3-8SCR) Clamp cG> I Yamp cH’ Screw <I> (SBPP4-1 (8) Remove the cables one by one from the LCD monitor arm to the base 3-16
[Page 34] No. 2D730-148E*0 (9) Push in the neck of the support rear cover to disengage it from the support front cover and remove it. (10) Remove the two M4 screws <A> to remove the support front cover. (11) Remove the four M6 bolts <B> to remove the LCD monitor arm. (12) Lift the LCD monitor arm and remove the three monitor cables from the arm Screw ==:A> (SBPP4-?2SCR) Support front cover- 3perating panel 1 -Support rear cover 3-17
[Page 35] 3.4.4 Adjusting the LCD monitor arm vertical movement resistance (1) Remove the two caps and the two M3 screws <A> to remove arm cover I. (2) Remove the two M3 screws <B> to remove arm cover 2. (PS3-8SCR) - Arm cover 2 (3) Using the M6 hex key, rotate the hexagonal socket head bolt to adjust the vertical movement resistance of the LCD monitor arm. Reducing the LCD monitor vertical Increasing the LCD monitor movement resistance vertical movement resistance 1 I Id bolt 3-18
[Page 36] 3.5 Removing the Operating Panel 3.5.1 For the SSA-770A The operating panel must be raised using the up/down slide lever in advance. (1) Removing the handle Refer to subsection 3.3. (2) Removing the monitor unit Refer to subsection 3.4. (3) Removing the right side cover (b) Remove the two M4 screws CA> to remove the cover. (4) Removing the left side cover (b) Remove the two M4 screws cB> to remove the cover. (5) Removing the right side cover for the peripheral units Remove the two M4 screws ~0 to remove the cover. * The cover can be removed by lifting it and pulling it toward you. (6) Removing the left side cover for the peripheral units Remove the two M4 screws CD> to remove the cover. * The cover can be removed by lifting it and pulling it toward you. (7) Removing the transducer cable arm Remove the two M4 bolts cE> to remove the transducer cable arm. (8) Removing the transducer cable hook Remove the two M4 bolts cF> to remove the transducer cable hook. (9) Removing the top cover (a) Remove the four caps. (b) Remove the four M4 screws <G> to remove the cover. (10) Removing the support front cover Remove the two M4 screws <H> to remove the cover. 3-19
[Page 37] No. 2D730-148E*0 0 Cap Support front cover 0 Screw <G=- \ (SBPP4-12SCR) Screw cH> (SBPP4-12SCR1_ T Left side Transducer cable arm (SB PP4-12SCR) ‘l., TRight side cover ‘\ ‘\ ‘\ -\ ‘1 Screw CA> (1 I) Removing the bellows Remove the two M4 screws <I>. (12) Removing the front/rear sliding wire (a) Remove the M3 screw cJ> to remove the retaining clamp for the front/rear sliding wire. (b) Loosen the screw cK> and rotate the wire retaining bracket to remove the end of the wire. (c) Rotate the wire to remove it from the up/down mechanism. 3-20
[Page 38] (13) Removing the up/down sliding wire (a) Remove the end of the wire from the lever. (b) Rotate the wire to remove it from the up/down mechanism. ,,----, Front/rear sliding wire Screw Cl> (SBPF ‘4-l 2SCR) Up/down sliding wire 1 Up/down sliding mechanism Front/rear sliding wire Wire retaining bracket 3-21
[Page 39] (14) Removing the cables Remove the connected cables and the M3 screw CL> to remove the USB fitting clamp. Screw CL> (SBPP3-IOSCR) Clamp (15) Remove the operating panel from the system (a) Remove the four M6 bolts CM>. (b) Loosen the two M6 screws <N>. Rotate the operating panel clockwise and lift it to remove it. 3-22
[Page 40] 3.5.2 For the SSA-700A (1) Removing the handle Refer to subsection 3.3.1. (2) Removing the monitor unit Refer to subsection 3.4. (3) Removing the right side cover Same as for the SSA-770A. (4) Removing the left side cover Same as for the SSA-770A. (5) Remove the screw CA> to remove the cover cl>. (6) Remove the screws cB> and CC> to remove the covers <2> and <3> (7) Remove the screw CD> to remove the cover <4>. (8) Remove the screw cE> to remove the cover <!?I> I Covercl> Cover <5> Screw cE> 3-23
[Page 41] 3.6 Disassembling the Operating Panel 3.6.1 Removing the lower cover (1) Remove the transducer holders (six), jelly holders (two), and knobs (four small and one large). Jelly holder (small) (2) Remove the four screws <A> from the right and left of the holder mounting base secured on the back of the panel. Screws CA> - Screws CA> 3-24
[Page 42] (3) Loosen the four screws cB> at the right and left of the holder mounting base. * The screws B include springs and cannot be removed. Loosen the screws so that they turn freely and protrude from the base. 3-25
[Page 43] (4) Hold the cover at the jelly holder mounting holes, apply a force so as to pull the cover toward you, and lift the cover to remove it. &AUTION: Th e cover is mounted by engaging the indentations in its front part with the tabs in the handle section. Therefore, do not attempt to remove the cover by lifting it from the front. Doing so may damage the cover. Pull the cover 3-26
[Page 44] 3.6.2 Removing the upper panel (1) Remove the two screws <A> and the three screws cB> and shift the panel in the direction indicated by the arrow. Screws CA> (2) Remove the two connectors 3-27
[Page 45] No. 2D730-148E*0 3.6.3 Removing the handle (1) Remove the five screws <A> on the back of the panel and remove the handle. Screws <A> Handle 3-28
[Page 46] 3.6.4 Removing the trackball section (1) Remove the connector connected to the trackball section and remove the trackball section. Trackball section Connector 3-29
[Page 47] 3.6.5 Removing the PWBs (B) and (G) (1) Remove the three connectors connected to the PWB (B). (2) Remove the twelve screws <A> and remove the PWB (B). (3) Remove the two screws cB> and remove the PWB (G). Connector PWB (B) Screw <A> / v Connectors (indica [ted by a * mark) Pm 0 Screws <B> 3-30
[Page 48] 3.6.6 Removing the keyboard (1) Remove the six flat head screws CA> and shift the keyboard. (2) Remove the connector. Screws CA> Screws <A>
[Page 49] 3.6.7 Removing the keyboard (1) Remove the five screws CA> and the two screws <B> and remove the keyboard plate B. (2) Remove the connectors from the FPC. (3) Remove the six screws CC> and remove the keyboard cover. Keyboard cover Screws CA> Screws CC> (4) Remove the keyboard. Keyboard 3-32
[Page 50] 3.6.8 Removing the LCD ASSY (1) Remove the USB rubber cap. (2) Remove the two connectors. (3) Remove the two screws <A> and remove the L-bracket L and the L-bracket R. (4) Remove the five screws <B> and remove the LCD ASSY. LCD ASSY USB rubber cap Connector Screws cB> 3-33
[Page 51] 3.6.9 Removing the PWBs (C), (D), and (E) (1) Remove the six knobs CA>, <B>, and CC> and the eight STC slide controls CD>. Kno (2) Remove the two connectors. (3) Remove the five screws <A> and remove the PWB (C). (4) Remove the eleven screws <B> and remove the PWB (D). (5) Remove the two screws CC> and remove the PWB (E). Screws CC> pwy CC) A Screws <B> Screws <A> Screws <A> Screis <B> Conn\ector 3-34
[Page 52] 3.6.10 Removing the near/away slide lever (1) Remove the two screws CA> and remove the clamp. (2) Remove the two screws cB> and remove the bracket. (3) Remove the two screws CC> and remove the lever. (4) Remove the two screws CD> and remove the wire. Screws CA> *acket 3-35
[Page 53] No. 2D730-148E”0 3.6.11 Removing the up/down slide lever * Only for the SSA-770A (1) Remove the screw <A> and remove the clamp. (2) Remove the two screws <B> and remove the up/down slide lever. Screw CA> - 3-36
[Page 54] 3.6. 12 Mounting the switch cover (1) Mount the cover (starting with the front section) by reversing the steps for removal. Engage the tabs in the handle section with the indentations in the cover cl> and then mount the cover c2>. (2) Secure the four screws cB> while pressing the cover from above. (3) Secure the four screws <A> at right and left of the holder mounting base on the back of the panel. (4) Install the four small rotary encoder knobs and the large rotary encoder knob. Tabs
[Page 55] 3.7 Removing the PWB Rack 1 ACAUTION: Two persons are required when removing the PWB rack. Confirm that the transducers have been removed. (b) Remove the two M4 screws <A> to remove the cover * The cover can be removed by pulling its bottom toward you (2) Removing the AC-DC metal retainer (a) Loosen the two M4 screws cB> to remove the AC-DC metal retainer. AC-DC metal retainer Screw <A> p’ 3-38
[Page 56] (3) Removing the right side cover (b) Remove the two M4 screws CC> to remove the cover. (4) Removing the right side cover for the peripheral units Remove the two M4 screws CD> to remove the cover. * The cover can be removed by lifting it and pulling it toward you. (5) Removing the PWB rack (a) Remove all the cables connected to the PWB rack. (b) Remove the two M4 screws cE> and extract the PWB rack. Right side cover for the peripheral units Screw CC>- (SBPPC12SCR) 3-39
[Page 57] 3.8 Removing the PWBs (2) Removing the PWBs (a) Loosen the four screws cB> to remove the PWB rack shield plate. (b) Release the card ejectors at the upper and lower ends of the PWB and extract the PWB. cover (SBPPC12SCR) PWB rack
[Page 58] 3.9 Removing the TI Confirm that the transducers have been removed. (2) Removing the TI (a) Remove the cable for the pencil transducer (PP) connected to the TI. (b) Loosen the two screws cB> to extract the TI. cover 3-41
[Page 59] 3.10 Removing the IO (2) Removing the IO (a) Remove all the cables connected to the IO (b) Loosen the two screws <B> to extract the IO. Screw cB> (two) Right side cover 3-42
[Page 60] 3.11 Removing the HDD Unit (a) Remove the two rubber caps <A> at the bottom of the cover. (2) Removing the right side cover (a) Remove the two rubber caps cB> at the bottom of the cover. (b) Remove the two M4 screws cB> to remove the cover. (3) Remove the power supply section rear cover Remove the four M4 screws CC> to remove the cover. (4) Removing the HDD unit Remove the four M4 screws CD> to remove the HDD unit Right Screw <A=- Power supply section rear cover Screw CC> 3-43
[Page 61] 3.12 Removing the Battery (2) Removing the battery (a) Remove the battery connector that is connected to the AC-DC unit. (b) Remove the battery of the AC-DC unit. 3-44
[Page 62] 3.13 Removing the AC-DC Unit (2) Removing the AC-DC unit (a) Remove all the cables connected to the AC-DC unit. (b) If the metal sheet for EMC compliance is installed, remove it by removing screws cB>. (c) Remove the four M4 screws CC> to remove the AC-DC unit. Metal sheet for EMC compliance (SBPP3-8SCR) Screw <A> J Cap 3-45
[Page 63] CAUTION: When replacing the AC/DC tray of the SSA-770A, check the revision number of the AC/DC tray supplied for replacement and follow the appropriate procedure according to the revision number. 1. When the new AC/DC tray is BSM31-2013E (no revision) The AC/DC tray is not compatible and must not be used to replace the existing tray. 2. If AC/DC tray BSM31-2013E-G or BSM31-2013E-I is supplied (1) Wind the cables around the core as described below. (a) Wind the FAN/RELAY signal cables (two lines) between the AC/DC tray and transformer tray ONE time around the core (TWO turns). * When the cable is passed through the core the first time it is counted as one turn. Therefore, winding of the cable one time corresponds to TWO turns. (b) Wind the LED signal cable (which is connected to the top of the AC/DC tray) between the AC/DC tray and the transformer tray ONE time around the core (TWO turns). * When the cable is passed through the core the first time it is counted as one turn. Therefore, winding of the cable one time corresponds to TWO turns. (2) The metal sheet for EMC compliance should not be installed. 1) (b) Core for the LED signal cable Wind the cable ONE time around the core (TWO turns). (1) (a) Core for the FAN/RELAY 0 signal cable Wind the cable ONE time around the core (TWO turns). 3-46
[Page 64] CAUTION: 3. When the new AC/DC tray is BSM31-2013E-H or-J or later (revision or later) (1) Wind the cables around the core as described below. Note the number of times each cable is to be wound around the core. (a) Wind the FAN/RELAY signal cables (two lines) between the AC/DC tray and transformer tray FOUR times around the core (FIVE turns). * When the cable is passed through the core the first time it is counted as one turn. Therefore, winding of the cable four times corresponds to five turns. (b) Wind the LED signal cable (which is connected to the top of the AC/DC tray) between the AC/DC tray and the transformer tray ONE time around the core (TWO turns). * When the cable is passed through the core the first time it is counted as one turn. Therefore, winding of the cable one time corresponds to TWO turns. (2) The metal sheet for EMC compliance should be installed. Press the metal sheet firmly against the rack so that the upper tab is flat and then secure the sheet using screws1 and 2. (2) Press the metal sheet for EMC compliance firmly against the rack so that this tab is flat and then secure the sheet using screws 1 and 2. /;1) (b) Core for the LED signal cable Wind the cable ONE time around the core (TWO turns). \I----:.:, (1) (a) Core for the FAN/RELAY Screw (M3*8 mm) Wind the cable FOUR times around the core (FIVE turns). Screw (M3*8 mm) 3-47
[Page 65] 3.14 Removing the PSA and PSD (1) Removing the left side cover and AC-DC unit Refer to subsection 3.13. c-3 Removing the FAN-BOX (a) Remove the cable of the FAN-BOX (P95). (b) Loosen the three M3 FAN-BOX mounting screws CA> that secure the PSA and PSD to remove the FAN-BOX. (3) Removing the PSA Loosen the three screws <B> and two screws CC> that secure the PWB rack, hold the handles at the upper and lower ends, and extract the PSA. (4) Removing the PSD Loosen the two screws CD> and the screw cE> that secure the PWB rack, hold the handles at the upper and lower ends, and extract the PSD. Sc;kw cA> (SBPP3-IOSCR) Screw <E> 3-48
[Page 66] 3.15 Removing the Transformer (a) Remove the two rubber caps <A> at the bottom of the cover. (2) Removing the right side cover (a) Remove the two rubber caps cB> at the bottom of the cover. (b) Remove the two M4 screws cB> to remove the cover. (3) Removing the power supply section rear cover Remove the four M4 screws CC> to remove the cover. (4) Removing the base rear cover Loosen the two M4 screws <D> to remove the cover. (5) Removing the transformer (a) Remove all the cables connected to the transformer. (b) Remove the four M4 screws cE> to extract the transformer. Left side section rear c Base rear Screw CC> Screw <A> -’ Cap CA> Screw CD> (SBPP4-12SCR) (SBPP4-12SCR) 3-49
[Page 67] 3.16 Removing the MO (1) Removing the cables (a) Remove the two rubber caps at the bottom of the left side cover. (b) Remove the two M4 screws <A> to remove the left side cover. (c) Remove the four M3 screws <B> to remove the internal cover. (d) Remove the internal SCSI cable (J505) and power cable (J502 or J503) that are connected to the MO. (2) Removing the MO (a) Remove the blank cover. (b) Remove the two M3 screws CC> to extract the MO :R) Cap I Screw CA> 3-50
[Page 68] 3.17 Removing the Front Panel 3.17.1 For the SSA-770A (a) Remove the two rubber caps CA> at the bottom of the cover. (2) Removing the right side cover (a) Remove the two rubber caps cB> at the bottom of the cover. (b) Remove the two M4 screws cB> to remove the cover. (3) Removing the left side cover for the peripheral units Remove the two M4 screws CC> to remove the cover. * The cover can be removed by lifting it and pulling it toward you. (4) Removing the left side cover for the peripheral units Remove the two M4 screws CD> to remove the cover. * The cover can be removed by lifting it and pulling it toward you. (5) Removing the tray Loosen the four M4 screws <E> to remove the tray. (6) Removing the top front cover Remove the four M4 screws cF> to remove the cover. 3-51
[Page 69] No. 2D730-148E”0 Right side cover for the peripheral units Left side cover for the peripheral units 7 (SBPPCI 2SCR) - Screw <F> Screw <E> (SBPPCI 2SCR) Cap L Screw <A> (SBPPCI 2SCR) 3-52
[Page 70] (7) Removing the cables connected to the MO Remove the four M3 screws cG> to remove the internal cover. Remove the internal SCSI cable (J505) and power cable (J502 or J503) that are connected to the MO. Remove the cables (J67, J68, J400, and J132) connected to the front panel. 63) Removing the front panel Remove the four M4 screws cH> to remove the front panel. .MO Internal cover - Screw <G> (PS3-GSCR) 3-53
[Page 71] 3.17.2 For the SSA-700A Same as for the SSA-770A. (2) Removing the left side cover Same as for the SSA-770A. (3) Removing the tray Same as for the SSA-770A. (4) Remove the screw <A> to remove the middle shelf. (5) Remove the screw cB> to remove the cover cl>. (6) Remove the screws CC> and CD> to remove the covers <2> and c3>. (7) Remove the screw cE> to remove the front cover. - Screw CA> 1 Covyrcl> Front cover Trpy - Screw cE> Screw CC> y, * : II $- Cover <2> 3-54
[Page 72] 3.18 Removing the Rear Panel (a) Remove the two rubber caps CA> at the bottom of the cover. (2) Removing the right side cover (a) Remove the two rubber caps cB> at the bottom of the cover. (b) Remove the two M4 screws cB> to remove the cover. (3) Removing the power supply section rear cover Remove the four M4 screws CC> to remove the cover. (4) Removing the cables Remove all the cables connected to the rear panel. (5) Removing the rear panel Remove the two M3 screws CD> to remove the rear panel. Cap CA> section rear cover L Screw CD> \ Screw CC> (SBPP3-1 OSCR) Screw CA> (SBPP4-12SCR) (SBPP4-12SCR) 3-55
[Page 73] 3.19 Removing the PHYSIO Unit 3.19.1 For the SSA-770A (2) Removing the right side cover for the peripheral units Remove the two M4 screws cB> to remove the cover. * The cover can be removed by lifting it and pulling it toward you. (3) Removing the metal fitting Remove the three M3 screws CC> to remove the metal fitting. (4) Removing the PHYSIO unit (a) Remove all the cables connected to the PHYSIO unit. (b) Remove the M4 screw CD> and M4 screw cE> to remove the PHYSIO unit. Screw CD> 1 (PTR4-IOSBNI) Metal fitting \ Right side cover for the peripheral units Right side cover -Screw cB> Screw <A> Cap 3-56
[Page 74] 3.19.2 For the SSA-700A Same as for the SSA-770A. (2) Removing the PHYSIO unit (a) Disconnect all cables that are connected to the PHYSIO unit. (b) Remove the screw CA> to remove the middle shelf. (c) Remove the screw cB> at the PHYSIO unit to remove the PHYSIO front cover, (d) Remove the PHYSIO unit. ! shelf Screw < :A> PHYSIO unit
[Page 75] 3.20 Removing the Caster Units 3.20.1 For the SSA-770A (a) Remove the two rubber caps CA> at the bottom of the cover. (2) Removing the right side cover (a) Remove the two rubber caps cB> at the bottom of the cover. (b) Remove the two M4 screws cB> to remove the cover. (3) Removing the front cover (required to remove the front caster unit) (a) When the blank cover has been mounted, loosen the M3 screw CC> and remove the blank cover with the printer base. (b) Remove the four M4 screws CD> to remove the cover. (4) Removing the power supply section rear cover (required to remove the rear caster unit) Remove the four M4 screws CD> to remove the cover, (5) Removing the base front cover (required to remove the front caster unit) (a) Remove the two M4 screws cF>. (b) Loosen the two M4 screws cG> to remove the cover. (6) Removing the base rear cover (required to remove the rear caster unit) Loosen the two M4 screws cH> to remove the cover. 3-58
[Page 76] r Right side cover Cap cB> crew cB> - Front cover Blank cover LScrew CD> 2SCR) (SBPP4-12SCPj */ (SBPP4-12SCR) I Base front cover L screw cG> 3-59
[Page 77] (7) Removing the blocks Remove the four M4 bolts <I> to remove the two blocks supporting the shaft of the caster unit. (8) Removing the caster units (a) Remove the eight M4 screws cJ>. (b) Remove the eight M8 screws cK> and remove the caster unit by sliding it toward the front of the system. * Remove the caster unit at the rear of the system using the same procedure. Screw cJ> (SBPP4-IGSCR) Block 1’ I 1. (6B8-20SDM) Caster unit / 3-60
[Page 78] 3.20.2 For the SSA-700A (1) Follow the same procedures as for the SSA-770A until the base covers are removed. (2) Remove the eight caster fixing screws (four each on the right and left casters, M8). Caster Caster fixing 3-61
[Page 79] 3.21 Cleaning the Air Filters (1) Removing and cleaning the air filters (for service engineers) (a) Remove the two M4 screws to extract both the front and rear filters. (b) Remove the dust from the meshes using a vacuum cleaner. (c) Reinstall the filters by reversing the steps for removal. Air filter J (2) Removing and cleaning the air filters (for the user and service engineers) (a) Remove the filter cover and then remove the air filter. (b) Remove the dust from the mesh using a vacuum cleaner. (c) Reinstall the air filters by reversing the steps for removal. CAUTION: Do not start the system with the filters removed. 3-62
[Page 80] 3.22 Cleaning the Panel 3.22.1 Cleaning of key tops (1) Remove dirty key tops by pressing them downward from the surface of the panel. (2) Clean the removed key tops. (3) After cleaning, insert the removed key tops from the back of the panel. CAUTION: Use water or mild detergent to clean the key tops. Completely wipe off the water from the key tops before reinstalling them. Do not use chlorine-based cleansers, acids, or organic solvents. Also, be sure to reinsert the key tops at their original positions with correct orientations. There is a possibility of reinserting the key tops incorrectly. 3.22.2 Cleaning procedures for parts other than key tops Use a clean dry cloth for cleaning other parts. If dirt is difficult to remove, use a cloth that has been moistened with water or mild detergent and then tightly wrung out. CAUTION: Do not pour water or mild detergent directly on any parts. Do not use volatile organic solvents such as benzin or paint thinner or rags impregnated with chemicals. 3-63
[Page 81] 3.23 Adjusting the Operating Panel Locking Wire 3.23.1 Adjusting the up/down slide wire * Only for the SSA-770A (1) Remove the two caps <A>. (2) Remove the two screws <A> (M4) and remove the right side cover. e cover Screw <A> Cap.cA> (3) Loosen the nut. (4) Rotate the adjuster until the lever contacts the gas spring. (5) After adjustment, tighten the nut 3-64
[Page 82] Gas spring 3-65
[Page 83] 3.23.2 Adjusting the near/away slide wire (1) Perform the procedures in subsections 3.6.1 and 3.6.4, and the steps (1) and (2) in subsection 3.6.5 of subsection 3.6 “Disassembling the Operating Panel”. (2) Loosen the nut. (3) When locking cannot be released, loosen the adjuster to adjust the wire. (4) When locking cannot be performed, tighten the adjuster to adjust the wire. (5) After adjustment, tighten the nut. Nut- 3-66
[Page 84] 3.24 Replacing the Operating Panel Locking Wire 3.24.1 Replacing the up/down slide wire * Only for the SSA-770A (1) Removing the covers (a) Remove the two caps <A>. (b) Remove the two screws CA> (M4) and remove the right side cover. (c) Remove the two screws cB> (M4) and remove the right side cover for the peripheral units. (d) Remove the two caps cB=-. (e) Remove the two screws CC> (M4) and remove the left side cover. (f) Remove the two screws CD> (M4) and remove the left side cover for the peripheral units. (g) Remove the four screws cE> (M4) and remove the tray. (h) Remove the four screws cF> (M4) and remove the top front cover. Right side cover for the peripheral units Right side cover Screw cB> I Left side cover for the peripheral units Cap <A> Screw CA> (SBPPC12SCR) Left Screw cE> ‘IS crew CC> Cap’cB> 3-67
[Page 85] Removing the up/down slide wire (at the support unit side) (a) Loosen the nut. (b) Remove the wire from the lever. (c) Rotate the adjuster and remove it from the bracket. (The monitor and covers are omitted from figure.) 3-68
[Page 86] (3) Removing the monitor unit IA ! CAUTION: Two persons are required when removing the monitor unit. (a) Loosen the two screws CA> and four screws cB> to remove the monitor rear cover. (b) Remove the support rear cover at the rear of the operating panel. To remove the support rear cover, slide it backward with holding and pushing the center part. cc> Loosen the two screws CC> to remove the pinch shield cover. (4 Remove the two screws CD> to remove the clamps. (e) Remove the signal, audio, and power cables that are connected to the monitor unit. 0-J Remove the four M6 bolts cE> to remove the monitor unit with the boss installed. (9) Remove the two M4 screws cF> to remove the support front cover. 3-69
[Page 87] Screw <A> Support front cover D”ll NC’ t \ I^_^ .---nn”\ Screw <B> Support rear cover cable - Screw CC> - Pinch shield cover Screw CD> Signal cable Clamp 3-70
[Page 88] (4) Perform the procedures in subsections 3.6.1 and 3.6.2 of subsection 3.6 “Disassembling the Operating Panel”. (5) Removing the up/down slide lever (a) Remove the screw CA> (M3) and remove the clamp. (b) Remove the two screws <B> and remove the up/down slide lever. Screw <A> -p - Clamp (6) Replacing the up/down slide wire Install the new up/down slide lever in the operating panel. (7) Reinstalling the covers Reinstalling the removed monitor and covers by reversing the steps for removal. 3-71
[Page 89] 3.24.2 Replacing the near/away slide wire 3.24.2.1 For the SSA-770A (1) Pull the near/away slide lever of the operating panel toward you to slide the operating panel and the monitor upward. 3-72
[Page 90] (2) Removing the handle (a) Remove the caps <A> and cB>. (b) Remove the four bolts <A> (M4) and remove the handle. Cap <A> Handle Bolt <A> ~ (6B4-12SDM) Cap cB> 3-73
[Page 91] (3) Removing the covers Remove the two caps <A>. Remove the two screws CA> (M4) and remove the right side cover. Remove the two screws cB> (M4) and remove the right side cover for the peripheral units. Remove the two caps cB>. Remove the two screws CC=- (M4) and remove the left side cover. Remove the two screws CD> (M4) and remove the left side cover for the peripheral units. (9) Remove the four caps CC>. (h) Remove the four screws cE> (M4) and remove the top cover. (0 Remove the four screws cF> (M4) and remove the tray. 0) Remove the four screws cG> (M4) and remove the top front cover. Cap CC> Top cover ; Left side cover for -Screw cE> the peripheral units (SBPPC12SCR) Screw CD> (SBPP4-12SCR) I Right side cover for the oerioheral units Screw CC> Top front cover I I* .b Right side -b- Screw cG> -b cover *“1 II (SBPP4-12SCR) I *\** ‘. Screw cF> Tray L Cap <A> Screw <A> (4) Perform the procedures in subsection 3.4 “Removing the Monitor Unit” ! CAUTION: Two persons are required when removing the monitor unit. IA I 3-74
[Page 92] (5) Removing the near/away slide wire (at the support unit side) (a) Remove the two screws CA> (M4) and lift the bellows. (b) Remove the screw cB> (M3) and remove the clamp. (c) Loosen the screw CC> and rotate the wire retaining bracket to remove the tip of the wire. (d) Loosen the nut and rotate the adjuster to remove the wire from the support unit. \S crew <A> (SBPP3-IOSC re - Support unit Adjuster Nut Wire retaining bracket 3-75
[Page 93] (6) Perform the procedures in subsections 3.6.1, 3.6.2, 3.6.3, and 3.6.4 and steps (1) and (2) in subsection 3.6.5 of subsection 3.6 “Disassembling the Operating Panel”. (7) Removing the near/away slide lever (a) Remove the two screws CA> and remove the clamp. (b) Remove the two screws cB> and remove the bracket. (c) Remove the two screws CC> and remove the lever. (d) Remove the two screws CD> and remove the wire. Screws CA> Clamp 7h acket (8) Replacing the near/away slide wire Install the new near/away slide lever in the operating panel. (9) Reinstalling the covers Reinstall the removed monitor and covers by reversing the steps for removal. 3-76
[Page 94] 3.24.2.2 For the SSA-700A (1) Removing the monitor unit Refer to subsection 3.4. (2) Removing the right side cover Same as for the SSA-770A. (3) Removing the left side cover Same as for the SSA-770A. (4) Remove the screws <A> and <B> to remove the middle shelf and the metal plate for retaining the middle shelf. (5) Remove the screw CC> to remove the cover cl> (6) Remove the screws CD> and cE> to remove the covers <2> and <3> (7) Remove the screw cF> to remove the cover c4>. (8) Remove the screw cG> to remove the cover <5=-. (9) Remove the screw cH> to remove the tray. (10) Remove the screw <I> to remove the front cover. Screw <A> Middle shelf eta1 plate for retaining the middle shelf Monitor- Screw CD> Cover <2> - Screw ct “1 Front cover - Cover <3> w Side cover 3-77
[Page 95] (11) Removing the near/away slide wire (at the support unit side) (a) Remove the screw CA> (M3) and remove the clamp. (b) Loosen the screw cB> and rotate the wire retaining bracket to remove the tip of the wire. (c) Loosen the nut and rotate the adjuster to remove the wire from the support unit. Screw CA> (SBPP3-1 OSCR) wire retaining bracket 3-78
[Page 96] (12) Perform the procedures in subsections 3.6.1, 3.6.2, 3.6.3, and 3.6.4 and steps (1) and (2) in subsection 3.6.5 of subsection 3.6 “Disassembling the Operating Panel”. (13) Removing the near/away slide lever (a) Remove the two screws CA> and remove the clamp. (b) Remove the two screws cB> and remove the bracket. (c) Remove the two screws CC> and remove the lever. (d) Remove the two screws CD> and remove the wire. Screws CA> Clamp 7h wire Screw CD> 7 Screw CC> / _ r acket (14) Replacing the near/away slide wire Install the new near/away slide lever in the operating panel. (15) Reinstalling the covers Reinstall the removed monitor and covers by reversing the steps for removal. 3-79
[Page 97] 3.25 Removing the CD-R Unit (1) Remove the cap and the two M4 screws <A> and remove the left side cover. (2) Remove the blank cover <A>. (3) Remove the four M3 screws <B> and remove the rack cover. Blank cover CA> Screw <B> (SBPP3-8SCR) - L Screw <A> (SBPP4-12SCR) (4) Secure the CD-R unit in place using the two M3 screws ~0. -Screw CC> (SBPP3-8SCR) 3-80
[Page 98] (5) Connect the power cable and the SCSl5OP cable to the CD-R unit. NOTE: Command entry Execute the following program after the CD-R unit is installed or uninstalled. (1) Display the Service menu. (2) Select [Utility] - [Command Line], enter the appropriak command, below and press [Go]. (a) After the CD-R is installed D:Wplio\TusSetupScript\EnableCD.cmd (b) After the CD-R is uninstalled D:\Aplio\TusSetupScript\DisableCD.cmd (3) Restart the system. 3-81
[Page 99] 3.26 Cleaning the MO Drive When dust or foreign matter in the air adheres to the lens actuator of the MO drive, the performance of the MO drive may be degraded. Clean the lens actuator periodically using the lens cleaner specified below. Note: Use the dedicated lens cleaner specified below. Product name Fujitsu product number Product number for ordering Lens cleaner 0240470 CA90002-C980 Open the shutter of the lens cleaner before use and check the condition of the brushes. When the bristles of the brush have spread, the lens cleaner cannot remove stains or dirt from the lens completely. In such cases, replace the lens cleaner. Clean the MO drive using the lens cleaner as outlined below. The cleaning procedure takes approximately 15 seconds. (1) Start up the system. (2) Insert the lens cleaner. The lens cleaner is automatically loaded. Once the lens cleaner is loaded, the positioner moves forward and backward and the lens is cleaned by the cleaning brush of the lens cleaner. (3) After cleaning, the lens cleaner is automatically ejected 3-82
[Page 100] 3.27 Cleaning the Palm Switch <<Items required for cleaning>> l Gauze l Air spray l Tweezers 3.27.1 Cleaning procedures for the old-type palm switch (1) Remove the connector from the panel and remove the palm switch. (2) Press the two tabs on the sides of the palm switch and remove the top cover of the palm switch. Then, remove the ball. (3) Disconnect the connection cable between the dial board and the main board. Do not forcibly disconnect the cable connector. (4) Clean the detection shafts and 45” bearing (shaded sections in the figure) with a piece of gauze. (5) Clean out the inside of the case using an air spray. Take appropriate measures to prevent dust from flying out. Check for foreign matter inside the case. (6) Clean the dial rotation surface using a cotton swab. (7) Clean the gap between the dial and the cover (circumference) using the air spray. (8) Clean the inside of the dial unit using the air spray. (9) Connect the cable between the dial board and the main board. Bend the cable so that it does not interfere with the ball. (10) After cleaning the ball, place it back into the palm switch and reinsert the top cover. (11) Confirm that there are no abnormalities in ball rotation and dial rotation. (12) Reconnect the connector to the palm switch and reinstall the palm switch at its original position on the panel. 3-83
[Page 101] Cover joining tab (also on the opposite side) 1 Detection shafll 3-84
[Page 102] 3.27.2 Cleaning procedures for the new-type palm switch (1) Fit the dial unit attaching/detaching tool (service part) into the groove on the top of the dial unit. Rotate the tool counterclockwise to remove the dial unit. (2) Remove the ball from the palm switch. (3) Clean the detection shafts and 45” bearing (shaded sections in the figure) with a piece of gauze. (4) Clean out the inside of the case using the air spray. Take appropriate measures to prevent dust from flying out. Check for foreign matter inside the case. (5) Clean the gap between the dial and the cover (circumference) using the air spray. (6) Clean the ball and place it back into the palm switch. Fit the cover locking tabs into the slots taking care not to allow the ball to interfere with the cable. (7) Fit the dial unit attaching/detaching tool into the groove on the top of the dial unit. Rotate the tool clockwise to securely mount the dial unit. (Rotate the dial unit attaching/detaching tool until it locks into place with an audible click.) (8) Confirm that there are no abnormalities in ball rotation and dial rotation. Rotate the dial unit attaching/ Dial unit attaching/ detaching tool in this direction detaching tool to remove the dial unit. Slot position for releasing the ial unit- Slot position for locking the dial unit Rotate the dial unit attachina/ detaching tool in this direction to mount the dial unit. 3-85
[Page 103] Cover locking slots CAUTION: When removing the top cover of the palm switch, take care not to damage the cable inside. Be sure to reinstall the palm switch on the panel in the correct orientation. Use water or mild detergent to clean the ball. Completely wipe off the water from the ball before reinstalling it. Do not use chlorine-based cleansers, acids, or organic solvents. 3-86
[Page 104] 4. OPERATION OF EACH PWB 4.1 Front-End Unit The front-end unit consists of the Transducer Interface board (TI), the three Transmit/Receive boards (TR), the Receive Control board (RC), and the Continuous Beamformer board (CB). (2) Composition Transducer Interface board (TI) The TI board includes the three array transducer connectors and the CW pencil transducer connector. This board enables switching between the transducers. Transmit/Receive boards (TR) Simultaneous transmission/reception for a maximum of 192 channels is performed on the three TR boards. In addition, the transmission delay control and reception gain control functions are provided. Receive Control board (RC) This board performs reception phasing/addition in quad signal processing (up to 192 channels). In addition, the Real Time Control circuit (RTC), which controls the front-end unit, is provided. Continuous Beamformer board (CB) This board performs reception phasing/addition control and quadrature detection for the SCW by inputting 64 channels. This board includes the signal processing circuit, ADC, and dedicated transmission/reception circuit for the Pencil CW. 4-l
[Page 105] (3) Flow of signals in the front-end unit (a) Transmission system Transmission data from the RTC in the RC board is transferred to the TXPG in the TR board. The transmission delay and transmission waveform data are set. The TXPG has a transmission reference clock (40 MHz, four-phase) and outputs the trigger signals HOUT/LOUT to the PULSER based on the clock. The PULSER outputs a bipolar transmission signal based on the trigger signal. The transmission signal is output to the transducer connector port that is selected in the transducer port selector of the TI board. (b) Reception system The reception echo signal flows in the direction opposite to the flow of the transmission drive signal via the same route and is input to the reception circuit in the TR board. The reception signal is input to the RXAR-ASIC (PREAMP-IC). The signal is amplified by the specified amount and is input to the RXBF-ASIC in the RC board. The input signals are AD-converted in the RXBF-ASIC and reception beam forming is performed. Beam forming is performed simultaneously in a maximum of four directions and the beams are output to the BE board. (c) SCW transmission/reception system The SCW transmission system is the same as that described in (a) above. The 64-ch reception signal output from the TR board is input to the CB board that is connected to the RC board in parallel. In the CB board, quadrature detection is performed in the demodulator after reception analog beam forming. Then baseband signal processing is performed and the signals are AD-converted. The I/Q signals that have undergone AD conversion are converted to serial data and are then output to the BE board. (d) Pencil CW transmission/reception system The CB board includes the pencil CW transmission/reception circuit. The transmission output signals in the CB board are passed through the pencil transducer connector connected to the TI board using the cable and are then output to the pencil transducer transmission elements. The echo signals received in the pencil transducer are also passed through the TI and are input to the reception system in the CB board. After the signals are amplified in the pencil CW receiver circuit in the CB board, they are passed through the demodulator or later blocks and are AD-converted. 4-2
[Page 106] (4) Differences in the PWBs in the PV series APL10 PV8000 Major differences SSA-390A Transducer Interface 1 PBCNN260 3 A 360-pin transducer connector is used (TI board) HVSW PWB 6 (not compatible with PV7000). ECHOCNN PBSEL BOX 2 A relay for switching the transducer is PBSELMB 1 used. PBCONT 1 High-density PWB Transmit Receive 3 PULSER 4 The line voltage can be switched (TR board) between two values for every rate in the dual pulser. High-density PWB (64 chs per PWB) A total of 192-ch signals can be transmitted and received simultaneously with the three PWBs. Continuous Beam former 1 CW RECEIVER 1 Elimination of the granddaughter board. (CB board) Reduction of mirror artifacts Receive Control DBF 8 The 192 chs x four beams can be (RC board) 1 T&R-CONT 1 received in the 12 beamformer ASICs. RPG 1 160-MHz crystal oscillator High-density PWB (nine + one) Total 6 PWBs Total PBSELBOX + 15 PWBs (5) Calibration and tests required when the board is repaired/replaced The outlines of repair and replacement of the board are described below. For details, refer to the service manual (maintenance volume: 2D730-149E). (a) When the TI board is replaced Calibration is not required. Execute the test program from the service menu and confirm that no errors occur. (b) When the TR/RC/CB board is replaced The calibrations given below are required. For the procedure for each calibration, refer to the service manual (maintenance volume). After calibration, execute the test program from the service menu and confirm that no errors occur. l ADC offset calibration : Required when the TR, RC, or HDD board is replaced l SCW TX power calibration : Required when the TR board, RC board, or power-supply unit (PSA) is replaced. 4-3
[Page 107] ‘R-O 1 TXPG + to BE board + to BE board Transducer Port Selector 192 ch to 192 ch’3 from/to BE -F board 64 ch ‘i Block diagram of the front-end unit + to BE board
[Page 108] 4.1.1 TI The TI board (Transducer Interface board) includes the three array transducer connectors and enables switching of the transducers to be connected to the pulser/preamplifier. (2) Description of functions and operation (a) Array transducer switching This board contains the three 360-pin ZIF connectors for the array transducers and switches between the 192-ch echo line and the required control line using a relay. The transmission/reception signals are transferred to the pulser/preamplifier mounted on the TR board through the Backplane board. (b) Pencil transducer transmission/reception signal connection The connector box for the pencil transducer is installed in the cabinet and is thus connected to the TI board via a cable. The transmission/reception signals coming from the pencil transducer are transferred to the pencil pulser/pencil preamplifier mounted on the CB board through the TI and Backplane boards. (c) Support of the M-TEE transducer l Thermistor line Transfers signals coming from the temperature detection thermistor embedded in the M-TEE transducer to the ADC mounted on the RC board. (Up to two thermistors are supported.) l Angle detection line Transfers signals coming from the angle detection potentiometer embedded in the M-TEE transducer to the ADC mounted on the RC board. (d) Support of transducer with impedance converter This board supplies power for collector bias of the impedance converter. An over- current detection circuit is provided. The emitter bias power supply is provided from the TR board. (e) Support of 1.5 D transducer including a HVSW l This board receives the control data for the HVSW embedded in the transducer (walking bus) from the RC board and transfers it to the transducer. l This board supplies f 100 V for the HVSW embedded in the transducer. (f) Power supply for circuit inside the transducer This board supplies + 5 V for the circuit inside the transducer. An over-current detection circuit is provided. 4-5
[Page 109] (g) Error detection The TI board includes detection circuits for the following errors a. HV fault error This error occurs if the voltage for the HVSW embedded in the transducer is not correct. b. HV detection fault error This error occurs under the following conditions. If this error occurs, the crowbar circuit operates and stops supplying f 100 V. l The f 100-V power for the HVSW embedded in the transducer is applied to a connector where no transducer is connected. l The f 100-V power for the HVSW embedded in the transducer is applied to a connector where a transducer that does not require a high voltage is connected. c. HV diode fault error The diode is set on the echo channel in order to clamp the echo channel to prevent application of the f 100 V. This error occurs if the diode is broken and a current resulting from the high voltage flows in the echo channel. d. Relay fault error This error occurs if a relay for switching the transducer has values that are different from those of the relay control register. e. Over-current fault error This error occurs if an over-current occurs in the power (+ 5 V for Zconv) supplied to the transducer. (h) Transmission disable signal generation The TI board transfers the transmission disable signal (TXDIS) to the TR boards under the following conditions. l An HV fault error is detected (a in (g)) l An HV detection fault error is detected (b in (g)) l An HV diode fault error is detected (c in (g)) l A relay fault error is detected (d in (g)) l An over-current fault error is detected (e in (g)) l The HVSW in the 1.5 D transducer is in transition l The transducer selection relay is in transition l The transducer ID of the selected transducer is changed l An empty transducer connector is selected 4-6
[Page 110] (3) Equivalent boards in PV8000 The TI board includes the functions of five boards: PBCONT, PBSELMB, HVSW, PBCNN260, and ECHOCNN. (4) Differences relative to PV8000 Item I Aplio I PV8000 Transducer selection switch 1Mechanical relay 1Semiconductor high-voltage switch Transducer connector 360-pin ZIF connector 260-pin ZIF connector Number of echo channels 192 ch 128 ch 4-7
[Page 111] Motor Control - Switch Analog Control - Relay Switch Echo Booster Supplies LED Driver & 3, * I v 3 3 Posltion Sensor /’ I’ 192 192 cw Rx +/- I’ 1 2, I ” 1 CW TX +/- 7 I I I Transducer ID 7 ESD & 7, I’ Filtering / RF TX/RX 192, 192, , I Transducer interface block diagram 4-8
[Page 112] 4.1.2 TR The two primary functions of this board are first, to generate and deliver the transmit signals to the TI board, and second, to receive, amplify, and deliver the echo signals to the RC board. (2) Description of functions and operation (a) Transmit pulser This is the circuit which drives the piezoelectric elements inside the transducer electrically. The number of transmission waves and the pulse duration are controlled by the TXPG. The transmission pulse (bipolar transmission waveform) is generated in the circuit by increasing the low-voltage power supply in the transformer. In addition, this pulser has the two blocks in the transformer primary-winding side. Transmission voltage switching for each rate is achieved by applying different voltages to each of the blocks (dual pulser). (b) Transmit pulse generation The transmit pulse generation section is comprised of the following subsections: transmit control bus (TC bus), /TXCL and /TCWR signals, digital control, PECL clocks (four blocks), PECL synchronize signals (two blocks), power supply, 64 pairs of transmit control signals (Hout and Lout), and a transmit pulse generation ASIC (TXPG). (c) Preamplifier and variable gain amplifier (functions in RXAR ASIC) They amplify the echo signals from the piezoelectric element. The gain in the preamplifier is fixed to +I4 dB. The variable gain amplifier sets the gain from 0 to +75 dB by changing the control voltage (VTGC) from 0 to 1 V. (d) Antialiasing filter (functions in RXAR ASIC) This filter prevents the aliasing that occurs when the analog echo signals are converted to digital signals in the RC board. The cutoff frequency can be changed from 4 to 16 MHz by changing the control voltage (VAAC) from 0 to 1 V. The RXAR ASIC is set as given below in the SCW mode. The gain is fixed to 20 dB and the antialiasing filter is bypassed. 4-9
[Page 113] (e) T/R switch The T/R switch is used to switch the impedance of the echo signal channel between the transmit event and the receive event. During the receive event, this switch short-circuits the secondary transformer to provide a low echo signal channel impedance. (f) Address decoder (EPLD) and TC bus The address decoder uses digital control lines to perform the required control for the TR board. The address decoder serves as the “brains” of the TR board and communicates with the RC board via the TC bus. The TC bus serves as a bidirectional interface with the RC board. This bus is the digital communication hub for the TR board, which is used to access the memory in the TR address decoder, ID EEPROM, and TXPGs. The TC bus is comprised of two data buses and one address bus. The data bits are broken down into two 16-bit buses, TCOD[15:0] and TCI D[15:0], where each data bus is then connected to two TXPGs. TCOD[15:0] is sent to TXPG ASIC 0, TXPG ASIC 2, and the ID EEPROM, while TCI D[15:0] is sent to TXPG ASIC 1 and TXPG ASIC 3. (g) Reception resonance Reception resonance has been achieved by inputting inductance into echo signal lines in parallel. A coil is connected between the echo signal line and GND. There are four other inductance coils that can be selected. The resonance can be changed by using these inductance coils in various combinations. (h) Fault detection and protection circuit The TR board generates two types of interrupts: transmission interrupts and clock detection interrupts. a. Transmission interrupt A transmission interrupt is generated when TXENO and TXENI are both “H” simultaneously. This occurs only if there is a hardware fault or if the address decoder test bit is set. The address decoder EPLD always monitors the TXENO and TXENI lines and sets the interrupt status bit when it senses that both signals become “H” simultaneously. When this bit is set, the TR board stops transmission by making TXENO and TXEN 1 “L” and transfers the transmission inhibit signal and the interrupt signal to the system simultaneously to indicate that a system malfunction has occurred. This disables transmission to all three TR boards, not just the TR board where the error occurred. b. Clock detection interrupt A clock detection interrupt is generated when the PECL clock in the front-end unit stops before the TXPG completes transmission and the transmission for all three TR boards stops. This is extremely dangerous for the TR board because it leaves the TXPG output signals which drive the FET of the transmission circuit in an unknown state. If the TXPG output signals are fixed to “H”, the FET of the transmission circuit may be broken. 4-10
[Page 114] (3) Comparison with PV8000 The TR board is equivalent to the PULSER-H in the PV8000, except for the following items. Aplio PV8000 (1) Number of 64 CH 32 CH transmission/reception circuit channels per TR board (2) Transmission Bipolar transmission can be Only unipolar transmission can be performed. waveform performed. (Bipolar transmission cannot be performed.) (3) Transmission voltage The transmission voltage The transmission voltage cannot be switched switching can be switched for every for every rate. (Only PW and CW rate. (A dual pulser circuit is transmission voltage switching can be mounted.) pet-formed.) (4) Transmission pulse The transmission pulse is The transmission pulse is generated by generation method generated by using a switching the high-voltage power supply in the transformer to raise the low FET. voltage. (5) Reception preamplifier A new amplifier (RXAR An antialiasing filter, which is an individual ASIC) is used. This unit part, is mounted on the DBF. includes a preamplifier, a The variable gain range is 48 dB. variable gain amplifier, and an antialiasing filter in a single package. The variable gain range is 75 dB. 4-11
[Page 115] ----------- T--,--------------1 h&it~lik Transmrt Pulse TXCL Synch PECL Clocks 1 and T/R Switch Hout/Lout 1 Generation I 2 4 64 Pairs \ I I i 64 TX/RX (64) I I 4 ‘\ b 64 I JTAG 1 TXPG TITRS 64 (4x16-CH) AA Transmit Voltage Digital Control Digital Control Power Supplies Echo Booster TC Bus Biasing Voltage Power Supplies P5A 1 P5D 1 M2-AR 1 P3.3-D I ,-------------,--+----------------- 1 Analog Control 1 I , DC “Itage 4 , Ligital Control Echo Booster 1 Biasing Voltage I Digital Control + ZA-RXAR RCFEMRS w/op-amp RCFEREF 4 b 1 VCM (64x1-ch) RCTRVMD I VMID b TRRCS RCTRAAC ’ Differential OP , VTGC 2, b Amps D RCTRTGC VAAC \ 2 b \ ’ RCTREBB Power Supplies I M14A I Power Supplies P5 AC I P3.3-AR 1 P32-AC M2-AR , M2-AC I .-------________ -I Digital Interface Power Supply I 1 P5D I EPLD I P5l A - 1 TC Bus 4 b b 1 Buffered TC Bus P5A L.. b r .. + JTAG -I v TTL Synch TTL Clocks P3.3-AC Power b Transceivers PSTRTXVO Supplies P3.3AR PECL to TTL b PSTRTXVI ’ M2-AC P5AQ 4 M2 - Synch b 2 M2-OAQ M14A I I ~TEy PECL b PECL \ ’ P3-3AQ TXVOi 6 + Buffers \4 b I Power Supplies \ M14AQ TXVI I P3.3D \ b 1 P5D PECL Clocks TR functional block diagram for transmit pulse generation 4-12
[Page 116] 4.1.3 RCB The RCB (Receive Control Board) controls the front-end section, AD-converts the reception signals, and performs receive beam forming. (2) Descriptions of functions and operations The RCB includes the following functions. (a) Front-end section scan control: RTC (Real Time Controller) block The transmission/reception in the front-end section is controlled according to control information from the RM. The RTC block consists of an FPGA (Field Programmable Gate Array, programmable IC) and SRAM. The sequencer set in the FPGA transmits data to each block of the front-end section according to the scan control information to be stored in the register in the FPGA and SRAM. (b) AD conversion and receive beam forming: RXBF (Receive Beam Former) block The reception signals obtained for every ch are AD-converted with a sampling frequency of 40 MHz and a quantification precision of 10 bits. The required reception delay is applied to each converted signal and addition is pet-formed. This block consists of 12 ASlCs (RXBF ASICs). (c) Analog control signal generation: Analog Control block The analog control signals required in the front-end section (listed below) are generated. They are obtained by performing DA conversion for the digital codes output from the RTC. l VTGC (time gain compensation) This signal controls the preamplifier gain depending on the depth of reception. This signal is input to the RXAR ASIC (preamplifier IC) in the TR board. l VAAC (antialias control) This signal is used to control the signal bandwidth before the AD conversion. This signal controls the cutoff frequency of the low-pass filter (antialias filter) in the RXAR ASIC. l TXV (transmit voltage control) This signal controls the transmission pulse voltage. The signal is input to the TR PWB via the power supply unit. l VMD (voltage for midpoint bias) This signal controls the bias point of the reception echo signal line between the TR board and RC board (between the RXAR ASIC and RXBF ASIC). The signal is input to the RXAR ASIC. 4-13
[Page 117] (cd) System clock generation: Clock Generation block The clock to be used in the system is generated. A 40-MHz clock is generated by dividing the 160-MHz crystal-oscillator frequency. (e) Self-diagnosis: Diagnostic Circuit block The line voltage and standard voltage in the front-end section are measured for self- diagnosis of the system. This block also includes the function for detecting the temperature of the TEE transducer head and angle of the piezoelectric element surface. The signals to be measured are selected by the multiplexer and are input to the AD converter. The converted data is transmitted to the RM via the RTC. A circuit which stops transmission forcibly when a certain regulation value is reached is provided for the transmission voltage and TEE transducer head temperature detection signals. (3) LED An LED (DSI) is set at the free edge on the RCB. If this LED lights, the FPGA of the RTC is configured (programmed). The FPGA is automatically configured when the power supply is input. Therefore, if the LED does not light even when the power supply is input, the RTC circuit is not configured and therefore the system does not operate. This problem may occur as a result of contact failure of the FPGA terminal, device failure, or damage to the ROM that stores the program. 0111 RTC (FPGA) k- (4) Error detection The RCB detects the errors described below. (a) Excessive transmission voltage detection The RCB monitors the transmission voltage output from the power supply unit to the TR board by the TXV as described before. If this control signal exceeds a certain set value, the RCB generates an error and stops transmission. The value is set by the register in the RTC. (b) TEE transducer head high-temperature detection The RCB is provided with a function that detects the temperature of the TEE transducer head. If the temperature exceeds 43”C, the RCB generates an error and stops transmission. In addition, since the RCB mainly controls the front-end unit, when an error occurs in the TI, TR, and CB boards, the RCB receives it temporarily. The RCB then outputs an interrupt request to the RM. 4-14
[Page 118] (5) Comparison with the PV series Circuit block Corresponding board Main differences in the RCB n the PV series RXBF [Number of chs] 128 (SSA-390A) + 192 [Hardware size] Eight DBFs (maximum) + one RCB [Data output method for parallel simultaneous reception] Time-sharing output is performed in the DBF. However, parallel output is performed in the RCB and thus the signal band width does not decrease. RTC ?PG [Hardware size] T&R-CONT Only two types of ICs (FPGA and SRAM) are used in the Aplio Diagnostic Circuit T&R-CONT [Measurement of voltages] A measurement system is not provided with the PV series. In the Aplio, various voltages can be measured. Analog Control T&R-CONT [Bias point control] ‘art of DEMOD The bias between the preamplifier and AD converter is fixed in the PV series. In the Aplio it can be controlled by the VMD. Clock Generation Clock generation [Crystal oscillator frequency] section of DEMOD 240 MHz + 160 MHz [System clock frequency] 20 MHz + 40 MHz (The sampling frequency is the same in both systems (40 MHz).) 4-15
[Page 119] > > Analog echo signals > > Digital RF echo signals to BE > > RXBF RXBF 192 ch (4 parallel data with different profiles) (ASK (ASK X X 12) 12) > > > > AAA AAA UP UP BUS BUS Header Header BUS BUS V V TC BUS to TI and TR, and CB HV MUX Control BUS to TI FE BUS . from BE < 4 VTGC VAAC RTC Analog Control TXV VMD (FPGA & SRAM) Measurement BUS from Front-End Block Measured results Front-End Clocks Back-End Clocks ADC Clocks Block diagram for RCB 4-16
[Page 120] 4.1.4 CB (1) General description The CB board consists of the SCW channel adding circuit (summation), SCW/CW quadrature detection circuit, pencil CW transducer transmission/reception circuit, clock generation circuit, bus interface circuit, and self-diagnosis circuit. (2) Functions (a) SCW summation This function applies a delay of less than one period of the carrier waveform to the reception signals output from the preamplifier of the TR board and performs algebraic summation. (b) Quadrature demodulation circuit The band-pass filter (BPF) removes the higher harmonic elements. The mixer performs quadrature demodulation with the carrier frequency. The BPF and high-pass filter (HPF) located behind the mixer removes clutter elements. The selectable gain AMP circuit adjusts the levels of the Doppler signals, which differ depending on the type of transducer. The ADC converts the Doppler signals that have undergone quadrature demodulation to digital signals. These digital signals are then sent to the BE board. The phase and gain difference information in the sin/cos circuit is sent to IDPROM (5. Interface Section) and is corrected automatically in the BE board. If the correction is performed incorrectly, mirror artifacts increase. (c) Pencil CW TX/RX circuit The pencil CW TX/RX circuit is connected to the pencil transducer via the TI board. Bipolar transmission is used. For reception, Pencil CW mode is switched to SCW mode after the preamplifier. If switching is performed incorrectly, echoes from the pencil transducer cannot pass. (d) Clock generation circuit (pluse generator/LO) This circuit generates the clock which drives the pencil transmission circuit in the pencil CW TX/RX circuit, and also generates the reference clocks for the two mixers. These clocks are derived from the TXPG ASIC. If the TXPG ASIC is broken, images cannot be displayed in Pencil CW mode and SCW mode. (e) Inter-face The interface circuitry includes an address decoder EPLD, an ID EEPROM, a JTAG controller PLD, clock drivers, and buffers. 4-17
[Page 121] (3) Description of operation (a) SCW summation The SCW reception signals from the transducer are amplified to 20 dB (fixed gain of the preamplifier in the TR board) and are input to the 2nd preamplifier in the CB. The CB has a 64-ch reception signal line. When the sampling marker is moved in the SCW, the delay tap is selected accordingly from among the 16 taps on the delay line in the eight 8 x 16 multiplexers including 8 chs. Then algebratic summation is performed and reception beam forming is implemented. The delay time within one cycle of the carrier waveform is sufficient because the pulsed wave is periodic and a minimum quantification step of 20 ns is implemented with a maximum delay of 500 ns. The reception signals after algebratic summation differ depending on the type of transducer. Therefore the signal levels are matched using the variable attenuator RF gain and an appropriate dynamic range is allocated. The reception signals of the pencil (CW) and summation output of the SCW are switched by the multiplexer. Control of the 8 x 16 multiplexer, selection of the RF Gain, and output switching control between the pencil and SCW are performed by the address decoder (EPLD) based on the code of the TC bus from the RC. (b) Quadrature demodulation section 1) Selectable BPF The address decoder selects an appropriate band-pass filter (BPF) depending on the carrier frequency. 2) Quadrature demodulator A clock with a go-degree phase difference in carrier frequency is input to the two diode mixers and quadrature demodulation is performed for reception signals passed through the selectable BPF. The clock is generated by the TXPG based on the system clock (four-phase, 40 MHz) from the RC and is supplied to the mixer by shifting the level using the driver. 3) Baseband filter The DC (carrier elements) and added elements are eliminated by the BPF behind the mixer. The HPF behind the BPF is a fourth-order active filter and allows clutter elements to be removed. The required Doppler elements can be extracted The LPF is an antialiasing filter for the ADC and the cutoff is fixed to 156 kHz. This filter is a fourth-order active filter. 4-18
[Page 122] 4) Selectable gain amp The Doppler signal level differs for each transducer. Therefore, the signal level can be selected from 0 to 16 dB in 2-dB steps. Selection is performed by the address decoder (EPLD) based on code of the TC bus (BBGAIN). 5) ADC After performing IO-MHz sampling in Z-A type ADCs, the frequency element 156.25 kHz or more is eliminated in the digital filter and decimation is performed at the 312.5 kHz. The signals are output as serial data. The signals in the two ADCs (in-phase/quadrature-phase) are synchronized with the system clock via the EPLD and are output to the BE board. The I/Q signals in the two ADCs are combined into one synchronization signal with the serial data and they are then transferred in the differential format. (c) Clock generation circuit (plus generator/LO) In order to generate the reference clocks for mixers and the pencil CW transmission clock, the TXPG is configured to operate in infinite number of pulses mode (INPM). The PECL clock (four-phase, 40 MHz) is input to the TXPG and the frequency is set based on the TC bus data. (d) Pencil CW transmitter/receiver I> Pencil CW transmitter The pencil CW transmission circuit drives the pencil transducer piezoelectric elements. The TXPG outputs the clocks Houtl and Lout1 to the ACT08. This is controlled by the transmission enable signal (TXENI). The ACT08 drives the MOSFET and performs bipolar transmission by driving the transformer alternately. The transmission signal is transferred to the pencil transducer via the TI board. 2) Pencil CW receiver The reception signal returned from the pencil transducer is input to the preamplifier of the CB board via the TI board. After the weak signal from the transducer is amplified in the preamplifier, the CW receiver is connected to the selectable BPF and quadrature demodulation circuit by switching the multiplexer. The signal sent via the same route as for SCW summation and is processed. 4-19
[Page 123] (e) Inter-face 1) CB address decoder (EPLD) The CB address decoder (EPLD) decodes data from the TC bus and generates all the control signals required for the TXPG, ID EEPROM interfacing, control of buffer direction, and CB board. The TC bus is a bidirectional bus and consists of an 18-bit address, 16-bit data, an enable signal, and a handshaking signal. 2) PECL clocks The clocks for the CB board are derived from the PECL clock (four-phase, 40 MHz) RCFECK [0:3] output from the RC and the sync signal RCFECKE. The PECL clock is used in the TXPG. The clock converted from RCFECKO and RCFECK2 to TLL is used in the CB board. 3) ID EEPROM The ID EEPROM stores information related to the serial number, hardware revision, firmware revision, board ID, and drawing number. This device also stores the IQ calibration data that is used to correct the gain and phase differences between the in-phase and quadrature-phase circuits in the quadrature demodulator (quadrature demodulator circuit) on the BE board. The capacity is 8 k x 8 bits and this device is accessed via the CB address decoder EPLD and the TC bus. 4-20
[Page 124] (4) Adjustment points (5) Differences relative to PV8000 The CB is equivalent to the CW-RECEIVER in PV8000. However, there are the following differences. Aplio PV8000 (1) Reception PREAMP The PREAMP (20 dB) is set in The PREAMP is set in the CWRECEIVER. the TR-B. (2) Reception route The signals are transmitted via The coaxial flat cable is connected. the backplane (no cables). PULECHO and CWECH064 are connected. (3) I/Q adjustment No adjustment The two VRs are adjusted on board. The BE reads the IDPROM data and corrects it automatically. (4) ADC output Serial differential output Parallel output (5) PWB configuration 64 chs, one board 32 chs + granddaughter 32 chs (6) Pencil pulser Provided Not provided (but provided with the pulser) (7) Clock 40 MHz peel, four phases 60 MHz, one phase; the logic circuit is operated at 10 MHz. (8) Diagnostic cir. Provided Not provided 4-21
[Page 125] r--------------------------------- t SCW Summation I x 32 x4 8x4 2nd / Prg;w 8xl6MPX -t- Digital Control / , PreAmp -8 x 16 MPX 6dB Pulse Generator/LO Synch PECL Clocks CW Receiver JTAG CW Transmitter I Quadrature Demodulator output Data 1 k-B ! Diiital Control ’ Interface Circuitry Diagnostic Circuitry TDigital Control ~pGe--s- - - - - - - - - - i P12Al/P12A2/P12A3 I P14.2AQ / -b Regulator b I Transceivers ‘3 1 M14.2AQ N1OA1/N1092/N1OA3/N12A1 -b Regulator 4 __.^ 14 I P7A P5Al/P5A2/P5A3/P501 ---. --. -b Regulator /4 I RCTC Bus N5Al/N5A2)5A3 I ‘3 I I P5A P5D Synch -b Filter b ’ P3.3A P3.3D PECL Clocks -b Filter b I PSTRTXVO PSTRTXVO -b Filter b I -------------------L------------ -- CB Board functional block diagram 4-22
[Page 126] 4.2 BE Unit This unit consists of the FE unit echo signal processing system, the video signal processing system, the audio signal processing system, and the real time manager that controls the system. (2) Composition This unit includes the following boards. l RM board l BE board l VI board l IQDAS board (option) (3) Flow of signals BE board VI board ,- - - - - - - - - - - - - - -, ,---------------_ PA Video signal processing W L- - - - - - - -------I IQDAS board ,---------------, I Audio signal processing l---------------l l---------------J Flow of signals in the BE unit (4) Equivalent units in the PV series This unit is equivalent to the following parts in the PV series. l Receiver of the T&R digital unit l Doppler unit l DSC unit l System unit 4-23
[Page 127] 4.2.1 BE The BE board consists of the following blocks and performs data processing in all image modes after performing phasing/addition. l DR (Digital Receiver) l EP (Echo Processor) l FP (Flow Processor) l SIP (Signal & Image Processor) l SC (Scan Converter) l PCI I/F The features of the architecture of this board are the miniaturization of circuits using an ASIC with a high level of integration and the provision of a buffer memory for each block to enable independent data processing. (2) Functions The composition and functions of each block in the BE board are described below. (4 DR This consists of four DR-ASICs for quad signal processing (QSP). The following functions are performed in all modes. l DC filtering Removes the DC components. l Digital TGC (time gain control) Controls the STC. l Baseband conversion Performs orthogonal detection depending on the reception center frequency data and converts the frequency to the baseband signal to obtain real and imaginary signals. l Decimation filter Performs decimation for input data with 1024 samples or less per vector. l Dynamic filter Sets the LPF to the echo signal depending on depth. In addition to the functions shown above, the following function is also performed in the PWD mode. l Range gating Range gating integration is performed for the sample volume. 4-24
[Page 128] (b) EP This consists of one EP-ASK and a persistence memory. EP operates in 2D-mode and M-mode and performs echo signal processing as outlined below. l Magnitude calculation Power operations are performed for real and imaginary input data. l Log compression Log compression is performed for the echo signal. l Blending Combination focus and other vector combination operations are performed. l Edge enhancement Edge enhancement is performed for the echo signal. l AGC Feedback processing and echo enhancement are performed for the echo signal. l Black hole and noise spike filtering Noise appearing as black and white points is eliminated. l Persistence Frame correlation processing is performed. (cl FP This consists of one FP-ASIC, a buffer memory, two FPGAs, a filter coefficient memory, and a persistence memory. FP operates in CDI mode and MCDI mode and performs color Doppler processing. l Corner turning Real and imaginary data input from the DR is stored in the buffer memory and the data is rearranged when processing is performed by the MTI filter. l Wall filtering MTI filtering is performed. FIO, which reduces motion artifacts, is also pet-formed. l Autocorrelation Auto-correlation is performed. l Spatial averaging The filter to be used in spatial averaging (smoothing) is set 4-25
[Page 129] l Hole filtering Processing for the filter and blanking, which remove clutter and noise, are pet-formed. l Persistence Frame correlation processing is performed for the color Doppler data. l Clutter processor Analysis is performed for the removal of clutter components by FIO Cd) SIP This consists of one DSP, one FPGA, and a buffer memory. SIP operates in trace image display modes (PWD, CWD, M, and MCDI) and at the time of waveform (such as ECG) display. In PWD and CWD modes, the following processing operations are performed. l Spectrum generation (FFT) Signal analysis is performed by the FFT and a spectrum is generated. l Audio left/right signal generation Separate left and right Doppler signals are generated. l Display column generation The space to be used for spectrum display is generated. l Derived parameter calculations Real-time calculation such as auto-trace are performed. In the M and MCDI modes and at the time of waveform display, the following processing operations are performed. l Display column generation The space to be used for waveform display (such as M, MCDI, and ECG) is generated. l Tissue/flow decision (only for MCDI mode) Whether the pixels displayed in MCDI mode represent tissue or blood flow is determined. 4-26
[Page 130] W SC This consists of the one SC-ASK, a CINE memory, and a buffer memory. SC operates in the 2D and CDI modes and performs the following 2D image system processing operations. l Scan conversion Ultrasound data is converted to TV format data l Tissue/flow decision Whether the pixels displayed in CDI mode represent tissue or blood flow is determined. l Peak hold (only for CDI mode) A maximum value is displayed for each pixel in CDI mode. l Frame interpolation (only for CDI mode) Frame interpolation processing is performed in CDI mode l 2D tine storage Holds the 256 MB tine memory. l Cine replay (under CPU control) Cine display is pet-formed. Operations such as playback speed are controlled by the CPU. (f) PCI I/F This function interfaces the HOST bus (PCI bus) and the following buses. l Imaging bus Data bus to be sent to the VI board. The imaging bus includes the 32-bit bus from the SC and the 18bit bus from the SIP. The bus clock for both buses is 40 MHz. l BE internal local bus Internal local bus used for controlling each ASIC. This bus operates at 40 MHz with a 32-bit width. l FE bus The FE bus is used by the CPU to access boards in the FE unit. This bus operates at 40 MHz with a 32-bit width. (3) Flow of signals The flow of signals for each mode is shown on the next and subsequent pages. 4-27
[Page 131] RC BOARD RTC SUBSYSTEM ----- TR BOARDS I TXPG ASICS 1 I---------------------.------------------ 1 ; DR ASIC 11 EPASIC l Il DC Digital Baseband I Cancel + Decimator + Mag Calc+ Log TGC + Conversion + Compress + Blending I II IL-------------------i, II I Scan Lateral Black Hole 11 Persist Convert 4 II EE Noise Spike+ Axial EE f II I ------------------ 1 SCASIC -------- BE BOARD Pixel RGB _ + Monitor l Unpack + Lookup Video Routing -b Color Key NTSC/PAL Graphics -b Mixer + Video --- Controller Formatter 1 Real-time Manager 1 VI BOARD 1 IO BOARD 1 Flow of signals in 2D mode
[Page 132] RC BOARD r RTCSU~~VSTEM TR BOARDS ’ rXTG%k- ’ TX T/R 1 BeaAt%mer 1 ’ Pulser + Switch _--- ’ A;$g ! IBeam%rme* Tuning 1 ARASICS ; I ---- Clutter- Glut -. ter Sianal BE BOARD Proc. Adaptive Control rDRAslc _ - - - - - - - - - - - - - - - - ,,3 I DC Digital Baseband Cancel + Decimator + DE;e;‘c Filter TGC -, Conversion + I FP ASIC Pixel I Peak Tissue/ Frame Scan Shading + - Pack 4 I Hold Flow f lnterp Storage 1 I SC ASIC --------------------- J + Monitor Video Routing NTSC/PAL Graphics Video ___ Controller Formatter 1 Real-time Manag! 1 VI BOARD 1 IO BOARD 1 Flow of signals in CDI (B) mode
[Page 133] RC BOARD r RTr SUBSYSTEM TR BOARDS ’ rXyG%&- ’ ’ TX Beamformer Control 1 1 Rx Beamformer Control 1 I I I Analog Controls )‘I Bean%rmer 1 BE Header Generation 1 1 FE Header Generation I I L----J BE BOARD ------------------~ , ,-------------- -DFASI C I II I I DC Range Digital Baseband Wall I ! b Cancel + Decimator + Gate Gain +Conversion + Filter - Filter Integ. I I L-------------------A ’ I ------------ A I I---- I L/R Audio Smooth IFFT Generate I Missing Signal SjWl” Median m 1 W-W f Filter f olumn 4 Estimate 4- I DSP b Column + unpack + Display Buffer Formatter Graphics NTSC/PAL Controller - Video + Formatter --- Real-time Manager Mic In Flow of signals in PWD mode
[Page 134] RC BOARD I IT: SUB-&STEM ----- TR BOARDS 1 TXPG ASKS I TX TX T/R I Beamformer 1 ’ Pulser -W Switch Rx ! Analog TGC f Tuning Beamformer’ 1 1 RXBF ASKS I I -----A LARAs’cs I I ---------------_---------------- I ERTSE - EP ASIC I I l I DC I l Black Hole Digital Baseband y?;eT’c -+ Cancel + Decimator + *Mag Calc+C Log Axial EE+ Noise I TGC W Conversion l empress+ Spike I I l I I I L----------------------------------- I ’ J _------------- l Display Peak I - Pixel Column f Detect/ 4 I Pack ’ I Generate Hold I DSP -------------A BE BOARD b Column + Unpack + Display + Video b Signal Buffer Formatter Routing Routing Graphics NTSC/PAL Controller RGB Lo&up -b co/$-e~“y + Video Formatter -- Real-time Mana jer 1 1 I I I I I Flow of signals in M mode
[Page 135] RC BOARD 1 IT: SUB-&STEM ----- TR BOARDS 1 TXPG ASICS 1 I TX Beamformer Control I I Rx Beamformer Control I I I Analog Controls )‘I Bean%rmer I ’ Pulser Tx I BE Header Generation 1 I FE Header Generation I I I L----J I b Beamforme I ; RXBF ASICS 1 -----1 LARAs’E 1 --------------_ BE BOARD r ii ,&c - - I DC t, Cancel + Digital + Baseband Decimator w Dynamic + TGC Conversion + Filter --------------- I -FpA%- I L-------------------J I -------- r - - -Dyp- Wall Auto- Axial VelNarl Keyhole I e Filter Corr. Filter + Power + Filter t I Cine I II Storage ! L------------------J Pixel Display Tissue/ ID Pack 4 Column Spatial Generate f Flow Filter ------------- A b Column + Unpack + + Video Buffer Formatter Routing NTSC/PAL Graphics Color Key + Video --- I I Mixer Formatter 1 Real~~r~~~ager 1 - Flow of signals in MCDI mode
[Page 136] (4) Equivalent units in the PV series This board is equivalent to the following boards in the PV series. l RECEIVER board in the T&R digital unit l DOP-IO board in the Doppler unit l DOP-PROC board in the Doppler unit l DOP-CONT board in the Doppler unit l DOP-ADAP board in the Doppler unit l BDSC board in the DSC unit l IM board in the DSC unit l CDSC board in the DSC unit 4-33
[Page 137] INPUTS FROM TR BOARDS CHO PCI DR ASICS ’ PCI INTER BUS (4) b FACE FE BUS TO - ID PROM LOCAL/MICROPROCESSOR BUS BOOT I I PROM FP ASIC FP FPGA I I I I SIP FPGA DSP IP DSP b LOCAL ROUTER 6701 SDRAM PHYSIO INPUT OUTPUT TO PHYSIO INPUT FROM CB BOARD TRIGGER CONTROL FE TEST BUS IP INTERFACE I TAG TES ‘p~co~!fiol 1 BE TEST BUS RCBECK Ei 40-MHz TTL CLOCKS I PECL CLOCK Block diagram of the BE board
[Page 138] 4.2.2 VI (Video Interface) The VI (Video Interface) board manages the system images from the BackEnd (BE) board, overlay graphics from the Realtime Manager (RM) board, and routes the system monitor video signal to the IO board. The VI also processes the NTSUPAL video signals in order to support external OEM devices. The VI manages the system audio and controls audio I/O to external devices. (2) Functions The functions implemented by the VI are outlined below: l Video processor display manager (ASIC) l Video processor image formatter (FPGA) l System video display mode control l Variable video resolution control l Bidirectional scaling engine for NTSUPAL l SVGA RGB encoding for system monitor l NTSC/PAL video decoding l NTSC/PAL video encoding l Simultaneous S-video, composite, and RGB analog output for NTSC/PAL l System audio control (I/O select, volume/tone control, LPF, HPF, etc.) l Doppler audio D/A conversion l System power supply control and status check l 64 bit, PCI bus interface l JTAG boundary scan interface 4-35
[Page 139] (3) Explanation of the block diagram The figure on the next page shows the top-level functional block diagram of the VI board. There are two separate PCI master/target devices on the VI board. One is the VPDM (video processor display manager) ASIC and the other is the VPIF (video processor image formatter) decoder FPGA. Each PCI master/target interface in the devices supports 64-bit address/data transactions at a 33-MHz bus speed. The PCI bus allows the VPDM to receive image data (e.g. 2D images, Doppler traces, ECG waveforms, etc.) from the BE board. It also allows both the VPDM ASIC and the decoder FPGA to communicate with each other. The digital SVGA link from the graphics subsystem of the RM to the VI carries 24-bit RGB data for the CPU overlay graphics and horizontal and vertical synchronization signals. The RM also sends the pixel clock signal to the VI, which is synchronized with the RGB data. The VPDM drives the system monitor through a triple 8-bit video DAC to display ultrasound images with superimposed overlay graphics. The maximum resolution supported will be 800x600 at a non-interlaced vertical refresh rate of 60 Hz for NTSC systems and 75 Hz for PAL systems. The 16-bit data bus from the VPDM to the VPIF encoder FPGA carries 16-bit YCbCr (4:2:2) non-interlaced video data with a resolution of 800x600 pixels. The VPDM sends video data that contains the video information displayed on the system monitor to the VPIF, where the data is converted to the NTSC/PAL image size and format and is then routed to the IO board. The NTSC/PAL video encoder receives image data in the 4:2:2 YCbCr format from the VPIF encoder FPGA and converts it to analog signals for an external monitor or a VCR The output analog signals are Y and C, color composite and RGB components. The NTSC/PAL video decoder converts analog Y/C signals from an external video device to image data in the 4:2:2 YCbCr format. Hsync, Vsync, field ID, and blanking information are included in the data stream. In VCR playback mode, the decoded VCR video data is scaled up and converted to non-interlaced video data in the VPIF decoder FPGA and is then sent to the VPDM via the PCI bus. The system audio control block provides input/output audio source select, volume and tone control, test tone generation, crossover networks, and audio D/A conversion. (4) Equivalent units in the PV series The functions of this PWB are equivalent to those performed by the RGB-CONV and TV- PROC in the PV series. 4-36
[Page 140] NTSUPAL encoder interface The “encoder” interface block refers to the external digital video encoder part used to generate NTSUPAL analog video signals. The 16 bit YCbCr data is converted into the CCIR-601 4:2:2 8-bit digital video format, including synchronization codes. This block also outputs disrete blanking, hsync, and vsync signal for diagnostics. The digital video output from the encoder FPGA includes additional blank lines and pixels in order to avoid problems with overscan in playback monitors. The NTSC timing is shown in the following table. Table: External NTSC Parameter I Pixel clock cycles I Time Pixel clock 1 74.07 ns (13.5 MHz) Line active 692 63.556 ps 81% Line FP I 2.074 ps Line sync I 64 I 4.74 ps Line BP I 74 I 5.48 ps Line blanking I 166 I 12.296 ps Line time 858 (H) 63.556 ps 15.7 kHz Line average data rate I 0.53 BytesJ(40.5 MHZ cycles) I Vertical active 234.5 H 14.904 ms 89% active Vertical FP 10 H 0.635 ms Vertical sync 3H 0.190 ms Vertical BP I 15 H I 0.953 ms Vertical blanking I 28 H I 2.22 ms Field time 262.5 H 16.683 ms (225,225 13.5 MHz clocks) 59.94 Hz 72% active Frame time 525 H 33.36 ms (450,450 13.5 MHz clocks) 29.97 Hz 72% active Image size 324,548 pixels 4-37
[Page 141] Detailed vertical timing for NTSC mode is defined in the following table, with the 525 total lines numbered from 0 to 524. This format included more blanked lines than the NTSC specification in order to avoid problems with overscan in external monitors. Table: NTSC vertical line timing First line 1 Last line 1 Line count 1 Function 523 I 20 I 23 Blanked lines per PAL specification 21 23 3 Extra blanked lines. 24 258 235 Active lines, even field 259 260 2 Extra blanked lines. 261 I 282 I 22 I Blanked lines per PAL specification. 283 I 286 I 4 I Extra blanked lines. 287 520 234 Active lines, odd field 521 522 2 Extra blanked lines. Timing for PAL signal generation is shown in the following table. Table: External PAL Parameter Pixel clock cycles Time Pixel clock 1 74.07 ps (13.5 MHz) Line active 688 50.963 ps 80% active Line FP 34 2.519 ps Line sync 64 4.74 ps Line BP I 78 I 5.78 ps Line blanking I 176 I 13.037 ps Line time 864 (H) 64 ps 15.625 kHz Vertical active 282.5 H 18.08 ms 90% active Vertical FP 6H 0.384 ms Vertical sync 2.5 H 0.160 ms Vertical BP 21.5 H 1.376 ms Vertical blanking 30 H 1.92 ms Field time 312.5 H 20 ms (270,000 13.5 MHz clocks) 50 Hz 72% active Frame time 525 H 40 ms (540,000 13.5 MHz clocks) 25 Hz 72% active Image size 388,720 pixels 4-38
[Page 142] Detailed vertical timing for PAL mode is defined in the following table, with the 625 total lines numbered from 0 to 624. This format included more blanked lines than the PAL specification in order to avoid problems with overscan in external monitors. Table: PAL vertical line timing First line 1 Last line 1 Line count 1 Function 623 I 21 I 24 Blanked lines per PAL specification 22 I 24 I 3 I Extra blanked lines. 25 307 283 Active lines, even field 308 309 2 Extra blanked lines. 310 I 334 I 25 I Blanked lines per PAL specification. 335 I 338 I 4 I Extra blanked lines. 339 620 282 Active lines, odd field 621 622 2 Extra blanked lines. 4-39
[Page 143] For System Monitor RGB 3 / VPDM b P 4 ,’ Buffer Display Manager PM) 24-bit RGB from RM Image 1 B-bit I 64-bit PCI Bus PCI Interface L 64-bit PCI Decoder FPGA 32-bit External Video outputs 1 1 Frame Buffer Analog Y In 8-bit YCbCr b NTSC/PAL Analog C In b Decoder - 1 Power Supply Control Serial l2C Bux System Audio Control Block I I T Audio Test Tone Test tone decontrol Crossover Generator hAtput T input Audio Control MUX VI board functional block diagram 4-40
[Page 144] 4.2.3 RM (Realtime Manager) This PWB contains the host CPU whose main function is to activate the software. This PWB also includes the external memory unit I/F such as that for the HDD and the input/output I/Fs such as those for SCSI, Ethernet, and USB. (2) Composition (a) CPU and chip set The two Intel Pentium III processors are mounted. The CPU functions as a dual- CPU board. The Intel840 chip set is used to control this PWB. (b) Main memory Up to four Rambus memory modules can be mounted. (c) Graphics card The graphics card connected by the 4X AGP I/F outputs various graphics and characters as 24-bit digital RGB signals. These graphics and characters are combined with the ultrasound image in the VI board and then displayed on the monitor. (d) Audio card This performs recording and playback for various audio signals. (e) I/Fs The I/F controller functions listed below are provided. The I/Fs are connected to each device via the BP board. l PCI : Connected to the BE and VI boards in the 64-bit PCI bus through the PCI bridge. (A 32-bit PCI bus is connected internally.) l IDE : Supports the UltraDMA66 and is used for connection to the HDD. l SCSI : Supports the Ultra Wide SCSI and is used for connection to the CD-R and MO drives. l USB : Used for connection to the control panel and external USB devices. l Ethernet : Supports 1O/l 00 Base-T Ethernet. 4-41
[Page 145] (3) LEDs The following five LEDs are provided at the free edge of the PWB in order to indicate the PWB status. l PWR (green) : Monitors the power supply status and lights when the power is ON. l BOOT OK (green) : Lights when the boot process is completed normally after the self-test function (POST function) when the power is turned ON. l BIOS ERR (amber) : Lights when an error is detected by the POST function. l IDE ACT (amber) : Lights when an IDE device (HDD) is accessed. l SCSI ACT (amber) : Lights when the SCSI device is accessed. (4) Jumper setting When this PWB is used, the following three jumpers are set. l WE-BB : BIOS boot block write enable jumper (No jumper installed.) l FRC-REC : BIOS force recovery jumper (No jumper installed.) l SPL-SCR : BIOS boot logo select jumper (Jumper always installed.) (5) Equivalent units in the PV series This PWB is equivalent to the CPU board in the PV series. 4-42
[Page 146] Digital 24-bit RGB bus to VI -7 600-800 MHz 600-800 MHz P6 FC-PGA FC-PGA Pentium III Pentium III t7-q INTEL 840 MCH Connector for w MAIN MEMORY (RAMBUS) 1 GBls off-the-shelf VERTICAL RIMM HOST 4X AGP 32 MB 2048 MB MAX. CONTROLLER 1.6 GB/s 533 MB/s 266 MB/s INTEL 840 FWH PCI 64 HUB PCI 32 HUB P64H ICH 82806AA 82801 AA 16 bit/66 MHz 8 bit/66 MHz HUBLINK I/F HUBLINK I/F ’ PI &P2 1 64-bit PCI BUS (33 MHz) -1 SM Bus 1 PWR Button PCI-32/33 AC’97 I/F cs4299 cs4299 P5 Block diagram of the RM 4-43
[Page 147] 4.2.4 IQDAS (IP) <introduction> IP PWB is included in UIIP-770A Part No. : PM30-32120-1 Part Name : IQDAS PWB (IQDAS stands for IQ Data Acquisition) Requirements : Aplio on & after V4.00 <Circuit blocks> 1) Data-IF section (between BE PWB) 2) Capture memory section 3) Temporary storage memory section 4) DMA Controller section 5) PCI-IF section <Outline of circuit functions> This board takes IQ (In Phase, Quadrature Phase) data or IM (Image Memory) data output from BE PWB SIP section into capture memory with real time by memory R/W control FPGA. It is possible to store the particular part of taken data into the temporary storage memory. Because of this temporary storage memory, it is possible to continue capturing data without stopping examination finished, several captured data in the temporary storage memory is transferred to the system memory by DMA via PCI bus. <Block diagram> BE board PCl64@33MHz - ’ PCI 16@40MHz I I I IQDAS board f-71 lMHz
[Page 148] <<Circuit functions>> Functional Block Major Function Back Plane CNN Connector with APL10 main unit PCI-PCI Bridge IP board is connected with Aplio via PCI bus. There is PCI-PCI Bridge on IP board for separate PCI bus inside IP board. PCI-LOCAL Bridge The Interface device between IP board internal PCI bus and Local bus. LOCAL SRAM 1 The working RAM for PCI-Local bridge device BOOT ROM IP board information is stored in this ROM. DIMM memory for data capture The memory to capture IQ data or IM data in real time 512 MB DIMM memory Temporary data storage memory Temporary data memory; data taken from the above DIMM memory (DIMM) 512 MB DIMM memory DATA INPUT IQ data/lM data input buffer (only Buffer) FPGA Capture memory, Temporary storage memory Read/Write control, DIMM initialization, Refresh function, Interrupt function CONFIG ROM 1 Configuration ROM for FPGA CLOCK DRIVER IP board basic clock generator (33 MHz oscillator) Making 66 MHz from 33 MHz (40 MHz from BE) PWB RESET IP board local reset DIP SW Several hardware settings: Do not change the original settings SIO : for PCI-Local Bridge device IC initial setting s4 : for PMC interrupt S5 : for PCI/PMC (future use) settings s7 : for FPGA settings s9 : for FPGA ROM/SDRAM settings S6 : for PROM FPGA JTAG settings S2/S3 : for Clock driver settings Sl : board Reset UXSI 1 : FPGA configuration switch Power supply circuit I3.3v 2.5v 5v 4-45
[Page 149] IQ data - I IM data >V > Mem Cont FPGA Xilinx XCV IOOOE >V 64 bit 33 MHz PCI Bus PCI-PCI / 4 b b Bridge / Local 64 bit 33 MHz PCI Bus PCI/Local / 4 b Bridge / I Cont I -1 PMC ; SDRAM; ------ IQDAS Board Function Block Diagram
[Page 150] 4.3 System-Related Boards 4.3.1 I/O The I/O board connects with various I/O devices (VCR, HDD, etc.) located both inside and outside the system and the BP (Back Plane board) and transmits signals and power between them. In addition, appropriate cable connections are provided using the required shields and decoupling in order to meet EMI, RFI, and ESD requirements. (2) Functions The following functions are provided. (a) Power Power is supplied to external units such as the control panel, Physio module, microphone, EIDE unit (HDD), and USB port. (b) Video System monitor output, S-Video input/output, RGB output, and C-Video output from the VI board are performed. (c) Audio The following signals are output from the VI board. l Speaker audio output and headphone audio output. l Microphone audio output. (d) Speakers The system speaker signals from the VI board in the audio power amplifier are output to the system speakers (one pair of tweeters and one pair of woofers). (e) Communication/control The following interface functions are provided. l USB hub and ports l System reset signal input l Front panel (for the user) and rear panel interfaces l EIDE hard drive l SCSI/RDD l EthernetTl O/l 00 l Serial port (COMI) l Parallel port (IEEE1284) l Status LEDs 4-47
[Page 151] (3) Statuses indicated by the status LEDs , LAN Tl O/T1 00 Status (yellow=TlO, green=TlOO) - RM Boot Status (orange=Boot, green=Boot OK) EIDE Disk Access (green, ON=Access) +5V Internal Disk Power Status (green, ON=OK) +12V Internal Disk Power Status (green, ON=OK) +5V Removable Disk Power Status (green, ON=OK) +I 2V Removable Disk Power Status (green, ON=OK) +3.3V Power Status (green, ON=OK) +5V Power Status (green, ON=OK) _ +12V Power Status (green, ON=OK) J5B J5A USB Power Status (green, ON=OK) Jl 0’9 cl 4-48
[Page 152]
[Page 153] 4.3.2 Front panel (1) Outline and functions The signals and power (+I2 V/+5 V) from the IO board are transmitted to this panel and are output to the operating panel, CNN for the footswitch, and removable disk drive. Also, headphone, microphone, and shutter input terminals are mounted. For details of the connections, refer to section 9. 4.3.3 Rear panel (1) Outline and functions The output signals for the external devices from the I/O board are transmitted to this panel. Also, an Ethernet terminal, a USB terminal (2CH), a C-Video output terminal, an S-Video output terminal, and a RESET switch are mounted. For details of the connections, refer to section 9. 4-50
[Page 154] 5. OPERATING PANEL Data communication with the main unit is performed using several signal lines and one USB cable. The USB-HUB is inside the control panel. The functions of this panel are executed independently via this HUB. +3.3-V, +5-V, and +12-V power supplies are used. (2) Descriptions of functions and operation The functions related to USB connection are described below. (a) CARD-PC section (RS-232C connection to the USB via a USB/serial conversion chip) l Displays menus on the TCS display by transmitting data from the main unit. l Converts information input by touch keys to information for switches on the TCS menu and transfers them to the main unit. (b) SUB-CPU section (RS-232C connection to the USB via a USB/serial conversion chip) l Transfers information input by various switches (other than the POWER switch) to the main unit. l Transfers information via the rotary encoder. l A/D-converts the rotary VR and STC sliding VR analog data and transfers them as digital information to the main unit. l Transfers information input by the footswitch to the main unit. l Lights and extinguishes various LEDs by transmitting data from the main unit. l Sounds the buzzer when the switch is pressed by transmitting data from the main unit. (c) Keyboard section (USB connection) l Transfers the data input by the keys on the keyboard (d) Trackball section (USB connection) l Transfers the information input by the trackball to the main unit. Note, however, that LED lighting and the dial section (encoder) are controlled by the sub CPU. The following operations are implemented and controlled using signal lines. They cannot be performed inside the panel. (e) Control of the POWER switch and LED lighting (f) Control of STANDBY LED lighting 5-l
[Page 155] (3) Additional terms TCS : Touch Command Screen. This module integrates the touch keyboard with the LED screen. The 10.4-inch LED has 640 x 480 pixels. CARD-PC : This module integrates the CPU, RAM, and video controller functions. USB-HUB : There is one USB cable from the main unit. However, the USB-HUB inside the control panel has five ports. Two of these are connected to the USB ports in the left side of the control panel. These USB ports can be used for connecting external USB devices. However, if the USB devices do not have external power sources, the total consumption currents for the devices connected to the two ports must not exceed 150 mA. (4) USB connection diagram USB-HUB (5 ports) Keyboard Trackball USB Port USB Port Left side of the control panel USB/SERIAL (RS-232C) conversion (4 ports) _____ ___________ Unused ______________________ SUB-CPU : USB connection SERIAL (RS-232C) connection 5-2
[Page 156] 6. POWER SUPPLY UNIT 6.1 Power Supply Unit Configuration The power supply unit of this system consists of the following four units. (1) Transformer tray (SSA-770A) Trans-unit (SSA-700A) (2) AC/DC tray (SSA-770A) AC/DC-unit (SSA-700A) (3) PSA unit (4) PSD unit or Trans-unit AC/DC tray or AC/DC-unit 6.2 Block Diagram of the Power Supply Unit The block diagram of the power supply unit of this system is shown below. System rack > 200 V (to monitor) 100 V x3 (OEM output) 48 V 9.5 A AC100 V - 5 outputs 230 V - 255 w t 1 6-l
[Page 157] 6.3 Unit Specifications (1) Line voltage 100-V system 90 to 132 VAC : 49 to 61 Hz 230-V system 198 to 264 VAC: 49 to 61 Hz (2) Power consumption 1500 VA max (3) Output voltage Transformer tray OEM output Output voltage 1: 1 Trans-unit Total 250 VA (when 100 VAC is input) 360 VA (when 120 to 230 VAC is input) MONITOR output 200 VAC AC/DC tray PSA +48 *I VDC 9.5 A (SSA-770A) PSD +48 *I VDC 7.2 A FAN +46 *I VDC 0.8 A (for high-speed rotation) +40 k2 VDC 1 A (for low-speed rotation) Battery charging +5.6 f0.3 VDC IA ACRMVRS +5.0 f0.2 VDC 0.35 A ACRMMSV +5.0 f0.2 VDC 0.05 A (AC on)/0.70A (AC off) AC/DC-unit PSA +48 Yl VDC 11 Amax (SSA-700A) PSD +48 *I VDC 8Amax FAN +48 *I VDC 1.2 A Battery charged +5.6 f0.3 VDC IA ACRMVRS +5.0 f0.2 VDC 0.35 A ACRMMSV +5.0 f0.2 VDC 0.05 A (AC on)/0.70A (AC off) PSA unit M2.0AQ -2.10 VDC 20.0 A P3.3A +3.30 VDC 6.5 A P3.3AQ +3.45 VDC 23.0 A P5A +5.00 VDC 0.60 A P5AQ -5.00 VDC 5.20 A P7A +7.00 VDC l.lOA M9.6A -9.60 VDC 0.94 A P14AQ +I 4.2 VDC 1.50 A M14AQ -14.2 VDC 1.50 A PI OOHV +I 00 VDC 0.05 A Ml OOHV -100 VDC 0.05 A EBN -3.00 VDC 1.50 A Variable control voltage VTXO +0.125 to 18.0 VDC 20 Al40 w VTXI +0.125 to 18.0 VDC 20 A/40 w PSD unit P2.5D +2.575 VDC 24.2 A P3.3D +3.40 VDC 18.1 A P5D +5.10 VDC 15.7 A P12D +12.0 VDC 5.00 A M12D -12.0 VDC 3.30 A 6-2
[Page 158] 6.4 Transformer Tray This unit has the following functions. (a) Isolation from the commercial power supply (isolation transformer) (b) Input voltage switching (transformer tap changer) (c) Power supply output to the optional units (OEM AC output) (d) Power supply output to the monitor (P41) (e) Power supply output to the AC/DC tray (J51) (f) LED for error display (2) The figure shown below is the drawing of the transformer tray (SSA-770A) and Trans-unit (SSA-700A). <<Transformer tray= P41 monitor AC output OEM AC output Circuit protector for OEM AC output Circuit protector for monitor AC output J51 /AC output for LED for error display -\I1 the AC/DC tray 6-3
[Page 159] ccTrans-unit>> ~ P41 monitor AC output j~p J52 relay control input TRS LED control input J51 AC output for AC/DC tray Circuit protector for monitor AC output LED for error display Circuit protector for OEM AC output ZAUTION: Voltage setting for replacing the transformer tray and TRANS-unit The switches used to select the output voltage according to the input voltage are installed on the transformer tray and the TRANS-unit. When the switches of the replacement units (transformer tray and TRANS-unit) are not set correctly, the power supply may not start up correctly, or a malfunction of the connection equipment may result. Be sure to check the settings. For the transformer tray 100 to 120 VAC “115” must be displayed on Sl and S3. aIll;: The direction of S2 must be the same as that of Sl. 230 VAC “230” must be displayed on Sl and S3. czll~~ The direction of S2 must be the same as that of 53. For the TRANS-unit 100 to 120 VAC 115 ‘D “115” must be displayed. 115 kz “115” must be displayed. The direction must be the same as that of Sl and 52. El z 230 VAC “230” must be displayed. “230” must be displayed. The direction must be the same as that of Sl and 52. 6-4
[Page 160] (3) Description of each connector (a) OEM AC output Output voltage: Total of the three connectors 250 V (when 100 VAC is input) 360 V (when 120 to 230 VAC is input) NOTE: When the power supply of the system is turned ON, the relay in the transformer tray is turned ON by the relay control signal coming from the AC/DC tray and the AC output is also set to ON. (b) P41 monitor AC output PIN No. Signal 1 180 VAC to 220 VAC 2 1 180 VAC to 220 VAC 3 1 Protective grounding (GND) 41 - (c) J51 AC output for the AC/DC tray cln the case of the transformer tray> cln the case of the Trans-unit> PIN No. Signal PIN No. Signal 1 180 VAC to 220 VAC 1 78 VAC to 121 VAC 2 1 180 VAC to 220 VAC 2 78 VAC to 121 VAC 3 I 3 78 VAC to 121 VAC 4 1 Protective grounding (GND) 4 78 VAC to 121 VAC 5 Protective grounding (GND) 6 Protective grounding (GND) 6-5
[Page 161] 6.5 AC/DC Tray This unit has the following functions. (a) Power supply output to the PSA unit (J36) (b) Power supply output to the PSD unit (J37) (c) Power supply output to each unit fan (J31 and J32) (d) Processing of error signals from each of the following units and error signal output to the system (J34) l PSA unit l PSD unit l Each fan sensor circuit (e) Backup battery input for the memory of the RM board (J40) (2) The figure shown below is the drawing of the AC/DC tray. (SSA-770A) NM31 -2013E*I or *G or before BSM31-2013E*H or *J or later (3) The figure shown below is the drawing of the AC/DC-unit. J31 Test pin J40 J34 J32 J37 J36 6-6
[Page 162] 6.6 PSAIPSD Unit The PSA/PSD unit is a DC/DC conversion-type multiple output power supply unit. This unit generates and supplies the DC power used to operate the system and the peripheral devices such as the system E/M (Electrical Module: electronic circuit board module), main panel, HDD, and FDD using the 48 VDC output from the AC/DC tray. The PSA/PSD unit consists of two modules: the analog power supply unit PSA (Power Supply Analog), which supplies power to the analog front-end circuitry, and the digital power supply unit PSD (Power Supply Digital), which supplies power to the digital back-end circuitry and the peripheral devices. (2) Specifications (a) PSA This is a multiple output power supply, including 12 fixed voltage outputs and 2 variable voltage outputs (14 outputs in total). The output power (with the suffix “AQ”) is supplied to the circuit handling echo signals in the system (such as the preamplifier) and therefore must have as low noise as possible. The variable output powers VTXO and VTXI (voltages for transmission) are used for transmission driving of transducers in the Aplio system and control the acoustic power emitted from the transducer directly. The Aplio system uses the dual pulser method and has two similar transmission drive pulsers. The two variable output powers VTXO and VTXI, which have the same output characteristics, are supplied to the drive pulsers. output output Supply direction output output Supply direction power name voltage circuit power name voltage circuit M2.0AQ -2.10 v Reception amplifier P14AQ -2.10 v Relay drive power etc. supply P3.3A +3.30 v F/E 3.3-V logic circuit M14AQ +3.30 v CB board -12-V etc. circuit etc. P3.3AQ +3.45 v Reception amplifier PI OOHV +100 v HVSW plus power etc. supply P5A +5.00 v F/E D/A circuit etc. MIOOHV -100 v HVSW minus power supply P5AQ +5.00 v Transmission EBN -3.00 v EchoBooster power predriver circuit etc. supply P7A +7.00 v CB board +5-V circuit VTXO +0.125 Pulser drive power to 18.0 V supply M9.6A -9.60 V CB board -5-V circuit VTXI +0.125 Pulser drive power to 18.0 V supply 6-7
[Page 163] (b) PSD This is a multiple output power supply, including 5 fixed voltage outputs. Some voltages are supplied to the digital circuit of the system back-end and others are supplied to the audio amplifier circuit and main panel. output output Supply direction output output Supply direction power name I voltage I circuit I power name I voltage I circuit P2.5D +2.65 V Rambus, BE-FPGA P12D +12.0 V Main panel and audio circuit etc. Audio amplifier power P3.3D +3.40 V AGP card, BE system M12D -12.0 v supply I/F, FPGA etc. P5D +5.10 V CPU, video DAC I I circuit etc. I I I (c) Control I/F The analog differential signal CVTX for controlling the VTX voltage is supplied to the PSA through the Backplane board from the RC board. Also, the ON/OFF control signal is supplied to the PSX from the AC tray and the PSX status signal I/F is supplied to the AC tray from the PSX. (3) Operational check (a) Precautions for operational checks Perform the operational check paying attention to the following points a. Do not touch the input 48-V line. An electric shock may result. (The PSA/PSD operates with 48 VI800 W.) b. Do not touch the output connectors during operation. An electric shock may result. (The PSA has outputs off 100 V.) c. Be careful when handling parts that may have a residual charge even after the power supply is turned OFF. An electric shock may result. (The residual charge in the output capacitors cannot be discharged immediately.) d. Do not disassemble units such as the PSX box by removing the cover etc. (The PSX is replaced as a unit.) 6-8
[Page 164] (b) Test point location Confirm the voltage at each output test point/GND test point using the tester. Be careful not allow the tester probe to cause a short circuit by creating a connection between the live line of the output and the metal cover, because the cover plate is connected to GND. (Even if the tester probe on the live line of the output comes into contact with the grounded metal cover, the output will not be short-circuited to the ground because an in- line hiah resistance of 4.7 k ohm is placed in series (with the live line of the output) insidethe PSX.) PSD PSA VTXO lol ;ND P14AQ P5AQ 0 vll2D EBN 0 ‘12D M9.6A 0 ‘5D M2.0AQ 0 ‘3.3D MIOOHV 1 0 ‘2.5D M14AQ VTXI P7A P3.3A 0 P5A P3.3AQ PI OOHV GND lol (4) Test point voltage in the normal status Output power When the unit is mounted on Output power When the unit is mounted on name in the PSA the system [rating * (5% + name in the PSA the system [rating * (5% + total fluctuation)] total fluctuation)] M2.0AQ I -1.99 to -2.61 V 1 P14AQ I +12.7 to +15.7 V P3.3A +2.93 to +3.67 V M14AQ -12.7 to -15.7 V P3.3AQ +3.17 to +3.78 V PI OOHV +3.59 to +4.15 v P5A +4.50 to +5.50 v MIOOHV +6.97 to +8.03 V P5AQ I +4.50 to +5.50 v I EBN I +2.65 to +3.35 V P7A +6.30 to +7.70 V VTXO Confirm that they change M9.6A -8.64 to +I 0.6 V VTXI according to the AOP setting. P2.5D +2.34 to +2.91 V P12D +10.9 to +13.1 v P3.3D +3.13 to +3.87 V M12D -10.9 to to -13.1 v P5D +4.69 to +5.61 V 6-9
[Page 165] (5) Block diagram (a) PSA block diagram Status Outaut PSA ENH Main Over-Current Switcher OverTemperature IC.501 Main Rect. 010 Dll H ;;t;; L-Q Switching : : : : Transformer D14 Filter Linear Reg. : : . : Tl L6 L7 - ICI : 8 : EBN : : 48\ I : : : : INPL JT D16 Filter Linear Reg. i i - IIIC --FE- L8 L9 - IC2 : :a : Pl4AQ 0 : : : : : : D18 ; . : m : PIOOHV : : : : : : -1-111 Bridge - D20 L12 Filter L13 LinegReg. i i - - ‘base-shift - : * : MIOOHV : : soft Switcher : : II-Q4 T2 Filter : : L32 L33 P7A P5A P5AQ T3 Ma9p D37 D39 Filter Linear Re - & - + - L36-L38 Q12 Qli n llIc- Mac#gw D45 D46 Filter & -+- L41-L43 T401 Internal Use Supplies PVCC PVEE D405 Filter Linear Reg. 4 A n L402 L403 - IC402 T I J” L 1 1 I D4$ 1 1 Filter L404 L406 --- VTXI r Pre-Regulator WX VOLTAGE CVTXI H/CVTXl L (Differential Analog Control Input) IC801 IC804 = 9.00 x CVTX 0 m CVTXOH/CVTXOL (Differential Analog Control Input) VTXO I a Pre-Regulator IC802 IC808 6-10
[Page 166] (b) PSD block diagram Status Output PSD-ENH Overvoltage UnderVoltage Main OverCurrent Switcher ’ OverTemperature IC21 :: : : Filter i i L22 L23 L28 : : l P3.3D i ; 48V INPUT I,,&/,! ’ :ge,;, L27 1 I j .M,2D Full Bridge Phase-shift Soft Switcher Q2-Q5 AUX 1 Tl Internal Use Supply pycc Switcher 1 Q6 - - Filter - Linear Reg. Lu L6 L8 IC2 6-11
[Page 167] 6.7 LEDs for Error Display The eight LEDs on the back of the transfer tray are used to indicate abnormalities and errors as shown below. For the procedure for identifying faults in the power system, refer to the service manual (maintenance volume). LED 2 LED B PSD Status BATTERY Status LED 5 LED A PSAIPSD FAN Status INPUT AC Status System power OFF overload) 48 V-Internal Fault OFF OFF Blinking Blinking Blinking Green ON - Green ON (AC/DC power (yellow) (green) (green) supply abnormality) AC failed or missing OFF OFF OFF Blinking Blinking OFF - OFF (AC/DC output (green) (green) voltage drop) Battery - - - - - - OFF - disconnection Battery charging - - - - - - Blinking - (green) Battery charging - - - - - - Green ON - cornDIeted 6-12
[Page 168] 7. MONITOR 7.1 CRT Monitor (1) Specification Line voltage 90Vto264V Power consumption 1.2 A to 0.6 A Monitor size 17-inch (300 mm x 225 mm) Dot pitch 0.25 mm Input video signal R, G, and B separated signals, 0.714 Vp-p, 75 R Input synchronization signal H-SYNC, V-SYNC separation, TTL level Display resolution 800 x 600,60 Hz/75 Hz Speaker 2 tweeters, 2 woofers (no audio signal processing in the monitor Mass : Approximately 24 kg A . CAUTION: The monitor is a heavy unit (20 kg or more). Two persons are required for removing and replacing the monitor. Otherwise, personal injury may result or the monitor may be damaged. (2) Monitor adjustment The brightness and contrast can be adjusted using the corresponding buttons on the front of the monitor. (3) Self-test pattern and degaussing function Pinhole buttons for test pattern display and degaussing are provided in the front part of the bottom of the monitor. (An implement with a sharp tip is required to press these buttons.) Left: Test button Right: Degaussing button Test pattern display is selected when the Test button is pressed and the pattern changes when the Degaussing button is pressed. It is possible to return to the original pattern by pressing the Degaussing button several times. 7-l
[Page 169] (4) OSD adjustment menu (a) Starting and terminating the menu Hold down the Test button on the bottom of the monitor and turn ON the breaker. Press the POWER switch to startup the system and then release the Test button. When the Degaussing button is pressed in this status, the adjustment menu shown below is displayed. The menu can be closed by selecting “EXIT 0SD”from MAIN MENU. The menu can be displayed again by pressing the Degaussing button again. When setting is completed, press the POWER switch of the system and then turn OFF the breaker. (b) Operating the menu The Degaussing button and the + and - switches for contrast and brightness are used for menu operation. The selection cursor is moved by pressing the + and - switches, and submenus are displayed by pressing the Degaussing button GEOMETRY COLOR TEMPERATURE ADVANCED CONTROLS MONITOR STATUS DEGAUSS EXIT OSD SERVICE MENU Degaussing button ON VERTICAL POSITION VERTICAL SIZE q SAVE & RETURN TO MAIN MENU 7-2
[Page 170] The detailed functions are displayed in submenus. The selection cursor is moved by pressing the + and - switches, and the adjustment menus for the functions are displayed by pressing the Degaussing button. The values in the adjustment menu are changed by pressing the + and - switches. To set a value, press the Degaussing button. After a value is set, the submenu closes and the next adjustment can be performed When “SAVE & RETURN TO MAIN MENU” is selected, the changes are set and the MAIN MENU is displayed. Note that when “RECALL” is selected. the changes are deleted and the MAIN MENU is displayed. I SIZE & POSITION HORIZONTAL SIZE VERTICAL POSITION VERTICAL SIZE - q SAVE & RETURN TO MAIN MENU - q RECALL Degaussing button Degaussing button Degaussing buttor The value is changed by pressing the + and -switches. GEOMETRY COLOR TEMPERATURE ADVANCED CONTROLS MONITOR STATUS DEGAUSS EXIT OSD SERVICE MENU 7-3
[Page 171] (c) Detailed explanation of each submenu The functions in each submenu are described below. 1) SIZE & POSITION The display position and range are adjusted. HORIZONTAL POSITION 1Adjustment of the display position in the horizontal direction HORIZONTAL SIZE Adjustment of the display range in the horizontal direction VERTICAL POSITION Adjustment of the display position in the vertical direction VERTICAL SIZE I Adjustment of the display range in the vertical direction 2) GEOMETRY The display distortion and skewing are adjusted. ROTATION 1Adjustment of skew in the display PINCUSHION Adjustment of barrel and pincushion distortions TRAPEZOID Adjustment of the trapezoidal distortion BALANCED PINCUSHION Adjustment of the balance of barrel and pincushion distortions PARALLELOGRAM Adjustment of parallelogram distortion 3) COLOR TEMPERATURE The color temperature is set. Setting of the temperature to 9300 K 1Setting of the temperature to 11000 K 13000 Setting of the temperature to 13000 K SERVICE The R, G, and B values can be set. CAUTION: The image quality adjustment for each transducer is performed at 13000 K. Therefore, make sure that the color temperature is not changed inadvertently. 4) ADVANCED CONTROLS Special functions of the monitor POWER SAVING ON Setting power saving mode to ON or OFF LANGUAGE Selecting the language to be used in the menus OSD TIMER I Thus function is not available for this monitor. Moire adjustment 7-4
[Page 172] 5) MONITOR STATUS The monitor status, such as the synchronization mode, is displayed. 6) DEGAUSS Degaussing is performed. 7) EXIT OSD The adjustment menu is terminated. 8) SERVICE The basic setting screen is displayed. CAUTION: Do not change items in SERVICE. Doing so may change basic settings such as the color temperature. 7.2 LCD Monitor (1) Specification Line voltage : 90Vto264V Power consumption : 45 W or less Monitor size : 17-inch Dot pitch : 0.264 mm x 0.264 mm Input signal : R, G, and B separated signals, 0.714 Vp-p, 75 R H-SYNC, V-SYNC separation, TTL level Maximum resolution : 1280 x 1024,60 Hz/75 Hz Speaker : x 2 (no audio signal processing in the monitor) Mass : Approximately 4.2 kg 7-5
[Page 173] (2) Adjustment menu (a) User mode The menu switches 0 0 on the front of the LCD monitor are used to display the user V@lA menu and to select menu items. The [-] switch and [+] switch are used to select the brightness, contrast, screen size, and y curve. User menu Menu item I Default value I Adjustment range Brightness I 30% I 0% to 100% Contrast I 100% I 0% to 100% Size I 1 I 1I213 2 1 I21314 * To reset the setting to the default value, press the [-] switch while holding down the [+] switch. (b) Service mode 1) Opening and closing the service mode menu Hold down both the [V] switch and the [+] switch on the front of the LCD monitor for 3 seconds to display the service mode menu (main menu). To close the service mode menu, select the exit key @ and press the [V] switch. VOA - + Exit key Menu switches [-] switch [+] switch 2) Menu operation Use the [+] and [-] switches for menu operation Press the [+] or [-] switch to move the selection cursor to the desired function and then press the [V] switch to set the selection. The submenu for the selected function is displayed to show more detailed functions. Use [+] or [-] to select the desired function and then press [VI. The adjustment menu with an icon is displayed. Change the value using [+] or [-] and then press [V] to set the new value. 7-6
[Page 174] MENU [VI <Brightness> Use [+] or [-] to change the value. MENU [VI 7-7
[Page 175] 3) List of functions available in the service mode menu Menu list Main menu Function Screen Color Other Information Language 7-8
[Page 176] 4) Details of the service mode menu l Screen Auto Adjustment: Used to automatically adjust flickering, bleeding, the image display position, and other screen conditions. * Be sure to use the adjustment guide program (loadadj.exe) for automatic adjustment. If automatic adjustment is not successful, use the following individual menu items. Clock : Used to eliminate vertical stripes on the image. Phase : Used to eliminate bleeding, flickering, or horizontal stripes if necessary after clock adjustment. Position : Used to adjust the image position. Resolution : Used to adjust the resolution when a part of the image is not displayed or unnecessary parts are displayed. Range : Used to adjust the signal output level so that all color levels are displayed. (0 to 255). Smoothing : Used to adjust smoothness of the characters and lines of the image. (5 steps, soft to sharp) Signal Filter : Used to eliminate image overlap or a bright vertical line on the right of the image. CAUTION: When the LCD monitor, RM PWB, VI PWB, or IO PWB has been replaced, be sure to run the Auto Adjustment program and the Range program to tune the signals for the diagnostic ultrasound system (only in the case of analog signal input). Be sure to quit service mode after the service mode menu is used. Otherwise, automatic power save mode is not started. 0 Color Brightness : Used to adjust the screen brightness. (0% to 100%) Contrast : Used to adjust the image contrast. (0% to 100%) Temperature : Used to select the color temperature. (1, 2, Off) Color temperature 1 is higher than 2. * The value is set to OFF automatically when the gain value is changed. Gamma : Used to set the y value. (1, 2, 3, 4, Fixed) Saturation : Used to adjust the saturation (vividness of colors). (-100 to 100) * The entire color scale may not be displayed when this function is used. Hue : Used to adjust the hue (coloring). (-100 to 100) * The entire color scale may not be displayed when this function is used. Gain : Used to adjust red, green, and blue individually to obtain the desired color. (0% to 100%) * Gain adjustment is disabled when the Temperature value is changed. Reset : Used to reset all color settings to the default values 7-9
[Page 177] l Other Screen Size : Used to change the image display size. (1, 2, 3) 1: 17 inches (full-size) 2: 16 inches 3: 15 inches (equivalent to the display size of the 17-inch CRT monitor) Border Intensity : Used to adjust the brightness of the borders (areas where no image is displayed). (0 (black) to 255 (white)) Menu Position : Used to adjust the menu display position. (Any position on the screen) Use the [+] or [-] switch to adjust the position. The adjustment direction (horizontal or vertical) can be selected using the [A] switch. Menu Off Timer : Used to adjust the time for which the menu is displayed (5 s, 10 s, 25 s, 50 s) Reset : Used to reset all settings to the default values 0 Information Information (l/2): Displays the currently selected input signal, resolution, horizontal frequency, and vertical frequency. Information (2/2) : Displays the serial number, hours of use, and firmware version of the LCD monitor. 0 Language Language : Used to select the language for menus. Power save mode 1) Forced power save mode Hold down both the [A] switch and the [-] switch for approximately 1 second to forcibly set the monitor to power save mode (backlight off). To cancel power save mode, hold down both the [A] switch and the [-] switch for approximately 3 seconds. Forced power save mode remains set after the power switch on the operating panel is turned OFF unless it is canceled as described above. 2) Automatic power save mode The monitor is set to power save mode (backlight off) automatically when no signal is input for 5 seconds or more. 7-10
[Page 178] (3) Status LED The LED at the lower front of the LCD monitor indicates the monitor status. LED indication I Status OFF AC power OFF Blinking green (ON/OFF at 2-s intervals) Forced power save mode Lights green 1 Startup, display ON Lights orange Automatic power save mode (No signal) Blinking orange (ON/OFF at 2-s intervals) Signal error Blinking orange (ON/OFF at l-s intervals) Automatic power save mode (Abnormal temperature inside the LCD monitor) (4) Starting up the adjustment guide file and adjusting image quality (a) Double-click “C:\LCD\Loadadj.exe”. (b) Confirm that the Auto Adjustment guide screen is started up. (c) Hold down both the [Y] switch and the [+] switch on the front of the LCD monitor for 3 seconds to display the service mode menu (main menu). 7-11
[Page 179] (d) Follow the instructions on the Auto Adjustment guide screen to start up the Auto Adjustment program. Press [A]. Press [v]. Auto Adjustment is started up. After the Auto Adjustment program is started up, the OSD menu icons are displayed in the following order. When adjustment is completed, the display returns to the Screen menu. 7-12
[Page 180] W To eliminate vertical stripes on the screen Click the [Next] button on the Auto Adjustment guide screen. If vertical stripes are not displayed on the screen, proceed to step (f). Following the instructions on the Clock Adjustment guide screen, select the “Clock” icon using the [A], [VI, [-I, and [+] switches and then press the [V] switch to display the clock adjustment menu. Adjust the sampling clock rate so that the vertical stripes disappear. When adjustment is completed, press the [V] switch to save the new value and return to the Screen menu. Clock adjustment menu (9 To eliminate flickering or bleeding Click the [Next] button in the Clock Adjustment guide screen shown in (e). If there is no screen flickering or bleeding, proceed to step (g). Following the instructions on the Phase Adjustment guide screen, select the “Phase” icon using the [A], [VI, [-I, and [+] switches and then press the [V] switch to display the phase adjustment menu. Adjust the phase until flickering or bleeding is eliminated. When adjustment is completed, press the [V] switch to save the new value and return to the Screen menu. 7-13
[Page 181] Phase adjustment menu (g) To adjust the display position Click the [Next] button in the Phase Adjustment guide screen shown in (f). If there is no problem with the display position, proceed to step (h). Following the instructions on the Position Adjustment guide screen, select the “Position” icon using the [A], [VI, [-I, and [+] switches and then press the [v] switch to display the position adjustment menu. Use the [A] switch to select vertical adjustment or horizontal adjustment, and then move the display position using the [-] or [+] switch. When adjustment is completed, press the [v] switch to save the new setting and return to the Screen menu. Press [v]. Horizontal position adjustment menu Press [&J &ss [*]. After selecting horizontal or vertical adjustment, adjust the posrtron usrng Vertical position adjustment menu 7-14
[Page 182] (h) Click the [Next] button in the Position Adjustment guide screen shown in (g). Following the instructions on the Adjust the Range automatically screen, select the “Range Adjustment” icon using the [A], [VI, [-I, and [+] switches and then press the [V] switch to display the range adjustment menu. Use the [-] or [+] switch to select “Execute” and then press [V] to start up the Range Adjustment function. The OSD menu icons are displayed in the order shown below. When adjustment is completed, the display returns to the Screen menu. Press [v]. Press [A]. Press [v]. Range Adjustment is executed. After Range Adjustment is started up, the OSD menu icons are displayed in the order shown below. When adjustment is completed, the display returns to the Screen menu. 7-15
[Page 183] (i) Close the service mode menu. Press [-] or [+I. Press [v]. The service mode menu closes. Click the [Close] button on the Adjust the Range Automatically screen shown in (h). (j) Select [Start] + [Shutdown...] + [Shutdown] to shut down the system. (k) Restart the system and confirm that there are no abnormalities in LCD monitor display. CAUTION: This image quality adjustment procedure is applicable only to the U ILM-770A/700A. Auto Adjustment and Range adjustment must be performed without fail. 7-16
[Page 184] 8. PHYSIO MODULE This module detects physiological signals (ECG, respiration, PCG, and pulse) using the ECG cable and sensors and transfers the data to the diagnostic ultrasound system (hereinafter referred to as “system”). Line voltage ECG (electrocardiogram) ICH (IM RESP (respiration) ICH) RESP (respiration) ICH DC IN (external inpuffECG) ICH (line voltage: f2.5 V) PCG (phonocardiogram) ICH Pulse (pulse) ICH AUX (external input) 1 CH (line voltage: f2.5 V) (2) Functions The following functions are provided. (4 Amplification of the electronic potential induced by the R and F electrodes of the ECG cable attached on the body surface and transfer of the data to the system in order to display the ECG on the monitor. Detection of the QRS waveforms and transfer of the signals to the system. Measurement of the electrical resistance in the chest with an alternating current (between electrodes R and F) and transfer of the data to the system in order to monitor the respiration cycle. (b) Transfer of respiration data from the thermistor sensor attached to the nasal cavity to the system in order to display the respiration cycle on the monitor. (cl Transfer of data from the PCG sensor attached to the chest to the system in order to display PCG waveforms on the monitor. (4 Transfer of data from the waveform sensor attached to the carotid artery and apex of the heart to the system in order to display pulse waveforms on the monitor. (e) Transfer of input signals (physiological signals) from two external medical devices to the system. The input signal from the one of the external medical devices can be output to the system as a trigger signal for imaging. (3) Operations This module consists of the amplifier section, power supply section, and digital section. The module is connected to the system via a cable and a 15pin D-sub connector. The module receives a 12-V power supply and waveform request command from the system and performs waveform (such as ECG and PCG) output to the system. The amplifier which amplifies the ECG, respiration, PCG, pulse, and external device signals is set inside this module. After importing the A/D values, the module calculates the sensitivity and position using the single-chip CPU inside the A/D converter and stores them in the RAM as digital values. QRS detection is performed by the hardware. The waveform data is output via a serial interface at a transmission rate of 312.5 Kbaud when a waveform request command is received from the system. QRS detection signals are output by setting QRS flags in the detection signals and waveform data (two series). Also, sensor connector insertion/removal information is transferred to the system. For the ECG connector, lead ON/OFF information, which indicates whether or not the three electrodes are attached to the patient properly, is output. 8-l
[Page 185] 8-2
[Page 186] 9. CONNECTION DRAWING FOR UNITS --------_____
[Page 187] Drawing Drawing No. number for number for Connection points SSA-770A SSA-700A 1 CM30-30335 CM30-32448 Audio IO to VCR 2 CM30-30372 CM30-32449 Audio VCR to IO 3 CM30-30340 CM30-30340 Ethernet IO to Rear 4 CM30-30337 CM30-30337 USB IO to Rear 5 CM30-30363 CM30-30363 USB IO to Rear 6 CM30-30338 CM30-30338 C.Video OUT IO (BNC) + Rear (BNC) 7 CM30-30333 CM30-32452 C.Video OUT Rear (BNC) to VCRin (Pin Plug) 8 CM30-30334 CM30-30334 Y/C OUT: IO to Rear 9 CM30-30392 CM30-32446 Y/C OUT: Rear to VCRin 10 CM30-30347 CM30-30347 Rear Panel Reset SW 11 CM30-30373 CM30-32447 Y/C PlayBack: VCRout to IO 12 BSM31-2541 BSM31-3931 C.VideoPlayBack: VCRout-IO Yin 13 CM30-30336 CM30-32450 VCR RemoteControl: Panasonic AG-MD835:9pin-25pin Straight 14 CM30-30415 CM30-32451 VCR RemoteControl: SONY SVO-9500 :9pin-25pin Cross 15 BSM31-0671 BSM31-1907 OEM AC Cable (VCR) 16 BSM31-0671 BSM31-0671 OEM AC Cable (Color Printer) 17 BSM31-3757 BSM31-3757 ColorPrinter Data (USB-USB) 18 CM30-30925 CM30-30925 B&W Printer Data (USB-USB) 19 CM30-30927 CM30-30927 OEM AC Cable (B&W Printer) 20 CM30-30304 CM30-30304 MonitorVideo 21 CM30-30303 CM30-30303 MonitorAudio 22 CM30-30302 CM30-30302 MonitorPower 23 CM30-30354 CM30-30354 VCR RemoteControl: SONY DSR-20:9pin-9pin Cross 24 CM30-30413 CM30-30413 Reset IO to Rear 25 CM30-30311 CM30-30311 AC-DC to IO control 26 CM30-30309 CM30-30309 AC-DC to PSA/PSD, RDD, IO Fan Connection Cable 27 CM30-30316 CM30-30316 Cardcage Cooling Fan 28 PM30-29870 PM30-29870 PSA/PSD Cooling Fan 29 CM30-30314 CM30-30314 IO Cooling Fan 30 CM30-30312 CM30-30312 RDD Cooling Fan 31 CM30-30328 CM30-30328 SCSI (White) 32 CM30-30327 CM30-30327 +5/12V Power Cable in FrontPanelBoard 33 CM30-30342 CM30-30342 SCSI Cable in FrontPanelBoard 34 CM30-30414 CM30-30414 Reset Signal Cable in IO Board 35 CM30-30330 CM30-30330 SCSI POWER CABLE (Black) 36 CM30-30394 CM30-30394 Mic., Shutter, Headphone cable in Front Panel Board 37 PM30-29980 PM30-29980 FOOT-SW CABLE 38 CM30-30306 CM30-30306 ControlPanel-FrontPanel 39 CM30-30317 CM30-30317 ControlPanel- USB DATA CABLE 40 CM30-30305 CM30-30305 ControlPanel- Power D-Sub CABLE 41 CM30-30331 CM30-30331 PHYSIO to IO 42 PM30-30000 PM30-30000 Pencil Probe Panel - TI Board 43 CM30-30351 CM30-30351 Pencil Probe Signal Cable in TI Board 44 CM30-30308 CM30-30308 HDD Power Cable 45 CM30-30307 CM30-30307 HDD Data Cable 46 BSM31-2587 BSM31-2587 AC POWER CABLE (100 V) 47 BSM31-2588 BSM31-2588 AC POWER CABLE (200 V) 9-2
[Page 188] 10. SOFTWARE ORGANIZATION 10.1 Overall Folder Configuration c: WINNT etc. Windows2000 standard folders Zdt Zdt installation destination Note that zdtexe is located in WINNT\system32. Temp Used as a work area for database system etc. A minimum of 1 GB or more is required. Maintenance including defragmentation of the C drive etc. is important. D: aplio Destination for Aplio system/application software aplio\bin Destination for executable binary files such as dll, ocx, and exe files aplio\config Configuration at the time of shipment and destination for activated options aplio\user\(login name) Destination for files defined by the user aplio\TUSService Destination for the service software The InnerVision contract file is located in aplio\TUSService\lVLicence. The service contract file is located in aplio\TUSService\ServiceLicence. aplio\TUSLog Destination for log file for application software database Database E: Image Destination for actual image files PIMS PIMS work area DDGW DDGW (option) work area IO-I
[Page 189] 10.2 Version Information The Aplio software version and the service software version are always displayed on the upper right of the service screen as shown below. For details, refer to the service manual (maintenance volume). 1o-2
[Page 190] 10.3 Error Handling 10.3.1 Error log operation Error log operation (error registration) is performed in the background as required. The recording destination is the same as for the application log of Windows. The data is located in C:\WINNT\system32\configMppEvent,Evt. The logged information can be checked using the log display function in the service menu (service menu main screen + Log). The user is notified of 5 types of errors as explained in subsection 10.3.4 “Error log operation (facility)“. The errors are registered and error messages are displayed in the dialog box or status bar. 1 o-3
[Page 191] 10.3.2 Error messages Each error message consists of the elements shown in table below. Item Contents Symbol Symbol name to be referred to from applications Severity Importance of error MessageText 1 Simple description of error (title) Facility Error log and operation Explanation Detailed description of error. Referred to from the Syngo error viewer. Action 1 Corrective action for error. Referred to from the Syngo error viewer. Information Additional information. Referred to from the Syngo error viewer 10.3.3 Error levels (severity) There are four error levels (same as Windows Native errors). Severity Meaning Error (STATUS-SEVIRITY-ERROR) An unrecoverable error has occurred. Warning (STATUS-SEVIRITY-WARNING) A recoverable error has occurred. Information (STATUS~SEVIRITY~INFORMATION) Processing has been completed. However, there is additional information. Success (STATUS-SEVIRITY-SUCCESS) 1 Processing has been completed. 1 o-4
[Page 192] Error log operation (facility) Target to be Operation notified FACILITY-MESSAGEBOX-DANGER Service engineer After an error message is and user displayed in the danger message box, the contents of the error are registered in the log. FACILITY-MESSAGEBOX-WARNING Service engineer After an error message is and user displayed in the warning message box, the contents of the error are registered in the log. FACILITY-MESSAGEBOX-CAUTION Service engineer After an error message is and user displayed in the caution message box, the contents of the error are registered in the log. FACILITY-MESSAGEBOX-NORMAL Service engineer After a message is displayed in the and user message box, the contents of the error are registered in the log. FACILITY-STATUSBAR Service engineer After a message is displayed in the and user status bar, the contents of the error are registered in the log. FACILITY-RELEVANT Service engineer The contents of the error are registered in the log. FACILITY-CSA Engineer -- 1 o-5
[Page 193] 11. SETTING THE PRESET MENU 11 .I Outline of Each Preset (1) Basic preset architecture This system contains the following presets. l System Preset l Exam Type (This is actually a system preset. However, it is important and is therefore included here also.) l Application Preset l Imaging Preset l 3D Preset (2) System preset System Preset includes a group of parameters to be set in the system. Some of the parameters included in System Preset are available only to the service engineer, but the parameters are generally available to the user. The user can change the parameters using the system preset editor. (3) Exam Type (a system preset) Exam Type is used to set the transducer, imaging preset, and application preset automatically. In this system, the Exam Type can be completely matched with the imaging preset name. When the Exam Type is selected on the Patient Information screen and the examination is started, the selected Exam Type is registered in the DICOM format. Using the Editor screen, the item (transducer, imaging preset, or application preset) to be activated when Exam Type is selected on the Patient Information screen can be set. (4) Application preset The application preset includes parameters such as Measurement, Body Mark, Annotation, and Report that are available to the user. The user can change the settings using the application preset editor. The application preset values at the time of shipment can be returned to those at the time of shipment at any time even if the values have been changed. Note, however, that if the values are returned to those at the time of shipment, the settings specified by the user are deleted. In order to prevent this information from being deleted, save it to external storage media. 1 l-l
[Page 194] (5) Imaging preset The imaging preset includes a group of parameters related to imaging. They are available to the user. However, there is no preset editor. User setting can be performed using the Save function after selecting the imaging parameter. Up to seven imaging presets can be registered for each Exam Type by the user and can be deleted and overwritten. Note, however, that the factory setting that is contained in each Exam Type cannot be deleted or overwritten. Imaging parameters can be preset independently for Fund., THI, CHI, Color, Power, etc. The 2D-IP is equivalent to B-IP in the previous system. There are five IP Types (A to E) and eight IP Indexes (1 to 8). Parameters such as the dynamic range value in 2D-IP are stored separately (in a separate file) from the imaging preset parameters. (6) 3D preset The 3D Preset consists of the following three presets. Although these presets have been set at the factory, they can be changed by the user. These presets can be set for each Exam Type and each Volume Type. l Initial display preset 0 Operation preset l Reconstruction preset 11-2
[Page 195] Testes MSK 2D-IP Measure Annotation B Image Image (User) (User) Preset 1 Preset 8 - 1 C (Userl) (User 4bdomen 4bdol nr ml Annotation Image Image Image (User) Preset Preset 1 Preset 8 v System (Factory) (Userl) (User@ I Preset 1 (User) Probe Info. System Preset (Factory) Figure 1 l-l Preset structure
[Page 196] Exam Type Imaging Preset Sub-mode CDI Parameters Adult Heart - Factory Setting -CDI Mode Filter : High Scale: High MAP : V/T1 Power Mode Filter : High Scale: High MAP : PI TDI Mode Jserl OB Factory Setting - CDI Mode Filter: High (Note that the CDI parameters are only examples and not recommended values.) Figure 1 l-2 Imaging preset structure 11-4
[Page 197] 11.2 Setting and Changing Presets (1) Preset selection on the Patient Information screen Exam Type Adult Heart The user selects an Exam Type on the Patient Information screen. Usually, the Exam Type used for the previous patient is displayed initially on the Patient Information screen. However, if an Exam Type has been saved using Application Preset Editor, the saved Exam Type is displayed as the default. If the displayed Exam Type is acceptable, click the [CONFIRM Start] button. To change the displayed Exam Type, select the desired Exam Type from the drop-down menu and click the [CONFIRM Start] button. The transducer, imaging preset, and application preset are set automatically according to the selected Exam Type. Note that the Exam Type selected on the Patient Information screen is also set as the DICOM Exam Type. 11-5
[Page 198] . Exam Type Preset Setting When an Exam Type is selected, the corresponding Transducer, Imaging Preset, and Application Preset are set automatically. Note: This table may be modified in the future without prior notice. 11-6
[Page 199] The meaning of each Exam Type (24 total) is given below for reference. 1) Abdomen General abdomen 2) Carotid Carotid artery 3) Thyroid Thyroid gland 4) Breast Mammary gland 5) OB Obstetrics 6) GYN Gynecology 7) Endo-vaginal Endovaginal 8) Fetal Heart Fetal circulatory system 9) Adult Heart General circulatory system 10) Pediatric Heart Pediatric circulatory system 11) Coronary Coronary 12) TCD Transcranial region 13) Neo-Head Neonatal head 14) Neo-General General neonatal abdomen 15) Neo-Hip Neonatal hip 16) PV-Venous PV vein 17) PV-Arterial PV artery 18) Digits Peripheral vessels (fingers) 19) MSK Orthopedics 20) Prostate Prostate 21) Kidney Kidney 22) Testes Testes 23) OTHER Other 24) M-TEE Transesophageal 11-7
[Page 200] l Application Preset Setting The measurements, body marks, and annotations that can be selected in each application preset are given below. Applicable region Note: This table may be modified in the future without prior notice. (*I): Only the arrow annotation is displayed on the touch panel (“2): No measurement items are displayed on the touch panel. 11-8
[Page 201] (2) Changing the imaging preset or application preset during an examination To change the imaging preset or application preset after clicking the [Start] button on the Patient Information screen, press the [Preset] switch on the touch panel, select the Imaging Preset page or Application Preset page by pressing the corresponding tab, and select the desired preset. l When the imaging preset is changed A dialog is displayed. The application preset and the transducer are not affected by the change of the imaging preset. l When the application preset is changed No dialog is displayed. The preset is changed directly. The imaging preset and the transducer are not affected by the change of the application preset. * If previous patient information is loaded on the Patient Information screen and the Exam Type is changed before the [Start] button is clicked, the imaging preset, application preset, and transducer setting are changed together accordingly. 11-9
[Page 202] 11.3 Preset Editors The system preset and application preset editors are described in this subsection. (There is no imaging preset editor because saving is performed on the touch panel.) (1) System Preset editor (not applicable to Exam Type) The system preset editor is opened by clicking “System Setting” in the popup menu. The submenu shown below is displayed on the monitor. Select an item in this submenu to open the editor. <<Popup menu>> II-IO
[Page 203] <<Example of the System Preset editor: “General Setting”>> . Functions and locations of the Save and Quit buttons (Common to all the pages of the System Preset Editor) Save : Saves the current settings. No confirmation message is displayed. To enable all of the settings, restart the system (restarting is not required for some setting items, however be sure to restart the system). Quit : Closes the editor screen. . Addition of preset items for service engineers (Common to all the pages of the System Preset Editor) When the service key ([Shift] + [<I) is pressed with the System Preset Editor screen displayed, the preset items for service engineers are displayed additionally. The preset items for service engineers disappear when the Editor screen is closed by pressing [Quit]. To display the preset items for service engineers on a different page of the System Preset Editor, press the service key ([Shift] + [c]) again. II-II
[Page 204] (2) Exam Type editor The Exam Type editor starts up when the Exam Type button in the popup menu is clicked. The following screen is displayed on the monitor. Saves the items currently selected in the Exam Type Application Preset area so that they will be displayed as the default. Resets the parameters Saves the parameters for the currently selected currently displayed in item to the factory settings. the Application Preset Resets all the parameter - Editor area. settings in the Exam Type Edit screen to the factory settings. Saves all the parameters for the item selected for Application Preset. At the time of shipment, each Exam type has only one Factory Setting. (The user can create presets.) Select the imaging preset, application preset, and transducer that correspond to the selected Exam Type and click [Save] at the right of the Exam Type field to set them. Exam Type : Can be selected from the items set at the time of shipment. Imaging Preset : The imaging preset related to the Exam Type can be selected. Presets defined by the user are also displayed here. If the Exam Type is changed, the imaging preset is also changed correspondingly. Application Preset: Can be selected from the list of all application presets. If the Exam Type is changed, the application preset is also changed accordingly. Probe Can be selected from all transducers. If the Exam Type is changed, the setting for Probe is also changed correspondingly. 11-12
[Page 205] (3) Application Preset editor This editor is started up in the same manner as the Exam Type editor in (2). Refer to the figure. To edit an application preset, select the desired application preset and also select the item group such as 2D-Meas. etc. Then set the parameters. Depending on the application preset, some item groups are not displayed (for example, OB Calc is displayed only when the OB application preset is selected). [Save] (located in the middle) Saves the parameters for each item group. [Save] (located at the bottom) Saves all of the parameters for all item groups (in particular, all of the parameters for the application presets) at one time. [Factory Setting] (located at the second position from the bottom) Returns the parameter settings to those at the time of shipment. [Factory Setting] (located at the bottom) Returns the parameter settings for all item groups to those at the time of shipment. 11-13
[Page 206] 11.4 Description of Preset Parameters The menu shown below is displayed in the popup menu. Example screens of each editor are shown. 11-14
[Page 207] (1) General settings TOSHIBA Organization Name The name set here is displayed in the ID area. The user can add items. Remove Used to delete an organization name from the list. System Location The hospital address etc. is entered here. (This is used in the maintenance menu.) Date The date is entered here. Time The time is entered here. Language (servicing) The language can be selected here (when multiple languages are supported). However, the type keyboard is not set here. KeyBoard (servicing) The keyboard-supported language can be selected. Height The unit used for the patient height on the Patient Information screen is set here. Weight The unit used for the patient weight on the Patient Information screen is set here. Data Format The format for the date displayed in the ID area is set here. Time Format The format of the time displayed in the ID area is set here. Trackball Sensitivity The sensitivity of the trackball is set here. 11-15
[Page 208] (2) Peripheral System Freeze at VIDEO PLAY Used to set whether or not the VCR is paused by pressing Freeze during VCR playback. Freeze at VIDEO REC Used to set whether or not the VCR is paused by pressing Freeze during VCR recording. VIDEO Counter Display Used to set whether or not the VCR counter is displayed. VIDEO Maker Used to set the VCR manufacturer. 11-16
[Page 209] (3) OUTPUT Title I Setting Clip Store Format The dynamic image recording format (AVI or DICOM) can be selected here. Sono Printer (B/W) (servicing) A B/W printer can be set here. The list includes the printers recommended by Toshiba. Son0 Printer (Color) (servicing) A color printer can be set here. The list includes the printers recommended by Toshiba. DICOM Printer (servicing) A DICOM printer can be set here. The list includes the DICOM printers set in the maintenance. The output is performed to the DICOM printer set here by operating the switches on the touch panel. DICOM Server (servicing) The DICOM server can be set here. The list includes the DICOM servers set in the maintenance. The output is performed to the DICOM server set here by operating the switches on the operating panel. Report Server (servicing) The server for DICOM SR can be set. This item is displayed only for the systems in which the DICOM option is installed. The parameters other than Clip Store Format are provided for use by service engineers only. 11-17
[Page 210] (4) Exam Information (System Preset) Operator Name Name of the operator. This is displayed on the Patient Information screen. The user can add items using the preset editor. Perform Physician Name of the physician who reads the ultrasound images. This is displayed on the Patient Information screen. The user can add items. Refer Physician Referring doctor (in the same hospital or satellite hospital). This is displayed on the Patient Information screen. The user can add items using the preset editor. Request Department Department requesting the examination. This is displayed on the Patient Information screen. The user can add items using the preset editor. Remove Used to delete items from the list using the preset editor. Emergency ID (servicing) A special character can be set as the end character of the ID for emergency patients. Unique character string : An arbitrary character can be set in the preset. Arbitrary character string : An arbitrary character can be set in the preset. Emergency Name (servicing) A special character can be set as the end character of the name for emergency patients. Unique character string : An arbitrary character can be set in the preset. Arbitrary character string : An arbitrary character can be set in the preset. 11-18
[Page 211] (5) Transducer M-TEE Pre-caution Used to set whether or not a warning message is displayed Message Display when the transducer surface temperature exceeds 42°C. .----------------------- .------------------------------------------------------------------------------- Icon Display Used to set whether or not the icon is displayed. .----------------------- Permission of Used to set whether or not the user is allowed to set the icon Icon Display display. Needle Adapter: Adapter Select Used to select the needle mark for PVT-770RT. PVT-770RT Toshiba Mark Placement Used to select the Toshiba mark (T mark) display position on the monitor. 11-19
[Page 212] (6) External Alpha-Protocol 1 COM Port 1 Used to select the COM port for the Alpha protocol. Card Reader Search ID Used to select whether or not to search the examination schedule acquired from the MWM for the data items that match the card reader information and to load them as patient information. 3D Positioning Used to select whether or not the 3D position sensor is used. Sensor II-20
[Page 213] (7) Report Refer 1 :o the installation manual for the USDI-772C (2C708-709EN). (8) Work (Irder EndoYsoinal Refer to the installation manual for the USDI-772C (2C708-709EN). 11-21
[Page 214] (9) Other Setting 2D Mode Change Simple Single When the 2D mode switch is pressed, Dual Method display or PW display is returned to Single display with the color mode and zoom settings retained. Example: Dual display for CDI + Single display for CDI --------------------------------- _--------------------------------------------------------------- 2D Single When the 2D mode switch is pressed, the display is returned to 2D Single display with color mode and zoom set to OFF. ANGLE COR 0 degree The angle display of the Doppler marker is switched alternately to 0” and the current angle value every time the Angle Cor dial is pressed. 60 degree The angle display of the Doppler marker is switched in the order 0” + 60” + 120” every time the Angle Cordial is pressed. Auto Refresh on Enable The [QuickScan Update] function is executed Unfreeze automatically when the freeze status is canceled. Disable The [QuickScan Update] function is not executed automatically when the freeze status is canceled. SH Display This item must be set to Disable at all times. Gain Comp w/o CHI Enable When the acoustic power is changed, the 2D gain is also corrected automatically. -------------------------------------------------------------------------------------------------- Disable When the acoustic power is changed, the 2D gain remains as is. Security Mode* Enable The security function (HIPAA) for protecting the patient information is set to ON. .-------------------------------------------------------------------------------------------------- Disable The security function (HIPAA) for protecting the patient information is set to OFF. * The settings take effect when the system is restarted after [Save] is pressed. 11-22
[Page 215] (10) Appl Meas. Number of Heart Cycles 1, 2, 3, 4, 5, 8, 7, 8 Specify the number of cardiac cycles in the for Heart Rate measurement range for the heart rate Measurement measurement. Erase in Cine Erase On Select whether the measurement marker is displayed or not during frame-advance or scroll playback. q : Not displayed 0 : Displayed Background Color Black Gray Specify the background color (black or gray) of the measurement display window. 11-23
[Page 216] (11) Meas. Window Title T Description Basic Measurement Vertical Line Specify whether or not vertical lines are to be displayed Window in the measurement display window. q : Displayed 0 : Not displayed Transparent Mode Specify whether the background of the measurement display window is transparent or not. q : Transparent 0 : Not transparent Realtime Display Specify the display mode of the measurement display window. Current: All the acquired data is displayed in the measurement display window only after the measurement data is fixed. All : All the currently available data is displayed in the measurement display window. 11-24
[Page 217] Title Description -i Application Measurement Vertical Line Specify whether or not vertical lines are to be displayed Window in the measurement display window*‘. q : Displayed 0 : Not displayed _---------------------------------------------------------------------- I- Transparent Mode Specify whether the background of the measurement display window is transparent or not. q : Transparent 0 : Not transparent I- .---------------------- _---------------------------------------------------------------------- Realtime Display Specify the display mode of the measurement display window. Current: All the acquired data is displayed in the measurement display window only after the measurement data is fixed*‘. All : All the currently available data is displayed in the measurement display windowe3. Font Type Specify the font size used in the measurement display window.*’ Value Size:14: The measurement item and unit are 13 pixels and the data is 14 pixels. Value Size:15: The measurement item and unit are 13 pixels and the data is 15 pixels. Bold: All the measurement items, units, and data are 13 pixels in bold. Highlight Specify whether or not the measurement result for the last measured item (the item for which the measurement mark is displayed in the window) is highlighted. (The ACT and TIC measurement displays are not highlighted.)*’ *I : This can be set for each Application Preset. *2: In vascular measurement, the system operates in “All” mode even when “Current” is selected. *3: For OB measurement, the system operates with the “Current” setting even when “All” is selected. 11-25
[Page 218] (12) 2D-Mode Meas. Select Tool to use for 3 Distances (fixed) Volume Measurements JOINT Baseline Right to Left The osseous acetabular angle (a) is displayed as viewed from the right with respect to the baseline, while the cartilaginous acetabular angle (8) is displayed as viewed from the left. Left to Right The osseous acetabular angle (a) is displayed as viewed from the left with respect to the baseline, while the cartilaginous acetabular angle (8) is displayed as viewed from the right. 11-26
[Page 219] (13) D-Mode Meas. RI talc. method*’ Ved RI is calculated using the Ved value. Vmin RI is calculated using the Vmin value. PI talc. methode2 Ved PI is calculated using the Ved value. Vmin PI is calculated using the Vmin value. Flow Volume Tool Auto Trace [Range] (Auto Trace) is set as the default when the Flow Volume, Vel Trace, or PI measurement is started. Vel Trace [Continuous Trace] (Freehand Trace) is set as the default when the Flow Volume, Vel Trace, or PI measurement is started. Flow Volume Calculation Vmean-peak The flow volume is calculated using the time- average value obtained by tracing the peak points (border) of the Doppler waveform. Vmean-mean The flow volume is calculated using the time- average value obtained by tracing the mean points (center of gravity) of the Doppler waveform. Velocity Value Signed Value The measured velocity values are output with [Available after PR]*3 plus/minus signs. Absolute Value The measured velocity values are output as absolute values. *I : This setting is common to RI calculations in the velocity trace, RI trace, and flow volume measurements in the basic measurement mode. *2: This setting is common to PI calculations in the velocity trace, PI trace, and flow volume measurements in the basic measurement mode. “3: Measurement tool available for Velocity Value 11-27
[Page 220] Preset Setting Preset Setting Measurement Tool Signed value Absolute value Velocity Vel Vel = abs (Vel) (DvelocityTool) PG PG Accel measurement Accel Accel (DaccelerationTool) Veil Veil = abs (Vell) Ve12 1 Ve12 = abs (Ve12) Time PGI PGI PG2 PG2 VelocityTrace Vmax Vmax = abs (Vmax) (DsplineTool), (DtraceTool), Vmin Vmin = abs (Vmin) (DlineTool), (DheartCycleTool) Ved Ved = abs (Ved) Vm Vm = abs (Vm) Vm-peak Vm-peak = abs (Vm-peak) Vm-mean Vm-mean = abs (Vm-mean) PI PI RI RI S/D S/D VelocityTrace (Cardiac) VP (Vmax) VP (Vmax) = abs (Vmax) (DsplineTool), (DtraceTool), VM VM = abs (VM) (DlineTool), (DheartCycleTool) = abs (VTI) MPG MPG HR (ECG) HR (ECG) RI (DRITool) 11-28
[Page 221] Preset Setting Preset Setting Measurement Tool Signed value Absolute value PI PI PI Vmax Vmax = abs (Vmax) Vmin Vmin = abs (Vmin) Ved Ved = abs (Ved) Vm 1 Vm = abs (Vm) Vm-peak Vm-peak = abs (Vm-peak) S/D S/D S/D (DSDTool) Vmax Vmax = abs (Vmax) Ved Ved = abs (Ved) dP/dt measurement Tool dP/dt dP/dt dt dt Veil Veil = abs (Vell) Ve12 Ve12 = abs (Ve12) dP/dt (-I,-3) measurement Tool dP/dt dP/dt dt dt Veil Veil = abs (Vell) Ve12 Ve12 = abs (Ve12) DcT measurement Tool Vel Vel = abs (Vel) DcT DcT 1 PHT Velocity Trace with DcT I VP = abs (VP) measurement Tool VM VM = abs (VM) PPG PPG MPG MPG DcT DcT PHT PHT 11-29
[Page 222] (14) Meas Report Measurement Value Mean of The mean of the measurements is displayed as Display Method Measurements the result in the Report screen. Most Recent The most recent measurement is displayed as the Measurement result in the Report screen. Measurement Value Only Mean value is Only the mean of the measurements is displayed Display Default displayed*’ when the Report screen is started. All value is displayed All the measured values and their mean are displayed when the Report screen is started. Display OB Ratios in OB Cl Specify whether the Cl result is displayed or not Reports on the Report screen. FL/AC Specify whether the FL/AC result is displayed or not on the Report screen. FL/BPD Specify whether the FL/BPD result is displayed or not on the Report screen. FL/HC Specify whether the FL/HC result is displayed or not on the Report screen. HUAC Specify whether the HC/AC result is displayed or not on the Report screen. Anatomy in OB Reports Type 1 The anatomy button is displayed on the Report screen in the Type1 format.*2 Type 2 The anatomy button is displayed on the Report screen in the Type2 format.? Hide The anatomy button is not displayed on the Report screen. Trend Format Single The trend graph is displayed in the Single format. Quad The trend graph is displayed in the Quad format. *I : When Carotid 1 or Carotid 2 is selected for Application Preset, “Measurement Value Display Default” is fixed to “Only Mean value is displayed”. “2: For display of Type 1 and Type 2, refer to subsection 12.9.5 “Anatomy function” of the operation manual <measurement volume>. II-30
[Page 223] (15) Meas. Report2 U/S GA On U/S GA is calculated Off U/S GA is not calculated. _----------------------------- GS Select the items to be used for calculating U/S GA. _----------------------------- CRL _----------------------------- BPD _----------------------------- OFD _----------------------------- HC _----------------------------- HA _----------------------------- AC _----------------------------- FTA _----------------------------- AA _----------------------------- FL _----------------------------- AXT 11-31
[Page 224] (16) Meas. Marker Marker Size [2D] Large, Middle, Small Specify the default size of the marker for 2D measurement. _-----------------------------~---------------------------------------------------------------- Auto-change Specify whether the marker is enlarged or not when the distance between the start point and end point exceeds a certain distance. Marker Style [2D] x, +, * Specify the shape of the marker for 2D measurement. Marker Size [M] Large, Middle, Small Specify the default size of the marker for M-mode measurement. Marker Style [M] x, +a * Specify the shape of the marker for M-mode measurement. Marker Size [Doppler] Large, Middle, Small Specify the default size of the marker for Doppler measurement. Marker Style [Doppler] x, +a * Specify the shape of the marker for Doppler measurement. Time/HR Tool Connect line On, Off Specify whether or not the horizontal line display connecting the paired vertical lines in time measurement or HR measurement is to be displayed. 11-32
[Page 225] (17) Body Mark BodyMark Size Regular Regular size Extended t 15times as large as the regular size Inherit Inherit ON Select this check box to cause the body mark to remain displayed when another image mode is selected. Erase in Freeze Off Erase Specify whether the body mark is erased or not when the image is unfrozen. q : The body mark is erased. 0 : The body mark is not erased. 11-33
[Page 226] (18) Annotation Erase in Mode change Erase The annotation is erased when another image mode is selected. Not Erase The annotation is not erased when another image mode is selected. Inherit When image mode is changed from 2D to 2D+PW, the annotation is displayed at the default position on the 2D+PW image. Erase in Freeze off Erase Specify whether or not the annotation is erased when image freezing is released. 0: rf The annotation is erased. 0 : The annotation is displayed. Cursor Type Arrow The arrow cursor is displayed when the annotation function is started. I Cursor The l-beam cursor is displayed when the annotation function is started. 11-34
[Page 227] (19) TIC Smoothing off, 3, 5,7,9, 11, 13, 15 Specify the smoothing of the graph line. Off : Smoothing is not performed. 3 to 15 : Smoothing is performed based on the specified number of points. 11-35
[Page 228] (20) OB Calc. For the OB measurement items for which two or more authors are registered, the author to be used can be selected from a drop-down menu. Gestational Age Specify the author of the formula to be used for calculating the gestational age. The specified author is also used as the default for generating the trend graph. Default Graph Specify the item to be displayed as the default for trend graph creation. EFW Specify the author of the formula to be used for calculating the estimated fetal weight. Extrapolated on Check this to extend the EFW display range. Gestational Age 23 Sestational sac diameter Hellman Rempen -------------. _------------------------------------ CRL Crown-rump length ASUM-VI Nelson Rempen Robinson 11-36
[Page 229] Title I Item Formal nomenclature Author Gestational BPD Biparietal diameter ASUM-VI CFEF Chitty (O-O) Chitty (O-l) Kurtz Merz Nicolaides Rempen Sabbagh Shepard OFD Occipitofrontal diameter ASUM-V2 Chitty Merz Nicolaides HC Head circumference ASUM-VI CFEF Chitty (pltd) Chitty (drvd) Merz _-----------_ ---------------------------------------- -------------------------------------- HA Head area Chitty _-----------_ ---------------------------------------- -------------------------------------- THD Thoracic diameter Hansmann Abdominal circumference ASUM-VI CFEF Campbell Chitty (pltd) Chitty (drvd) Deter Merz Nicolaides Shepard APTD Anteroposterior trunk diameter Tokyo TTD Transverse trunk diameter Tokyo FTA Fetal trunk cross-sectional area Osaka AA Abdominal area Chitty APAD Anteroposterior abdominal Merz diameter 11-37
[Page 230] Gestational Age TAD Transverse abdominal diameter Merz CFEF _-------------- ------------------------------------------------------------------------------- FL Femur length ASUM-VI CFEF Chitty Hohler Merz Nicolaides O’Brien Warda ------------------------------------------------------------------------------- OOD Outer orbital diameter Jeanty Mayden _-------------- ------------------------------------------------------------------------------- Humerus Humerus ASUM-V2 Merz _-------------- ------------------------------------------------------------------------------- Clavicle Clavicle length Yarkoni ------------------------------------------------------------------------------- Kidney Kidney length Bet-tagnoli ------------------------------------------------------------------------------- Radius Radius Met-z ------------------------------------------------------------------------------- Ulna Ulna Jeanty Met-z ------------------------------------------------------------------------------- Tibia Tibia Jeanty Met-z ------------------------------------------------------------------------------- Fibula Fibula Met-z ------------------------------------------------------------------------------- CER Cerebellum Goldstein ------------------------------------------~~~~~~~~------------------------ Foot Foot length Mercer ------------------------------------------------------------------------------- AXT APTD*TTD Tokyo 11-38
[Page 231] Gestational Age Default Measurement items of the AA Graph default graph AC APAD AXT BPD CER CRL Clavicle EFW FL FTA Fibula Foot GS HA 1: (BPD OFD) Humerus’ Kidney OFD Radius TAD THD Tibia Ulna EFW EFW Estimated fetal weight Campbell [AC] Hadlockl [AC, FL] Hadlock [BPD, AC, FL] Hadlock [HC, AC, FL] Hadlock [BPD, HC, AC, FL] Hansman [BPD, THD] Merzl [BPD, AC] Met-22 [AC] JSUM [BPD, AC, FL] Shepard [BPD, AC] Tokyo [BPD, APTD, TTD, FL] Osaka [BPD, FTA, FL] 11-39
[Page 232] (21) D-OB Calc. This menu is used to specify the default author data to be displayed on the trend graph for Doppler OB measurement. Trend Graph : Specify the author to be displayed as the default for trend graph creation Default Graph : Specify the item to be displayed as the default for trend graph creation. Trend Graph Umb A RI Umbilical artery RI Mai Schaffer Umb A PI Umbilical artery PI Mai Harrington Schaffer ------------_ --------------------------------------- .--------------------------------------- MCA RI Middle cerebral artery RI Mai Schaffer ------------_ --------------------------------------- .--------------------------------------- MCA PI Middle cerebral artery PI Mai Harrington Schaffer II-40
[Page 233] Trend Graph Fetal Ao RI Fetal aorta RI Mai Schaffer _----------------- _-----------------------------------. .-------------------------------------- Fetal Ao PI Fetal aorta PI Mai Harrington Schaffer _----------------- _-----------------------------------. .-------------------------------------- Lt Uterin RI Left uterine artery RI Schaffer Lt Uterin PI Left uterine artery PI Schaffer Rt Uterin RI Right uterine artery RI Schaffer Rt Uterin PI Right uterine artery PI Schaffer Default Graph Default Measurement items of the Fetal Ao RI Graph default graph Fetal Ao PI Lt Uterin RI Lt Uterin PI MCA RI MCA PI Rt Uterin RI Rt Uterin PI Umb A RI Umb A PI 11-41
[Page 234] (22) OB Meas. Method of OB weeks Mean of GA The mean of the GA measurement results is used to calculate LMP and EDD. _----------------------------- Current GA The most recent GA measurement result is used to calculate LMP and EDD. _----------------------------- U/S GA U/S GA is used to calculate LMP and EDD. Select Tool to use for OB Ellipse Select the tool for the area/circumference _----------------------------- Circumference measurement (HC, HA, AC, FTA, AA, or TC Continuous Trace Measurements _----------------------------- measurement). Cross _----------------------------- Spline Trace Select Tool to use for OB Continuous Trace Select the tool for the Vel Trace measurement _----------------------------- Vel Trace Measurements (UmbA, MCA, Fetal Ao, Lt Uterin, or Rt Uterin Spline Trace _----------------------------- measurement). Line Trace _----------------------------- Range RI talc. method for OB Ved The Ved value for the measured item is used for RI calculation. _----------------------------- Vmin The Vmin value for the measured item is used for RI calculation. PI talc. method for OB Ved The Ved value for the measured item is used for PI calculation. Vmin The Vmin value for the measured item is used for PI calculation. 11-42
[Page 235] (23) OB Meas. 11-43
[Page 236] Method for AFI 4 Distances AFI is displayed only after all the four distances are measured. .--------------------------------- _--------------------------------------------------------------------- : 4 Distances AFI is displayed even before all the four distances are measured. SD Display 3nlOff Specify whether or not the SD value is displayed in the OB measurement display window. GA Calculated by 3nly GA Chart If the data type is set to “GA” or “FG + GA”, the GA value is calculated based on the GA data regardless of whether or not the clinical age is input. If the data type is set to “FG”, the GA value is not calculated before the clinical age is input. Once the clinical age has been input, the GA value is calculated based on the FG data. If the data type is set to “Identical”, the GA value is calculated based on the Identical data regardless of whether or not the clinical age is input. The trend graph is generated based on the FG data. _--------------------------------------------------------------------- -G Chart with Clinical Age If the data type is set to “FG”, the GA value is not calculated before the clinical age is input. Once the clinical age has been input, the GA value is calculated based on the FG data. If the data type is set to “GA”, the GA value is calculated based on the GA data before the clinical age is input. When the clinical age has been input, the GA value is not calculated. If the data type is set to “FG + GA”, the GA value is calculated based on the GA data before the clinical age is input. Once the clinical age has been input, the GA value is calculated based on the FG data. If the data type is set to “Identical”, the GA value is calculated based on the Identical data regardless of whether or not the clinical age is input. The trend graph is generated based on the FG data. The user can view the detailed contents of the two options for “GA Calculated by”. 11-44
[Page 237] (24) LV (2D) This menu is used to preset the measurement type and the measurement method to be used when 2D-mode LV measurement is started from the Cardiac measurement menu. Method Teichholz Select the LV measurement method to be Cube displayed in the application measurement (2D Gibson mode) menu. Single Plane Biplane Bullet Parallel Tool Type 3 sections Three segments (interventricular septal thickness, LV endocardial diameter, and LV posterior wall thickness) are measured in parallel. 4 sections Four segments (RV diameter, inter-ventricular septal thickness, LV endocardial diameter, and LV posterior wall thickness) are measured in parallel. Parallel Tool Initial Angle -90 to 90 The default angle of the measurement marker for parallel measurement can be specified in IO” increments. MOD Simpson Tool Type Continuous The endocardium can be traced freehand for LV measurement (MOD Simpson method). Spline The endocardium can be traced using the spline trace method for LV measurement (MOD Simpson method). 11-45
[Page 238] (25) Cardiac Doppler This menu is used to preset the measurement method etc, to be used when Doppler-mode measurement is started from the Cardiac measurement menu. E Vel Tool Vel The E wave of the LV inflow is measured using the velocity measurement method. DcT E-wave measurement on the E-wave waveform for LV inflow is performed together with DcT measurement. Output Data of DcT Specify whether PHT is to be measured together Tool with DcT. q With PHT q : PHT is measured together. 0 : PHT is not measured together. 11-46
[Page 239] Method Teichholz Specify the LV measurement method to be displayed in the application measurement (M Cube mode) menu. Gibson Next Sequence Type 1 Each time the m switch is pressed in LV measurement, the item to be measured next is selected automatically in the range from the inter-ventricular septum to the posterior wall both for end diastole and end systole. 2 When the m switch is pressed in LV measurement, the endocardium only is selected as the item to be measured next for end systole. Parallel Tool Type 3 sections Three segments (inter-ventricular septal thickness, LV endocardial diameter, and LV posterior wall thickness) are measured in parallel. 4 sections Four segments (RV diameter, inter-ventricular septal thickness, LV endocardial diameter, and LV posterior wall thickness) are measured in parallel. 11-47
[Page 240] (27) LV Mass. This menu is used to preset the measurement type and the measurement method to be used when LV Mass measurement is started from the Cardiac measurement menu. LV Mass Method LV Mass AL Specify the LV Mass measurement method to be LV Mass TE displayed in the application measurement menu (Extra). Mass Calc Method [2D] ASE-Cube Specify the LV Mass calculation method to be Penn-Cube used in the LV measurement (2D mode). AV-Cube Teichholz Not Display The LV Mass data is not displayed in the measurement display window or on the Report screen. Mass Calc Method [M] ASE-Cube Specify the LV Mass calculation method to be used in the LV measurement (M mode). Penn-Cube AV-Cube Teichholz Not Display The LV Mass data is not displayed in the measurement display window or on the Report screen. 11-48
[Page 241] (28) Carotid1 Doppler Display type of Velocity*’ Velocity The velocity (cm/s) is displayed as the measurement result. _----------------------------- Frequency The frequency (kHz) is displayed as the measurement result. Display items of Velocity Vmax [Fmax] Specify the items to be displayed as the results of Trace Velocity Trace measurement. Vmin [Fmin] The settings for Vmax [Fmax], Vmin [Fmin], Ved Ved [Fed] [Fed], and Vmean [Fmean] are fixed to q Vmean [Fmean] PI q : The item is displayed. RI 0 : The item is not displayed S/D RI talc. method*3 Ved RI is calculated based on Ved. Vmin RI is calculated based on Vmin. PI talc. methods4 Ved PI is calculated based on Ved. _----------------------------- Vmin PI is calculated based on Vmin. Combinations of CCA Prox Specify one of the CCA measurement regions for ICAICCA Mid calculating ICAKCA. Dist ALL*5 ICA Prox Specify any number of the ICA measurement regions for calculating ICAKCA. Mid r Dist 11-49
[Page 242] *I : [Velocity] is selected as the default. *2: All the items are set to q (to be displayed) as the default. *3: [Ved] is selected as the default. *4: [Vmin] is selected as the default. “5: When “ALL” is selected, setting for ICA is not necessary. Combinations of “Prox-Prox”, “Mid-Mid”, and “Dist-Dist” are set automatically. (29) Carotid1 Meas. Tool Area*’ Ellipse The area is measured using the Ellipse method. Continuous Trace The area is measured using the Continuous Trace Spline Trace The area is measured using the Spline trace Cross The area is measured using the Cross method. Velocity Trace’;’ Continuous Trace The velocity is measured using the Continuous Trace method. Range The velocity is measured using the Auto Trace method with the range specified. Spline Trace The velocity is measured using the Spline Trace *I : “Ellipse” is selected at the factory before shipment. *2: “Range” is selected at the factory before shipment. II-50
[Page 243] (30) Carotid2 Doppler Display type of Velocity*’ Velocity The velocity (cm/s) is displayed as the measurement result. _----------------------------- Frequency The frequency (kHz) is displayed as the measurement result. Display items of Velocity Vmax [Fmax] Specify the items to be displayed as the results Trace Vmin [Fmin] of Velocity Trace measurement. The settings for Vmax [Fmax], Vmin [Fmin], Ved Ved [Fed] [Fed], and Vmean [Fmean] are fixed to q Vmean [Fmean] PI q : The item is displayed. RI 0 : The item is not displayed. S/D RI talc. method*3 Ved RI is calculated based on Ved. _----------------------------- .-------------------------------------------------------------- Vmin RI is calculated based on Vmin. PI talc. method4 Ved PI is calculated based on Ved _----------------------------- Vmin PI is calculated based on Vmin. “I : [Velocity] is selected as the default. *2: All the items are set to q (to be displayed) as the default. “3: [Ved] is selected as the default. *4: [Vmin] is selected as the default. 11-51
[Page 244] (31) Carotid1 Meas. Type Doppler* Velocity Trace Tracing is performed in the Velocity Trace mode. _---------------------------------------------------------------------------------------- PS Velocity The peak systolic velocity is measured in Velocity measurement mode. _---------------------------------------------------------------------------------------- ED Velocity The end-diastolic velocity is measured in Velocity measurement mode. * “Velocity Trace” is selected as the default 11-52
[Page 245] (32) Carotid2 Meas. Type Title Item I Description Right CV Doppler*’ Velocity Trace The velocity is traced using the Velocity Trace _____________---________________________------------------------------------------------------ PS Velocity The peak systolic velocity is measured using the Velocity measurement method. _____________---________________________------------------------------------------------------ ED Velocity The end diastolic velocity is measured using the Velocity measurement method. Right CV %Stenosis*2 %Stenosis Area The stenosis ratio is calculated using the %Stenosis Area method. _____________---________________________------------------------------------------------------ OXStenosis Distance The stenosis ratio is calculated using the OhStenosis Distance method. Left CV Doppler*3 Velocity Trace The velocity is traced using the Velocity Trace PS Velocity The peak systolic velocity is measured using the Velocity measurement method. ,____________---________________________------------------------------------------------------ ED Velocity The end diastolic velocity is measured using the Velocity measurement method. Left CV %Stenosis*4 %Stenosis Area The stenosis ratio is calculated using the %Stenosis Area method. .____________---________________________------------------------------------------------------ %Stenosis Distance The stenosis ratio is calculated using the OHStenosis Distance method. “I : “Velocity Trace” is selected at the factory before shipment. *2: “%Stenosis Distance” is selected at the factory before shipment. “3: “Velocity Trace” is selected at the factory before shipment. “4: “%Stenosis Distance” is selected at the factory before shipment. 11-53
[Page 246] (33) Carotid2 Meas. Tool This menu is used to preset the default measurement method to be started when the Velocity Trace measurement switch is pressed in the Carotid2 menu of Vascular measurement mode. Velocity Trace’;’ Continuous Trace The area is measured using the Continuous Trace _-----------------------------~---------------------------------------------------------------- Range The velocity is measured using the Auto Trace method with the range specified. Spline Trace The velocity is measured using the Spline Trace “I : “Range” is selected at the factory before shipment. 11-54
[Page 247] (34) ACT (option) If the optional ACT measurement program is installed, defaults for ACT measurement can be preset. On this screen, 3 pages are provided for ACT measurement. 0 Page 1 Annular Tracking Disable Disables the annular tracking assessment function during (Advanced) ACT measurement. -------------------------------------------------------------------------- Enable Enables the annular tracking assessment function during ACT measurement. The [Advanced], [Copy to File], [Annular Disp.], and [Trajectory Disp.] switches are displayed on the touch panel for LV ACT. Contour Display Off Turns OFF the contour display. -------------------------------------------------------------------------- On Turns ON the contour display. Annular Display on Off Turns OFF display of the annular position mark. -------------------------------------------------------------------------- Advanced mode” On Turns ON display of the annular position mark. Average Off Specifies the measurement and calculation methods for LV functions. The LV functions are measured and calculated using the data for only the cardiac cycle (between two successive end diastoles) containing the currently displayed frame. -------------------------------------------------------------------------- On Specifies the measurement and calculation methods for LV functions. The LV functions are measured and calculated by averaging the data for the specified number of cardiac cycles. *I : This item is enabled only when “Annular Tracking” is set to “Enable”. II-55
[Page 248] l Page 2 Heart Beat [Beat] 1 to 5 Sets the number of cardiac cycles included in the frame range (from the measurement start frame to the measurement end frame). Graph fitting Off Turns OFF graph fitting using the Fourier approximation. On Turns ON graph fitting using the Fourier approximation. The order of graph 1 to50 Specifies the order of the Fourier approximation when fitting “Graph fitting” is set to “On”. RR Interval 0, 5, 10, 15, 20, 25, Sets the acceptable time difference between the R-R Differential Time 30, 35, 40, 45, 50 intervals in the two-chamber and four-chamber cross- Limit [%] sectional images in Biplane Volume measurement.*’ *I : For details, refer to subsection 11.2 “Operating Methods” in the application volume of the operation manual. 11-56
[Page 249] l Page 3 Graph 1 to Graph 4 X-Pos. Left etc.*’ Specifies the types of graphs and measurement results to be displayed on the screen. It is also possible to select not to display graphs or measurement results. If the same graph type is selected for more than one of “Graph 1” to “Graph 4”, only a single graph of the selected type is displayed. The graphs are displayed in a row at the top of the screen, with Graph 1 located at the leftmost position. *I : Refer to subsection 11.5 “Arithmetic Operation Method for ACT Measurement” of the operation manual <measurement volume>. 11-57
[Page 250] 12. BACKING UP THE USERS DATA 12.1 Backing Up the Preset Data 12.1.1 Backup The information in “D\aplio\user” must be backed up. These files can be backed up together with the system-specific information on an MO using the user maintenance and the service menu. 12.1.2 Restoration The backed up information is restored to “D:\aplio\user”. Only the information included in the service menu need be restored. Restoration can be performed only when the system information of the system where the backup was performed and that where the restoration is to be performed match. If an attempt is made to restore information to a system other than the system where backup was performed, a warning dialog is displayed. Therefore, perform restoration carefully to the correct system. 12.2 Backing Up the Image Data 12.2.1 Backup Back up the image data registered on the hard disk by saving it on the MO using Patient Browser. The data is saved in the DICOM format. For the operating procedures for Patient Browser, refer to the operation manual ccApplications>>. 12.2.2 Restoration Restore the image data saved on the MO by transferring it to the hard disk using Patient Browser. The data must be saved in the DICOM format. For the operating procedures for Patient Browser, refer to the operation manual ccApplications>>. 12-1
[Page 251] 13. NETWORK CONNECTION 13.1 Changing the PC Name If it is necessary to change the PC name for compatibility with the network environment, the database must be deleted and then created again. (1) Connect the network cable to the corresponding connector at the back of the system and turn ON the power supply to start the system. After starting the system, display the service menu. Input the command “stopzjs” on Utilities Command Line and click Go to stop the system software. The service screen also disappears. (2) Select the Start button + Program + Accessory + Command prompt and then execute the “tus-rmdb.js” command to delete the database. After performing deletion, confirm that the database is deleted completely by following the procedures below. l Right-click the Start button and select Explorer from the menu. Confirm that no file or folder exist in the directories “D:\database” and “E:\lmage”. l If any file or folder exist in the directory “D:\database”, select all with the Ctrl + A keys to delete them. Perform the same procedures for the directory “E:\lmage”. 13-1
[Page 252] (3) Set the network. Perform settings for the PC name and TCP/IP as indicated. For these settings to become effective, the system must be restarted. However, the system must not be restarted yet. When the PC name is changed, input “tus-chnm.js PC name” at the command prompt (2) above. (This work also can be performed by selecting Control panel + System + Advanced + Environment setting and by setting the PC name to VERSANT-DBID-NODE.) Terminate the command prompt. (4) Reboot the system by clicking the Start button and selecting Restart from the menu (5) After restarting the system, execute the “tus-mkdb.js” command to create the database again. The display shown below is then displayed automatically. Confirm that the five databases are created. The display is an example with the PC name set as “toshibal”. <<Database list display example>> VERSANT Utility DBLIST Version 5.2.2.0.0 Copyright (c) 1989-l 999 VERSANT Corporation ID = 6 DB name = ArchiveStore@toshibal creator = toshiba date created = Thu Apr 05 15:46:51 2001 db type = GROUP DATABASE db version = 5.2.2.0 ID = 7 DB name = Local@toshibal creator = toshiba date created = Thu Apr 05 15:47:29 2001 db type = GROUP DATABASE db version = 5.2.2.0 ID = 8 DB name = SchedulerStore@toshibal creator = toshiba date created = Thu Apr 05 15:48:15 2001 db type = GROUP DATABASE db version = 5.2.2.0 ID = 9 DB name = PimsExamArch@toshibal creator = toshiba date created = Thu Apr 05 15:48:41 2001 db type = GROUP DATABASE db version = 5.2.2.0 ID = 10 DB name = ExchangeboardQtoshibal creator = toshiba date created = Thu Apr 05 15:49:59 2001 db type = PERSONAL DATABASE db version = 5.2.2.0 13-2
[Page 253] (6) Confirm that the network is connected. (7) Shut down the system by clicking the Start button and selecting Shutdown from the menu. Then restart the system and confirm that the system starts normally. 13-3
[Page 254] 14. UPGRADING THE SOFTWARE This subsection describes the standard procedure for upgrading the software. If special operation is required, the corresponding instructions will be provided. (1) Turn ON the power supply to start the system. Perform backup as required and display the service menu screen. Input the command “stopzjs” on the Utilities Command Line and click Go to stop the system software. The service menu screen disappears. (2) Select the Start button + Program + Accessory + Command prompt and then execute the “tus-stopAplio.js” command. Terminate the command prompt. (3) Reboot the system by clicking the Start button and selecting Restart from the menu. (4) After restarting the system, upgrade the software. An example of the upgrade procedure is given below. l Delete the “Zm/Orca Combined...” and “Zhui” applications using “Add/Remove Programs” of “Control Panel”. l Start Explorer. If “C:\Hui” or “C:\Zdt” exists, delete the whole folder. l Execute the “.setup.bat” command on the APPLI DISK MO. (5) After upgrading the software, execute the “tus-restartAplio.js” command and terminate the command prompt again. (6) Shutdown the system by clicking the Start button and selecting Shutdown from the menu. Then restart the system and confirm that it starts normally. (7) Display the service menu and perform restoration as necessary. If restoration is performed, the system shuts down. Restart the system and complete the procedure. 14-1
[Page 255] 15. SELECTION OF THE PANEL SWITCH LAYOUT The panel switch layout can be selected from among the following five types according to user preference. (1) TMSJ abdomen A (2) TMSJ abdomen B (3) Cardiac (4) TMSE/TAMS abdomen (5) TAMS left-handed 15.1 Standard Panel Switch Layout The switches on the main panel are arranged as shown below at the factory. 15.2 Changing the Panel Configuration File (1) Start Explorer from the service menu. (2) Open \\config\TUSPanel\TUSPanelPattern. The following batch files are displayed. Batch file name I Purpose Standard.bat 1 Sets the panel switch layout to “Standard”. TMSJ-Abd-A. bat Sets the panel switch layout to “TMSJ abdomen A”. TMSJ-Abd-B.bat 1 Sets the panel switch layout to “TMSJ abdomen B”. Cardiac.bat Sets the panel switch layout to “Cardiac”. TMSE-TAMS-Abd.bat Sets the panel switch layout to “TMSE/TAMS abdomen”. TAMS-Left-Hand.bat 1 Sets the panel switch layout to “TAMS left-handed”. (3) Execute the desired batch file. (4) Reboot the system. 15-1
[Page 256] 15.3 Panel Switch Layout and Switch Functions After Configuration Change Replace the key tops as shown in the figure according to the selected layout. Confirm that the configuration has been changed correctly. If the configuration has not been changed correctly, correct it. (1) TMSJ abdomen A (TMSJ-Abd-A.bat) (Pattern A) The functions of the seven locations indicated in the figure below are changed. For cl > and c2>, replace the elliptic label below the switch. For c3> to c7>, replace the key tops. Cl> 1 2D FREQ 1 Same as the 2D FREQ switch before the layout change <2> 2D FOCUS Same as the 2D FOCUS switch before the layout change <3> UPDATE Same as the UPDATE switch before the layout change <4> 1 FULL 1 Same as the FULL switch before the layout change <5> 1 DUAL 1 Same as the DUAL switch before the layout change <6> 1 STORE 1 Same as the STORE switch before the layout change <7> 1 FREEZE 1 Same as the FREEZE switch before the layout change 15-2
[Page 257] TMSJ abdomen B (TMSJ-Abd-B.bat) (Pattern B) The functions of the eight locations indicated in the figure below are changed. For cl> and c2>, replace the elliptic label below the switch. For <3> to <8>, replace the key tops. <I> 1 2D FREQ 1 Same as the 2D FREQ switch before the layout change <2> 1 2D FOCUS 1 Same as the 2D FOCUS switch before the layout change <3> 1 UPDATE 1 Same as the UPDATE switch before the layout change <4> 1 FULL 1 Same as the FULL switch before the layout change c5> 1 DUAL 1 Same as the DUAL switch before the layout change <6=- OUTPUT D Same as the OUTPUT C switch before the layout change <7> 1 OUTPUT C 1 Same as the STORE switch before the layout change <8=- 1 FREEZE 1 Same as the FREEZE switch before the layout change 15-3
[Page 258] (3) Cardiac (Cardiacbat) (Pattern C) The functions of the five locations indicated in the figure below are changed. Replace the key tops for cl>, c2>, and c5>. cl> 1 White switch 1 No function is assigned <2> 1 CHI 1 Sets the system to CHI mode. <3> OUTPUT B Same as the OUTPUT A switch before the layout change (Auto Print) <4> OUTPUT A Same as the OUTPUT B switch before the layout change (B/W printer) <5> 1 VCR 1 Same as the VCR switch before the layout change 15-4
[Page 259] (4) TMSE/TAMS abdomen (TMSE-TAMS-Abd.bat) (Pattern D) The functions of the three locations indicated in the figure below are changed. Replace the key tops for c2> and c3>. <I> White switch No function is assigned. (DICOM Server switch in the standard configuration) <2> 1 OUTPUT D 1 Same as the DICOM Server switch before the layout change <3> OUTPUT C Same as the OUTPUT C switch before the layout change 15-5
[Page 260] (5) TAMS left-handed (TAMS-Left-Handbat) (Pattern E) The functions of the eight locations indicated in the figure below are changed. Do not replace the key tops for cl > and c4>. I I I c4> <5> <6> White switch No function is assigned. (DICOM Server switch in the standard configuration) CHI Sets the system to CHI mode. White switch Same as the DICOM Server switch before the layout change <4> OUTPUT A Same as the STORE switch before the layout change <5> FREEZE Same as the FREEZE switch before the layout change <6> OUTPUT B Same as the OUTPUT A switch before the layout change (Auto Print) c7> OUTPUT D Same as the OUTPUT C switch before the layout change <8> OUTPUT C Same as the OUTPUT B switch before the layout change (B/W printer) 15-6
[Page 261] 15.4 Attaching the Label Attach the label that matches the selected pattern (one from among A to E) to the main panel. Label attachment position 15-7
[Page 262] 16. DATE/TIME SETTING CHANGE To change the system date and time, use the procedure described below. 16.1 To Change the System Time Only (1) press E on the main panel. (2) Select [General Preset] from the System Setting pop-up menu. (3) In the “Time” field of the General Setting screen, change the setting of hour/minute/second. 16.2 To Change the System Date and Time (1) Activate the maintenance menu (for service engineers). For the activating procedures, refer to the service manual (2D730-149E). (2) Select [Command Line] in “Utilities”. 16-1
[Page 263] (3) Use the keyboard to enter “stopzjs” in the Command field and select [Go]. * Execution of this command takes approximately 2 minutes. Do not perform other key operations until execution of the command is completed. (4) The Aplio software is terminated and a Windows screen appears. (5) Remove the rubber cap at the left side of the operating panel and connect the USB mouse. * This is because the m and m switches are not operable when the Aplio software is terminated. The trackball is operable though. (6) Select [Start] + [Programs] + [Accessories] + [Windows Explorer] to open the Windows Explorer window. 16-2
[Page 264] (7) Select [My Computer] + [Control Panel] + [Date/Time] to open the Date/Time window. Change the date and time in this window. CAUTION: Never change the system date or time to differ from the factory setting by 24 hours or more. If the system date and time is changed from the factory setting by 24 hours or more, system performance cannot be assured and the HDD will have to be replaced. (8) Select [Start] + [Shutdown...] + [Shutdown] to shut down the system. (9) Start up the system again. 16-3
[Page 265] 17. 3D POSITION SENSOR 17.1 3D Position Sensor Unit The position sensor receiver of the 3D position sensor detects the magnetic field emitted by the position sensor transmitter. Based on this information from the magnetic field, the control unit of the 3D position sensor measures the three-dimensional position and direction of the position sensor receiver with respect to the transmitter, and transmits the results to the diagnostic ultrasound system (hereinafter referred to as “the system”). (2) Functions As shown in figure 17-1, the position sensor receiver detects the magnetic field in the three directions generated by the position sensor transmitter, and the control unit measures the three-dimensional position (x, y, z) of the position sensor receiver in the orthogonal 3D space around the position sensor transmitter. The control unit also measures the directions (a, 8, r) that represent the rotation angles of the coordinate system formed by the position sensor receiver as shown in figure 17-2. The control unit then transmits the obtained positional and directional data to the system. receiver Figure 17-1 Positional data Figure 17-2 Directional data 17-1
[Page 266] (3) Operations As shown in figure 17-3, the 3D position sensor consists of the control unit, the position sensor transmitter, the position sensor receiver, and the power unit. They are connected to the system by a cable and a g-pin D-sub connector. The power unit is supplied with 100/240 VAC by the system and supplies DC voltage to the control unit. The control unit receives the positional data transmission request command from the system and transmits a pair of positional and directional data sets (x, y, z: a, 8, y) via a serial interface (RS-232C) at IO-ms intervals. The data must be input to the COM2 port on the system, which is a serial interface*. The control unit calculates the position and directions based on the amount of current that is induced in the position sensor receiver in each of the three magnetic field directions x, y, and z that are generated at different times. * For a protocol, select a COM port not used by the VCR or 3D position sensor. The FLYKTDBY switch, which is intended to switch the status of the control unit between running and standby, must be set to FLY at all times. 1 AC1 00-240V Aplio Power unit COM2 I I D-Sub9 RS-232C POWER &OR b receiver Control unit rl Figure 17-3 System configuration 17-2
[Page 267] The intensity of the magnetic field generated by the position sensor transmitter becomes lower as the measurement point is moved away from the transmitter. The position sensor receiver must be within the spherical range from 21 cm to 76 cm from the center of the position sensor transmitter as shown in figure 17-4. In addition, the position sensor receiver must be within the front hemisphere of the magnetic field, which is one of the two hemispheres defined by the positive and negative directions on the x-axis, in order to assure the accuracy of the positional data. If the position sensor receiver is moved across the front and rear hemispheres during data acquisition, image reconstruction becomes impossible. ? :----- ---_ -. -. ‘\ ‘\ ‘\\ ‘1 Transmitter of the ‘, position sensor , 1 I _-: ir’ : I’ I’ If I’ /’ Back I’ , /’ /’ ,’ ,’ __dC .- i---- Figure 17-4 Position sensor transmitter The position sensor receiver is attached to a transducer using the dedicated attachment as shown in figure 17-5 (the transducer shown in the figure is the 375AT). Refer to the relevant operation manual for the procedure for attaching the position sensor receiver. Attachment Position sensor receiver Figure 17-5 Attaching the position sensor receiver to a transducer 17-3
[Page 268] The acquired 2D ultrasound image is accompanied with the size information on the 2D image itself and the positional and directional information of the position sensor receiver with respect to the position sensor transmitter. This information is used to reconstruct a 3D image. Therefore, it is necessary to correct the positional relationship between the position sensor receiver and the 2D-ultrasound image using a conversion table called a calibration matrix. The shape of the transducer is taken into consideration when the calibration matrix is prepared. Figure 17-6 Positional data 17-4
[Page 269] 17.2 Magnetic Field Distortion Check Tool The magnetic field distortion check tool is used to confirm that the magnetic field generated by the transmitter of the position sensor is not affected by the magnetic field environment in which the Aplio is installed. This check should be performed in the following cases. l By the service personnel during the installation of the Aplio system l By the user after the system installation environment is changed. l By the service personnel during maintenance work. l By the user as a part of daily checks. (2) Functions Since the operating principle of the 3D position sensor is based on the magnetic field as described in subsection 17.1, the magnetic field distortion in the area in which data acquisition is performed must be below the specified level. The magnetic field distortion check tool is designed to check whether the magnetic field distortion in the possible acquisition range is acceptable or unacceptable. (3) Procedure Place the position sensor receiver at each of positions A, B, and C that are located on a straight line at regular intervals. At each position, the receiver must be placed in the same orientation. Based on the positional and directional information at each check position, the magnetic field distortion check tool obtains the following three parameters: 1) direction at each position, 2) vector(p) between A and B, and 3) vector(q) between B and C. If the magnetic field is not distorted, these three parameters should be equal. Based on this rule, the check tool analyzes any errors in the actual magnetic field environment as compared with the ideal and determines whether or not the level of magnetic field distortion is acceptable. Figure 17-7 Principle of magnetic field distortion check 17-5
[Page 270] (4) Magnetic field distortion check For details of the magnetic field distortion check procedure, refer to one of the manuals listed below. (a) Installation manual for the UIFR-771A (2C708-634E) (b) Operation manual for the UIFR-771A (2B708-114E) (c) Operation manual <<Applications volume>> for the SSA-700A/SSA-770A 17-6
[Page 271] LEGAL MANUFACTURER 1385, SHIMOISHIGAMI, OTAWARA-SHI, TOCHIGI-KEN 324-8550, JAPAN "Made for Life" is a trademark of Toshiba Medical Systems Corporation.